文献速递骨肌MR

深度学习增强的加速2D TSE与3D超分辨率Dixon TSE用于快速全面膝关节评估

Deep Learning-Enhanced Accelerated 2D TSE and 3D Superresolution Dixon TSE for Rapid Comprehensive Knee Joint Assessment.
PubMed 文献 · Invest Radiol · 2024-08-28

前瞻性纳入19例有症状成人,在3T扫描仪上比较深度学习重建的4倍加速2D TSE和基于深度学习增强6倍加速2D Dixon TSE的3D超分辨率重建与传统2倍加速2D TSE的膝关节MRI。DL增强2D TSE图像质量和诊断性能与标准方案相似;3D SRR Dixon TSE可行,但软骨、肌腱和骨受重建和混叠伪影影响,诊断信心较低;两种DL方案均使采集时间缩短44%。

要点

  • DL增强4倍加速2D TSE的图像质量和解剖可见性与标准2倍加速2D TSE非常相似。
  • 3D SRR Dixon TSE总体图像质量评分较低,因边缘锐度下降和伪影(P < 0.001)。
  • 标准2D TSE评估肌腱的诊断信心显著高于3D SRR Dixon TSE(P < 0.01)。

核心结论

  • DL增强4倍加速2D TSE方案提供与标准2D方案相似的图像质量和诊断性能;DL增强6倍加速2D Dixon TSE的3D SRR用于多对比3D膝关节MRI可行,诊断性能与标准2倍加速2D膝关节MRI相当;但重建和混叠伪影需进一步解决以确保软骨、肌腱和骨的可靠显示和评估;两种DL增强方案均使检查较传统2倍加速常规2D TSE膝关节MRI快44%。

研究设计与样本

  • 前瞻性、经伦理批准的研究,纳入19例有症状成人受试者,在临床3T扫描仪上行膝关节MRI;每位受试者单次 session 内接受3种DL增强采集方案:临床标准2倍平面内并行成像加速2D TSE方案(5个序列,11分23秒)作为参考、DL重建的4倍加速2D TSE方案(2倍并行成像联合2倍同时多层加速,5个序列,6分24秒)、基于DL增强6倍加速(3倍并行成像联合2倍同时多层)2D Dixon TSE MRI的3D SRR方案(6个绕相位编码轴旋转的各向异性2D Dixon TSE采集,6分24秒)。

关键影像表现/征象

  • DL增强4倍加速2D TSE方案提供与标准2D TSE方案非常相似的图像质量和解剖可见性。
  • 3D SRR Dixon TSE方案因边缘锐度下降和伪影存在,总体图像质量评分较低(P < 0.001)。
  • 所有方案的主观信噪比、对比分辨率、液体亮度和脂肪抑制均为良好至优秀。
  • 对1位读者,3D SRR方案的Dixon方法脂肪抑制显著优于标准2D TSE方案所用的频谱脂肪饱和(P < 0.05)。
  • 3D SRR Dixon TSE对膝关节结构的显示与标准方案非常相似,但软骨、肌腱和骨受重建和混叠伪影影响(P < 0.001)。

诊断标准与关键数值

  • 图像质量采用5分Likert量表评估。
  • 解剖可见性和诊断信心采用5分Likert量表评估正常和病理膝关节结构。
  • 记录膝关节内部紊乱的存在和位置,包括软骨缺损、半月板撕裂、韧带、肌腱和肌肉撕裂以及骨损伤。
  • 读者间和读者内一致性采用加权Fleiss-Cohen kappa(κ)统计量评估。
  • P值小于0.05认为有统计学意义。

临床应用与阅片要点

  • DL增强4倍加速2D TSE方案图像质量和诊断性能与标准2倍加速2D TSE相似,可使检查时间缩短44%。
  • 3D SRR Dixon TSE方案可行,诊断性能与标准2倍加速2D膝关节MRI相当,但软骨、肌腱和骨因重建和混叠伪影显示受影响。
  • 标准2D TSE评估肌腱的诊断信心显著高于3D SRR Dixon TSE(P < 0.01)。
  • 两种读者对所有方案和所有膝关节结构的诊断信心均高,但软骨和肌腱除外。
  • 使用任一方案评估膝关节内部紊乱的读者间和读者内一致性为中等至几乎完美(κ = 0.67-1.00)。

局限性

  • 重建和混叠伪影需进一步解决以确保软骨、肌腱和骨的更可靠显示和评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to evaluate the use of a multicontrast deep learning (DL)-reconstructed 4-fold accelerated 2-dimensional (2D) turbo spin echo (TSE) protocol and the feasibility of 3-dimensional (3D) superresolution reconstruction (SRR) of DL-enhanced 6-fold accelerated 2D Dixon TSE magnetic resonance imaging (MRI) for comprehensive knee joint assessment, by comparing image quality and diagnostic performance with a conventional 2-fold accelerated 2D TSE knee MRI protocol.

