双轴蝴蝶扫描剂量缩减模式下神经血管内治疗脑软组织图像质量评估

Image Quality Evaluation for Brain Soft Tissue in Neuroendovascular Treatment by Dose-Reduction Mode of Dual-Axis "Butterfly" Scan.
PubMed 文献 · AJNR Am J Neuroradiol · 2025-02-03

研究前瞻性纳入40例神经血管内治疗患者,比较常规锥形束CT与蝴蝶扫描剂量缩减模式(中剂量45 mGy、低剂量30 mGy)的图像质量。中剂量蝴蝶扫描减少伪影、增强对比度并改善皮髓质交界分辨(枕叶除外);低剂量组伪影明显减少且对比度改善(枕叶除外),但皮髓质交界分辨未见显著改善。

要点

  • 中剂量蝴蝶扫描(45 mGy)可减少伪影、增强对比度并改善皮髓质交界分辨,枕叶除外。
  • 低剂量蝴蝶扫描(30 mGy)伪影明显减少且对比度改善,但皮髓质交界分辨未见显著改善。
  • 枕叶区域因骨干扰易出现对比度下降和皮髓质交界分辨受损。

核心结论

  • 即使剂量缩减,蝴蝶扫描的特殊轨迹仍可减少伪影、改善对比度并增强皮髓质交界分辨。
  • 剂量缩减的影响在枕叶区域更明显,骨干扰导致对比度下降和皮髓质交界分辨受损。

研究设计与样本

  • 前瞻性研究纳入计划接受神经血管内治疗的患者,共40例,中剂量组20例、低剂量组20例。
  • 患者交替分配接受中剂量蝴蝶扫描(原始剂量的70%,45 mGy)或低剂量蝴蝶扫描(原始剂量的50%,30 mGy)。
  • 采用5分制评分系统评估伪影、对比度及皮髓质交界分辨,并与常规锥形束CT比较。

关键影像表现/征象

  • 与常规锥形束CT相比,中剂量蝴蝶扫描组伪影减少、对比度增强、皮髓质交界分辨改善,枕叶除外。
  • 与常规锥形束CT相比,低剂量蝴蝶扫描组伪影明显减少、对比度改善,枕叶除外,但皮髓质交界分辨未见显著改善。

诊断标准与关键数值

  • 图像质量采用5分制评分系统评估伪影、对比度和皮髓质交界分辨。
  • 中剂量蝴蝶扫描为原始剂量的70%(45 mGy),低剂量蝴蝶扫描为原始剂量的50%(30 mGy)。

临床应用与阅片要点

  • 中剂量蝴蝶扫描(45 mGy)在减少伪影、增强对比度和改善皮髓质交界分辨方面优于常规锥形束CT,枕叶除外。
  • 低剂量蝴蝶扫描(30 mGy)可明显减少伪影并改善对比度,但皮髓质交界分辨未见显著改善。
  • 枕叶区域因骨干扰易出现对比度下降和皮髓质交界分辨受损。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Flat panel conebeam CT (CBCT) is essential for detecting hemorrhagic complications during neuroendovascular treatments. Despite its superior image quality and trajectory over conventional CBCT (circular scan), the dual-axis butterfly scan has a slightly higher radiation dose relative to conventional CBCT. This study evaluates the image quality in dose-reduction mode to uncover the appropriate radiation dose for the butterfly scan.

MATERIALS AND METHODS: We prospectively included patients who were scheduled for neuroendovascular treatment and underwent conventional CBCT and the dose-reduction mode of the butterfly scan. Two reduced radiation dose modes were used for the butterfly scan: medium-dose butterfly scan (70% of the original dose, 45 mGy) or low-dose butterfly scan (50% of the original dose, 30 mGy). The enrolled patients were assigned alternately to undergo either the medium- or low-dose butterfly scan. We evaluated and compared artifacts, contrast, and discrimination of the corticomedullary junction between conventional CBCT and one of the dose-reduction modes of the butterfly scan, with a 5-point scale scoring system.

RESULTS: Twenty patients were enrolled in each of the medium- and low-dose groups, totaling 40 patients. Compared with conventional CBCT, the medium-dose butterfly group exhibited reduced artifacts, enhanced contrast, and corticomedullary junction discrimination (except in the occipital lobe). While the low-dose butterfly group exhibited markedly reduced artifacts and improved contrast (except in the occipital lobe), a significant improvement in corticomedullary junction discrimination was not observed.

CONCLUSIONS: Even with dose reduction, the specialized trajectory of the butterfly scan enables artifact reduction, contrast improvement, and enhanced corticomedullary junction discrimination. However, the impact of the reduced dose was more noticeable, particularly in the occipital region where susceptibility to bone interference resulted in decreased contrast and compromised corticomedullary junction discrimination.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。