S-Detect技术在不同平面与BI-RADS分类对乳腺病灶诊断价值的对比分析
该回顾性队列研究纳入2019年1月至2022年1月711例患者的756个乳腺病灶,评估S-Detect在径向与抗径向平面的诊断效能并与BI-RADS比较。88.4%病灶S-Detect跨平面结果一致,一致组中S-Detect准确率91.2%、AUC 0.916,均高于BI-RADS的84.9%和0.859;不一致组中S-Detect两个平面准确率和AUC均低于BI-RADS。病灶大小、边缘不清或成角、后方回声增强特征与跨平面诊断不一致显著相关。
要点
- S-Detect跨平面结果一致时诊断准确率91.2%、AUC 0.916,高于BI-RADS的84.9%和0.859。
- S-Detect跨平面结果不一致时,径向与抗径向平面准确率分别为47.7%和52.2%,均低于BI-RADS的69.3%。
- 病灶大小、边缘不清或成角、后方回声增强与S-Detect跨平面诊断不一致显著相关(p<0.05)。
核心结论
- S-Detect在跨平面结果一致时诊断精度优于BI-RADS,但在跨平面结果不一致时诊断效能受损,此时应谨慎参考S-Detect结果。
研究设计与样本
- 回顾性队列研究,纳入2019年1月至2022年1月711例患者的756个乳腺病灶,使用S-Detect评估径向与抗径向平面病灶,并以BI-RADS分类进行对比分析。
关键影像表现/征象
- 756个乳腺病灶中,668个(88.4%)S-Detect跨平面结果一致,88个(11.6%)不一致。
- 一致组中S-Detect诊断准确率91.2%、AUC 0.916,BI-RADS准确率84.9%、AUC 0.859(准确率p=0.045,AUC p=0.036)。
- 不一致组中S-Detect径向平面准确率47.7%、AUC 0.503,抗径向平面准确率52.2%、AUC 0.497;BI-RADS准确率69.3%、AUC 0.739(径向准确率p=0.014,抗径向准确率p=0.039;径向AUC p=0.042,抗径向AUC p<0.001)。
- S-Detect跨平面诊断不一致与病灶大小、边缘不清或成角、后方回声增强特征显著相关(p<0.05)。
临床应用与阅片要点
- 当S-Detect在径向与抗径向平面结果一致时,其诊断准确率和AUC高于BI-RADS,可参考其分类。
- 当S-Detect跨平面结果不一致时,其诊断准确率和AUC低于BI-RADS,应谨慎参考S-Detect结果。
- 对于病灶较大、边缘不清或成角、后方回声增强的病灶,S-Detect跨平面结果可能不一致,阅片时需注意。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: S-Detect, a deep learning-based Computer-Aided Detection system, is recognized as an important tool for diagnosing breast lesions using ultrasound imaging. However, it may exhibit inconsistent findings across multiple imaging planes. This study aims to evaluate the diagnostic performance of S-Detect in different planes and identify factors contributing to these inconsistencies.
MATERIALS AND METHODS: A retrospective cohort study was conducted on 711 patients with 756 breast lesions between January 2019 and January 2022. S-Detect was utilized to assess lesions in radial and anti-radial planes. BI-RADS classifications were employed for comparative analysis. The diagnostic performance was compared within each group, and p-values were computed for intergroup comparisons. Univariable and multivariable analyses were conducted to identify factors contributing to diagnostic inconsistency in S-Detect across planes.
RESULTS: Among 756 breast lesions, 668 (88.4%) exhibited consistent S-Detect outcomes across planes while 88 (11.6%) were inconsistent. In the consistent group, the diagnostic accuracy and area under the curve (AUC) of S-Detect were significantly higher than those of BI-RADS (accuracy: 91.2% vs. 84.9%, p = 0.045; AUC: 0.916 vs. 0.859, p = 0.036). In the inconsistent group, the diagnostic accuracy and AUC of S-Detect in radial and anti-radial planes were lower than those of BI-RADS (accuracy: 47.7% for radial, 52.2% for anti-radial vs. 69.3% for BI-RADS, p = 0.014, p-anti = 0.039; AUC: 0.503 for radial, 0.497 for anti-radial vs. 0.739 for BI-RADS, p = 0.042, p-anti <0.001). Diagnostic inconsistency in S-Detect across planes was significantly associated with lesion size, indistinct or angular margins, and enhancement posterior acoustic features (p < 0.05).
CONCLUSION: S-Detect has outperformed BI-RADS in diagnostic precision under conditions of inter-planar concordance. However, its diagnostic efficacy is compromised in scenarios of inter-planar discordance. Under these circumstances, the results of S-Detect should be carefully referenced.
引用
- Zhang P, Zhang M, Lu M, et al. Comparative Analysis of the Diagnostic Value of S-Detect Technology in Different Planes Versus the BI-RADS Classification for Breast Lesions. Acad Radiol. 2024. PMID: 39138111. doi:10.1016/j.acra.2024.08.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。