超高分辨率与标准分辨率光子计数探测器CT血管成像在尸体灌注模型中对股动脉支架成像的比较
在单具新鲜冷冻人体尸体上建立股部体外灌注模型并置入7枚外周支架,采用剂量匹配的120kVp光子计数探测器CT血管成像,比较标准分辨率模式与超高分辨率模式在低、中、高剂量(CTDIvol=3、5、10 mGy)及四种血管卷积核下的图像质量与管腔可见性。结果显示两种模式管腔可见性相同,超高分辨率模式在(超)锐卷积核下对比噪声比显著提高,且读者主观偏好超高分辨率图像。
要点
- 标准分辨率与超高分辨率模式在各剂量水平下管腔可见性相同。
- 超高分辨率模式联合低剂量与超锐卷积核时对比噪声比提升幅度最大。
- 采用超锐卷积核时,超高分辨率低剂量扫描的对比噪声比与标准分辨率高剂量扫描相当。
核心结论
- 超高分辨率模式的对比噪声比提升在联合低辐射剂量与超锐重建时最高,且超高分辨率模式的主观图像质量总体优于标准分辨率图像,提示进一步降低剂量的潜力。
研究设计与样本
- 在单具新鲜冷冻人体尸体上经腹股沟与腘窝入路用蠕动泵建立股部体外灌注,于股浅动脉置入7枚外周支架,行剂量等效120kVp光子计数探测器CT血管成像,标准分辨率与超高分辨率模式各在低、中、高剂量(CTDIvol=3、5、10 mGy)下采集并用四种血管卷积核重建。
诊断标准与关键数值
- 管腔可见性在所用剂量水平下标准分辨率与超高分辨率模式相同。
- BV60(超锐卷积核)在3 mGy时对比噪声比超高分辨率模式19.9 ± 1.9 vs 标准分辨率模式15.7 ± 1.6,p < 0.046。
- BV76(超锐卷积核)在3 mGy时对比噪声比超高分辨率模式8.0 ± 0.6 vs 标准分辨率模式5.4 ± 0.3,p < 0.001。
- BV76对比噪声比相对提升低剂量扫描48% vs 高剂量扫描36%,p < 0.033。
- 采用超锐卷积核时,超高分辨率低剂量扫描对比噪声比8.0 ± 0.6 vs 标准分辨率高剂量扫描9.1 ± 1.1,p > 0.760。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND AIMS: The aim of this study was to investigate the imaging performance and quality differences of PCD-CT in standard resolution mode (SR) versus ultra-high resolution mode (UHR) in the lower extremity runoff of dose-matched CTAs in a human cadaveric model.
METHODS: Extracorporeal perfusion of the upper leg was established in one fresh-frozen human cadaver via inguinal and popliteal accesses using a peristaltic pump. Seven peripheral stents were deployed in the SFA. Photon-counting CTAs were performed under contrast perfusion in SR and UHR mode with dose-equivalent 120kVp acquisition protocols (low-/ medium-/ high-dose: CTDIVol=3, 5, 10 mGy) and reconstructed with four vascular convolution kernels. Lumen visibility and contrast-to-noise ratio were compared using analyses of variance. Subjective image quality was assessed using a pairwise, forced-choice comparison software.
RESULTS: Lumen visibility was equal for SR and UHR at the used dose levels. CNR increase by UHR was significant for (ultra-)sharp convolution kernels BV60 (3 mGy; UHR vs. SR, 19.9 ± 1.9 vs. 15.7 ± 1.6, p < 0.046) and BV76 (8.0 ± 0.6 vs. 5.4 ± 0.3, p < 0.001). The relative CNR increase was higher for low-dose than high-dose scans (BV76: 48% vs. 36% at high dose, p < 0.033). The CNR of the low-dose scan in UHR mode was comparable to the high-dose scan in SR mode when the ultra-sharp kernel was used (8.0 ± 0.6 vs. 9.1 ± 1.1, p > 0.760). Among UHR examinations, a significant increase in CNR could only be measured in BV76 (8.0 ± 0.6 (3 mGy) vs. 12.4 ± 0.9 (10 mGy), p < 0.001). Readers preferred subjective image quality of UHR for all kernels with BV76 being ranked highest.
CONCLUSION: The CNR increase in UHR mode is highest when combining low radiation dose and ultra-sharp reconstructions. Meanwhile, the subjective image quality in UHR mode generally supersedes SR images, suggesting further dose reduction potential.