脑性瘫痪儿童髋关节移位超声测量的信度与准确度
本研究纳入34例脑性瘫痪儿童共59个髋关节,其中43个髋关节重复扫描以评估重测信度,两名评估者手动测量超声髋关节移位百分比(MPUS),并与骨盆正位X线片测量的金标准MP比较。结果显示重测、评估者内和评估者间ICC分别为0.90、0.94和0.82,测量标准误≤3.0%,两名评估者MPUS与X线MP的平均绝对差分别为6.2%和7.6%,在10%临床可接受误差内的比例分别为84.7%和73.7%。结论认为超声扫描具有可重复性,MPUS可被可靠且准确地测量。
要点
- 超声髋关节移位百分比(MPUS)重测信度ICC为0.90,评估者内信度ICC为0.94,评估者间信度ICC为0.82。
- MPUS测量标准误在重测、评估者内和评估者间评估中均≤3.0%。
- 评估者R1的MPUS与X线MP平均绝对差为6.2%±4.9%,84.7%的测量在10%临床可接受误差内;评估者R2为7.6%±6.1%和73.7%。
核心结论
- 超声扫描具有可重复性,超声髋关节移位百分比(MPUS)可被可靠且准确地测量,有望用于脑性瘫痪儿童髋关节移位的评估。
研究设计与样本
- 前瞻性体内先导研究,纳入34例脑性瘫痪儿童(23例男孩、11例女孩,年龄8.9±3.1岁),共59个髋关节接受一次超声扫描,其中43个髋关节扫描两次以评估重测信度;两名评估者(R1和R2)手动测量MPUS,以骨盆正位X线片测量的髋关节移位百分比(MP)为金标准。
诊断标准与关键数值
- 重测信度(R1)ICC2,1为0.90,评估者内信度(R1)ICC2,1为0.94,评估者间信度(R1 vs R2)ICC2,1为0.82;三种评估的MPUS测量标准误均≤3.0%;MPUS与X线MP的平均绝对差:R1为6.2%±4.9%,R2为7.6%±6.1%;MPUS在10%临床可接受误差内的比例:R1为84.7%,R2为73.7%。
临床应用与阅片要点
- 作者认为超声扫描可重复,MPUS可被可靠且准确地测量;测量前需手动选取感兴趣图像、裁剪感兴趣区域并去除髋部超声图像上的软组织,随后由定制软件自动测量MP。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Hip migration percentage (MP) measured on anteroposterior pelvis radiographs is the gold standard to assess the severity of hip displacement in children with cerebral palsy (CP). Repeated exposure of these children to ionizing radiation under a hip surveillance program is undesirable. Recently, a semi-automatic approach to measure MPUS on ultrasound (US) images was validated in a phantom study. This pilot in vivo study applied the previous phantom method and aimed to determine the reliability and accuracy of the MPUS.
METHODS: Thirty-four children (23 boys and 11 girls) aged 8.9 ± 3.1 y old and diagnosed with CP were recruited. A total of 59 hips were scanned once, while 43 of these were scanned twice to evaluate the test-retest reliability. Two raters (R1 and R2) manually measured MPUS; procedures included selecting images of interest, cropping a region of interest and removing soft tissues on hip US images. Custom software was developed to measure MP automatically after the manual pre-image processing.
RESULTS: The intra-class correlation coefficients (ICC2,1) for the test-retest (R1), intra-rater (R1) and inter-rater (R1 vs R2) reliabilities were 0.90, 0.94 and 0.82, respectively. The standard error of measurement of MPUS for all three evaluations was ≤3.0%. The mean absolute difference between MPUS and MPX-ray and the percentage of MPUS within clinical acceptance error of 10% for R1 and R2 were (R1: 6.2% ± 4.9%, 84.7%) and (R2: 7.6% ± 6.1%, 73.7%), respectively.
CONCLUSION: This study demonstrated that US scans were repeatable and MPUS could be measured reliably and accurately.
引用
- Pham TT, Le LH, Andersen J, et al. Reliability and Accuracy of Ultrasound Measurement of Hip Displacement in Children with Cerebral Palsy. Ultrasound Med Biol. 2024. PMID: 39098471. doi:10.1016/j.ultrasmedbio.2024.07.002
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。