志愿者及代谢功能障碍相关脂肪性肝病患者剪切波黏弹性成像的Visit间可重复性

Between-Visit Reproducibility of Shear Wave Viscoelastography in Volunteers and Patients With Metabolic Dysfunction-Associated Steatotic Liver Disease.
PubMed 文献 · J Ultrasound Med · 2024-08-08

前瞻性横断面研究纳入10名志愿者和20例经活检证实的MASLD或MASH患者,间隔至少2周行两次肝脏超声弹性成像,评估6项剪切波黏弹性参数的可重复性。结果显示检查技师对参数无影响(P > .05),各参数ICC为0.78~0.96,CV为10.1%~32.2%,RDC为27.7%~89.2%,Bland-Altman分析显示两次Visit间一致性高。结论认为同一超声机器上间隔2周两次Visit的黏弹性参数测量具有高可重复性和一致性。

要点

  • 间隔至少2周两次Visit的剪切波黏弹性参数ICC为0.78~0.96,SWA的ICC最高(0.96)。
  • 各参数CV为10.1%~32.2%,SWA的CV最低(10.1%),黏度的CV最高(32.2%)。
  • 检查技师对黏弹性参数无影响(P > .05),线性混合模型已将其作为混杂因素纳入。

核心结论

  • 同一超声机器上间隔2周两次Visit的超声剪切波黏弹性参数测量具有高可重复性和一致性。

研究设计与样本

  • 前瞻性横断面研究,经机构审查委员会批准,纳入10名志愿者和20例经活检证实的MASLD或MASH患者,间隔至少2周行两次肝脏超声弹性成像,使用研究型超声扫描仪采集射频数据计算6项参数。

诊断标准与关键数值

  • SWS的ICC为0.89(95%CI:0.79-0.95),CV为11.9%,RDC为33.0%,Bland-Altman平均偏差为-0.05 m/s(95%一致性界限:-0.45,0.35)。
  • E的ICC为0.81(95%CI:0.79-0.95),CV为23.3%,RDC为64.5%,Bland-Altman平均偏差为-0.61 kPa(95%一致性界限:-5.33,4.10)。
  • G的ICC为0.90(95%CI:0.80-0.95),CV为24.2%,RDC为66.9%,Bland-Altman平均偏差为-0.25 kPa(95%一致性界限:-2.06,1.56)。
  • SWA的ICC为0.96(95%CI:0.93-0.98),CV为10.1%,RDC为27.7%,Bland-Altman平均偏差为-0.01 Np/m/Hz(95%一致性界限:-0.27,0.26)。
  • SWD的ICC为0.78(95%CI:0.60-0.89),CV为29.0%,RDC为80.3%,Bland-Altman平均偏差为-0.09 m/s/kHz(95%一致性界限:-7.09,6.91);黏度的ICC为0.90(95%CI:0.80-0.95),CV为32.2%,RDC为89.2%,Bland-Altman平均偏差为-0.33 Pa/s(95%一致性界限:-2.60,1.94)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To assess the reproducibility of six ultrasound (US)-determined shear wave (SW) viscoelastography parameters for assessment of mechanical properties of the liver in volunteers and patients with biopsy-proven metabolic dysfunction-associated steatotic liver disease (MASLD) or metabolic dysfunction-associated steatohepatitis (MASH).

METHODS: This prospective, cross-sectional, institutional review board-approved study included 10 volunteers and 20 patients with MASLD or MASH who underwent liver US elastography twice, at least 2 weeks apart. SW speed (SWS), Young's modulus (E), shear modulus (G), SW attenuation (SWA), SW dispersion (SWD), and viscosity were computed from radiofrequency data recorded on a research US scanner. Linear mixed models were used to consider the sonographer on duty as a confounder. The reproducibility of measurements was assessed by intraclass correlation coefficient (ICC), coefficient of variation (CV), reproducibility coefficient (RDC), and Bland-Altman analyses.

RESULTS: The sonographer performing the exam had no impact on viscoelastic parameters (P > .05). ICCs of SWS, E, G, SWA, SWD, and viscosity were, respectively, 0.89 (95% confidence intervals [CI]: 0.79-0.95), 0.81 (95% CI: 0.79-0.95), 0.90 (95% CI: 0.80-0.95), 0.96 (95% CI: 0.93-0.98), 0.78 (95% CI: 0.60-0.89), and 0.90 (95% CI: 0.80-0.95); CVs were 11.9, 23.3, 24.2, 10.1, 29.0, and 32.2%; RDCs were 33.0, 64.5, 66.9, 27.7, 80.3, and 89.2%, and Bland-Altman mean biases and 95% limits of agreement were -0.05 (-0.45, 0.35) m/s, -0.61 (-5.33, 4.10) kPa, -0.25 (-2.06, 1.56) kPa, -0.01 (-0.27, 0.26) Np/m/Hz, -0.09 (-7.09, 6.91) m/s/kHz, and -0.33 (-2.60, 1.94) Pa/s, between the two visits.

CONCLUSION: US-determined viscoelastography parameters can be measured with high reproducibility and consistency between two visits 2 weeks apart on the same ultrasound machine.

来源:J Ultrasound Med,查看原文。版权归原作者所有。