文献速递腹部外周血管CT

低剂量超高分辨率光子计数探测器CT用于内脏动脉CT血管成像的初步研究

Low-dose Ultra-high-resolution Photon-Counting Detector CT for Visceral Artery CT Angiography: A Preliminary Study.
PubMed 文献 · Acad Radiol · 2024-08-06

前瞻性纳入57例患者分别接受全剂量与低剂量方案,采用光子计数探测器CT超高分辨率模式行腹部CT血管成像,比较0.2 mm与1 mm重建图像质量。低剂量扫描图像质量在超重患者中未受显著影响,0.2 mm重建可改善小动脉显示,低剂量超高分辨率模式足以用于内脏动脉临床评估。

要点

  • 低剂量与全剂量方案在相同层厚重建下多数内脏动脉信噪比与对比噪声比无显著差异,仅0.2 mm重建的肝总动脉、脾动脉及双侧肾动脉信噪比全剂量组较高。
  • 0.2 mm重建图像主观质量优于1 mm重建,尤其在小动脉的容积渲染图像上。
  • 低剂量扫描在超重患者中未显著影响图像质量。

核心结论

  • 光子计数探测器CT低剂量超高分辨率扫描模式足以用于内脏动脉的临床评估。

研究设计与样本

  • 前瞻性研究,全剂量组与低剂量组各纳入57例患者,2023年4月至2023年9月期间使用光子计数探测器CT系统(NAEOTOM Alpha)超高分辨率模式行腹部CT扫描。

关键影像表现/征象

  • 0.2 mm重建组的信噪比和对比噪声比低于1 mm重建组(所有动脉及噪声p < 0.001)。
  • 定性评估中0.2 mm重建组图像质量高于1 mm重建组,尤其在小动脉的容积渲染图像上(p < 0.001)。
  • 相同层厚下,除0.2 mm重建组肝总动脉、脾动脉及双侧肾动脉的信噪比外,全剂量组信噪比和对比噪声比未显著高于低剂量组。
  • 相同重建组内正常体重与超重患者比较,低剂量扫描未显著影响超重患者图像质量。
  • 九条内脏动脉可见度评分在相同层厚重建下全剂量与低剂量间无显著差异,除外肠系膜上动脉(轴位0.2 mm和1 mm重建p = 0.002和0.007;冠状位0.2 mm和1 mm重建p = 0.002和0.007)和胃左动脉(容积渲染0.2 mm和1 mm重建p = 0.002和p < 0.001)。

临床应用与阅片要点

  • 采用0.2 mm层厚重建可增强动脉显示,尤其有利于小血管的评估。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To validate the image quality of low-dose ultra-high-resolution (UHR) scanning mode of photon-counting detector CT (PCD-CT) for visceral artery computed tomography angiography (CTA).

MATERIAL AND METHODS: We prospectively enrolled 57 patients each in the full dose (FD) and low-dose (LD) protocols, respectively, to undergo abdominal CT scans using the UHR mode on a PCD-CT system (NAEOTOM Alpha), between April 2023 and September 2023. Both the FD data and LD data were then reconstructed into two series of images: (a) 0.2 mm slice thickness, reconstruction kernel Bv48, quantum iterative reconstruction (QIR) 4; (b)1 mm slice thickness, Bv40, QIR 3. The signal-to-noise ratios (SNRs) and contrast-to-noise ratios (CNRs) of seven arteries were objectively measured. The image noise, vessel sharpness, overall quality, and visibility of nine arteries were subjectively assessed by three radiologists.

RESULTS: The SNRs and CNRs of 0.2 mm reconstruction set was inferior to that of 1 mm reconstruction set (p < 0.001 for all the arteries and noise), however, the image quality of 0.2 mm reconstruction set was higher than that of 1 mm reconstruction set in qualitative evaluation especially for tiny arteries in Volume-rendered (VR) image (p < 0.001). The SNRs and CNRs were not significantly higher for FD group than LD group on the same slice thickness except for SNRs of common hepatic artery, splenic artery and bilateral renal arteries in 0.2 mm reconstruction set. In the comparison on image quality between normal weight and overweight patients within the same reconstruction set, the results showed that low-dose scan did not significantly impact the image quality in overweight patients. The ratings of visibility of nine visceral arteries were not significantly different among FD and LD at the same thickness reconstruction set except for superior mesenteric artery (p = 0.002 and 0.007 for 0.2 mm and 1 mm reconstruction set in axial image; p = 0.002 and 0.007 for 0.2 mm and 1 mm reconstruction set in coronal image, respectively) and left gastric artery (p = 0.002 and p < 0.001 for 0.2 mm and 1 mm reconstruction set in VR image, respectively).

CONCLUSION: The low-dose UHR scanning mode of PCD-CT has proven to be adequate for the clinical evaluation of visceral arteries. Utilizing a reconstruction with a slice thickness of 0.2 mm could enhance arterial depiction, particularly for small vessels.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Wang L, Hu Y, Zhong J, et al. Low-dose Ultra-high-resolution Photon-Counting Detector CT for Visceral Artery CT Angiography: A Preliminary Study. Acad Radiol. 2024. PMID: 39112295. doi:10.1016/j.acra.2024.06.044

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。