PCD-CT在腹部成像中相比EID-CT个体化kV适配对比剂注射方案可减少对比剂用量
本研究比较了门静脉期腹部CT中PCD-CT与EID-CT的对比剂用量及图像质量,PCD-CT采用总体重适配方案,EID-CT采用总体重与kV适配方案。基于91例EID-CT和102例PCD-CT扫描,PCD-CT的总碘负荷降低20.1%,且SNR和CNR显著更高,所有扫描主观图像质量均满足诊断要求。
要点
- PCD-CT较EID-CT总碘负荷降低20.1%,同时SNR和CNR显著更高。
- PCD-CT采用总体重适配对比剂方案,EID-CT采用总体重与kV适配个体化方案。
- 所有扫描主观图像质量均满足诊断要求。
核心结论
- PCD-CT使用临床个体化扫描与对比剂注射方案,可在减少对比剂用量的同时提供较EID-CT更高的SNR和CNR。
研究设计与样本
- 研究比较了EID-CT(2022年11月至2024年3月)与PCD-CT(2023年9月至2023年12月)的连续增强腹部门静脉期CT扫描;EID-CT采用基于总体重和kV的个体化采集与对比剂注射方案,PCD-CT采用总体重适配对比剂注射方案;共纳入91例EID-CT扫描和102例PCD-CT扫描。
诊断标准与关键数值
- PCD-CT较EID-CT总碘负荷降低20.1%;PCD-CT的SNR为9.9 ± 1.7,EID-CT为9.1 ± 1.8(p < 0.001);PCD-CT的CNR为5.1 ± 1.7,EID-CT为4.3 ± 1.3(p < 0.001);主观图像质量采用5点Likert量表由2名专家读者基于诊断信心评估,所有扫描均具有足够诊断图像质量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study aims to demonstrate reduced iodine contrast media (CM) in routine abdominal CT scans in portal venous phase (PVP) using a photon-counting detector CT (PCD-CT) compared to total body weight (TBW) and kV-adapted CM injection protocols on a state-of-the-art energy-integrating detector CT (EID-CT) while maintaining sufficient image quality (IQ).
MATERIALS AND METHODS: Consecutive contrast-enhanced abdominal PVP CT scans from an EID-CT (Nov 2022-March 2024) and a PCD-CT (Sep 2023-Dec 2023) were compared. CM parameters (total iodine load (TIL), iodine delivery rate (IDR) and dosing factor (DF)) were reported. An individualized acquisition and CM injection protocol based on TBW and kV was applied for the EID-CT and a TBW adapted CM injection protocol was used for the PCD-CT. Objective IQ was evaluated with mean attenuation (Hounsfield Units, HU), signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR)). Subjective IQ was assessed via a 5-point Likert scale by 2 expert readers based on diagnostic confidence.
RESULTS: Based on 91 EID-CT scans and 102 PCD-CT scans a TIL reduction of 20.1 % was observed for PCD-CT. PCD-CT demonstrated significantly higher SNR (9.9 ± 1.7 vs. 9.1 ± 1.8, p < 0.001) and CNR (5.1 ± 1.7 vs. 4.3 ± 1.3, p < 0.001) compared to EID-CT. Subjective IQ assessment showed that all scans had sufficient diagnostic IQ.
CONCLUSIONS: PCD-CT allows for CM reduction while providing higher SNR and CNR compared to EID-CT, using clinical individualized scan and CM injection protocols.
引用
- Hoeijmakers EJI, Stammen L, Wildberger JE, et al. PCD-CT enables contrast media reduction in abdominal imaging compared to an individualized kV-adapted contrast media injection protocol on EID-CT. Eur J Radiol. 2024. PMID: 39133989. doi:10.1016/j.ejrad.2024.111680