血管衰减、虚拟单能级和重建卷积核对冠状动脉周围脂肪组织衰减的影响:离体猪心能量积分探测器CT与光子计数探测器CT研究

Effect of Vessel Attenuation, VMI Level, and Reconstruction Kernel on Pericoronary Adipose Tissue Attenuation for EID CT and PCD CT: An Ex Vivo Porcine Heart Study.
PubMed 文献 · AJR Am J Roentgenol · 2024-08-21

本研究利用离体猪心模型,探究血管衰减、虚拟单能级(VMI)及重建卷积核对冠状动脉周围脂肪组织(PCAT)衰减和脂肪衰减指数(FAI)的影响。结果显示,PCAT衰减和FAI受血管腔衰减、VMI水平及重建卷积核影响;在能量积分探测器CT和光子计数探测器CT高VMI水平下,仅少数冠状动脉周围体素参与FAI测量。

要点

  • PCAT衰减和FAI受血管腔衰减、VMI水平及重建卷积核影响,测量时需标准化参数。
  • 在EID CT和PCD CT高VMI水平下,仅少数冠状动脉周围体素参与FAI计算。
  • 使用更锐利的卷积核时,PCAT衰减和FAI与血管衰减的关联逐渐减弱。

核心结论

  • PCAT衰减和FAI受血管腔衰减、VMI水平及重建卷积核影响;对于多能EID CT和高VMI水平的PCD CT,仅少数冠状动脉周围体素参与FAI测量。

研究设计与样本

  • 离体猪心模型研究;对猪心右冠状动脉注射生理盐水或不同稀释度对比剂,使血管衰减范围达0-1000 HU;随后依次行EID CT(120 kVp)和PCD CT(140 kVp)扫描,CTDIvol恒定为10 mGy;EID CT多能图像采用Qr40卷积核重建;PCD CT虚拟单能图像(40-80 keV,间隔10 keV)分别采用Qr40、Bv40和Bv56卷积核重建。

关键影像表现/征象

  • 右冠状动脉衰减与邻近PCAT衰减之间存在相关性(r=0.3-1.0)。
  • 血管衰减与FAI之间存在相关性(r=-0.9至1.0)。
  • 使用逐渐锐利的卷积核时,上述关联逐渐减弱。
  • EID CT血管衰减增加及PCD CT VMI水平增加时,FAI直方图峰值衰减值右移。
  • PCD CT使用更锐利卷积核时,FAI直方图峰值衰减值左移,且落入FAI计算阈值范围内的体素百分比增大。

诊断标准与关键数值

  • FAI计算中符合阈值范围并入组的直方图体素百分比:EID CT为9-38%,PCD CT在70-80 keV VMI水平下为6-39%。

临床应用与阅片要点

  • PCAT衰减和FAI测量需考虑血管腔衰减、VMI水平及重建卷积核对结果的影响。
  • 在EID CT和PCD CT高VMI水平下,仅少数冠状动脉周围体素参与FAI计算。

局限性

  • 本研究为离体猪心模型研究。
  • 原文结论指出这些发现可能有助于标准化PCAT衰减和FAI测量的采集与重建参数。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Pericoronary adipose tissue (PCAT) attenuation and the fat attenuation index (FAI) may serve as markers of inflammation and the risk of adverse cardiac events. However, standardization of relevant CT acquisition and reconstruction parameters is lacking. OBJECTIVE. The purpose of this study was to investigate the influence of vessel attenuation, the virtual monoenergetic image (VMI) level, and the reconstruction kernel on PCAT attenuation and FAI by use of energy-integrating detector (EID) and photon-counting detector (PCD) CT systems in an ex vivo porcine heart model. METHODS. The right coronary artery (RCA) of a porcine heart was injected with saline or varying contrast medium dilutions to achieve vessel attenuation ranging from 0 to 1000 HU. After each injection, the heart was sequentially scanned by EID CT at 120 kVp and PCD CT at 140 kVp at a constant CTDIvol of 10 mGy. For EID CT, polychromatic images were reconstructed with a Qr40 kernel. For PCD CT, VMIs (obtained at 40-80 keV in 10-keV increments) were reconstructed with Qr40, Bv40, and Bv56 kernels. ROIs were placed to measure RCA and PCAT attenuation. FAI was determined using software; histogram analysis was performed to assess voxel attenuation in the volumes of interest for FAI calculation. RESULTS. Correlations were observed between attenuation in the RCA and attenuation in the adjacent PCAT (r = 0.3-1.0) and between vessel attenuation and the FAI (r = -0.9 to 1.0). For PCAT attenuation and the FAI, these associations became progressively weaker when progressively sharper kernels were used. For increasing vessel attenuation on EID CT and for increasing VMI level on PCD CT, FAI histograms showed right shifts in peak attenuation values; the percentage of histogram voxels that met the threshold range for inclusion in FAI calculation was 9-38% for EID CT and 6-39% for PCD CT at VMI levels of 70-80 keV. For PCD CT, use of sharper kernels was associated with left shifts in peak attenuation values and greater percentages of voxels within the threshold range for inclusion in FAI calculation. CONCLUSION. PCAT attenuation and the FAI are influenced by vessel lumen attenuation, the VMI level, and the reconstruction kernel. A minority of pericoronary voxels contribute to FAI measurements for polychromatic EID CT and for PCD CT at high VMI levels. CLINICAL IMPACT. These findings may help standardize acquisition and reconstruction parameters for PCAT attenuation and FAI measurements.