MATERIALS AND METHODS: This prospective, ethics-approved study included 19 symptomatic adult subjects who underwent knee MRI on a clinical 3 T scanner. Every subject was scanned with 3 DL-enhanced acquisition protocols in a single session: a clinical standard 2-fold in-plane parallel imaging (PI) accelerated 2D TSE-based protocol (5 sequences, 11 minutes 23 seconds) that served as a reference, a DL-reconstructed 4-fold accelerated 2D TSE protocol combining 2-fold PI and 2-fold simultaneous multislice acceleration (5 sequences, 6 minutes 24 seconds), and a 3D SRR protocol based on DL-enhanced 6-fold accelerated (ie, 3-fold PI and 2-fold simultaneous multislice) 2D Dixon TSE MRI (6 anisotropic 2D Dixon TSE acquisitions rotated around the phase-encoding axis, 6 minutes 24 seconds). This resulted in a total of 228 knee MRI scans comprising 21,204 images. Three readers evaluated all pseudonymized and randomized images in terms of image quality using a 5-point Likert scale. Two of the readers (musculoskeletal radiologists) additionally evaluated anatomical visibility and diagnostic confidence to assess normal and pathological knee structures with a 5-point Likert scale. They recorded the presence and location of internal knee derangements, including cartilage defects, meniscal tears, tears of ligaments, tendons and muscles, and bone injuries. The statistical analysis included nonparametric Friedman tests, and interreader and intrareader agreement assessment using the weighted Fleiss-Cohen kappa (κ) statistic. P values of less than 0.05 were considered statistically significant.

RESULTS: The evaluated DL-enhanced 4-fold accelerated 2D TSE protocol provided very similar image quality and anatomical visibility to the standard 2D TSE protocol, whereas the 3D SRR Dixon TSE protocol scored less in terms of overall image quality due to reduced edge sharpness and the presence of artifacts ( P < 0.001). Subjective signal-to-noise ratio, contrast resolution, fluid brightness, and fat suppression were good to excellent for all protocols. For 1 reader, the Dixon method of the 3D SRR protocol provided significantly better fat suppression than the spectral fat saturation applied in the standard 2D TSE protocol ( P < 0.05). The visualization of knee structures with 3D SRR Dixon TSE was very similar to the standard protocol, except for cartilage, tendons, and bone, which were affected by the presence of reconstruction and aliasing artifacts ( P < 0.001). The diagnostic confidence of both readers was high for all protocols and all knee structures, except for cartilage and tendons. The standard 2D TSE protocol showed a significantly higher diagnostic confidence for assessing tendons than 3D SRR Dixon TSE MRI ( P < 0.01). The interreader and intrareader agreement for the assessment of internal knee derangements using any of the 3 protocols was substantial to almost perfect (κ = 0.67-1.00). For cartilage, the interreader agreement was substantial for DL-enhanced accelerated 2D TSE (κ = 0.79) and almost perfect for standard 2D TSE (κ = 0.98) and 3D SRR Dixon TSE (κ = 0.87). For menisci, the interreader agreement was substantial for 3D SRR Dixon TSE (κ = 0.70-0.80) and substantial to almost perfect for standard 2D TSE (κ = 0.80-0.99) and DL-enhanced 2D TSE (κ = 0.87-1.00). Moreover, the total acquisition time was reduced by 44% when using the DL-enhanced accelerated 2D TSE or 3D SRR Dixon TSE protocol instead of the conventional 2D TSE protocol.

CONCLUSIONS: The presented DL-enhanced 4-fold accelerated 2D TSE protocol provides image quality and diagnostic performance similar to the standard 2D protocol. Moreover, the 3D SRR of DL-enhanced 6-fold accelerated 2D Dixon TSE MRI is feasible for multicontrast 3D knee MRI as its diagnostic performance is comparable to standard 2-fold accelerated 2D knee MRI. However, reconstruction and aliasing artifacts need to be further addressed to guarantee a more reliable visualization and assessment of cartilage, tendons, and bone. Both the 2D and 3D SRR DL-enhanced protocols enable a 44% faster examination compared with conventional 2-fold accelerated routine 2D TSE knee MRI and thus open new paths for more efficient clinical 2D and 3D knee MRI.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Smekens C, Beirinckx Q, Bosmans F, et al. Deep Learning-Enhanced Accelerated 2D TSE and 3D Superresolution Dixon TSE for Rapid Comprehensive Knee Joint Assessment. Invest Radiol. 2024. PMID: 39190787. doi:10.1097/RLI.0000000000001118

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。