场循环磁共振成像用于识别0.2 T以下的微小缺血性卒中
前瞻性研究纳入9例亚急性缺血性卒中患者,在发病后1-6天于0.2 mT至0.2 T的四个至六个演化场强下行场循环成像(FCI),发现梗死区在0.2 T以下表现为T1高信号,与基线影像所见梗死区对应,两名独立阅片者间一致性为86%(Cohen κ=0.69);结论认为全脑FCI可在低至0.2 mT的场强下通过T1弛豫机制识别亚急性缺血性卒中。
要点
- 亚急性缺血性卒中在0.2 T以下FCI上表现为T1高信号,与基线影像梗死区对应。
- 梗死区与对侧正常脑组织的T1对比度随场强降低(0.2 mT至0.2 T)而增大。
- 两名独立阅片者视觉确认梗死区的阅片者间一致性为86%(Cohen κ=0.69)。
核心结论
- 全脑FCI可在低至0.2 mT的场强下通过T1弛豫机制识别亚急性缺血性卒中。
研究设计与样本
- 前瞻性研究,连续筛查单中心卒中单元2018年2月至2020年3月及2021年4月至12月收治的成年患者,9例确诊缺血性卒中患者于发病后1-6天完成FCI检查。
关键影像表现/征象
- FCI扫描在0.2 T以下显示与基线影像所见梗死区对应的T1高信号区,两名对临床图像设盲的独立阅片者视觉确认的阅片者间一致性为86%(Cohen κ=0.69)。
诊断标准与关键数值
- FCI扫描在0.2 mT至0.2 T之间的四个至六个演化场强下采集,演化时间从5至546毫秒共五个,生成T1图;梗死区与对侧组织对比度随场强降低而增大(r[24]=-0.68;P<.001);梗死区与未受累组织的T1离散斜率中位数分别为0.23(IQR,0.18-0.37)与0.35(IQR,0.27-0.43),P=.03。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 本研究为单中心研究,样本量小(9例),且所有参与者均为男性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Field-cycling imaging (FCI) is a new technology developed at the University of Aberdeen that measures change in T1 relaxation time constant of tissues over a range of low magnetic field strengths (0.2-200 mT) by rapidly switching between different fields during the pulse sequence. This provides new sources of contrast, including some invisible to clinical MRI scanners, and may be a useful alternative imaging modality for stroke. Purpose To test whether a prototype whole-body FCI scanner can be used to identify infarct regions in patients with subacute ischemic stroke. Materials and Methods This prospective study screened consecutive adult patients admitted to a single center stroke unit from February 2018 to March 2020 and April to December 2021. Included participants with confirmed ischemic stroke underwent FCI 1-6 days after ictus. FCI scans were obtained at four to six evolution fields between 0.2 mT and 0.2 T, with five evolution times from 5 to 546 msec. T1 maps were generated. The Wilcoxon signed-rank test was used to compare infarct region and contralateral unaffected brain, and Spearman rank correlation was used to examine associations between infarct to contralateral tissue contrast ratio and field strengths. Two independent readers blinded to clinical images rated the FCI scans. Results Nine participants (mean age, 62 years ± 16 [SD]; all male) successfully completed FCI. FCI scans below 0.2 T exhibited hyperintense T1 regions corresponding to the infarct region identified at baseline imaging, visually confirmed with 86% interrater agreement (Cohen κ = 0.69). Infarct to contralateral tissue contrast ratio increased as magnetic field decreased between 0.2 mT and 0.2 T (r[24] = -0.68; P < .001). T1 dispersion slopes differed between infarct and unaffected tissues (median, 0.23 [IQR, 0.18-0.37] vs 0.35 [IQR, 0.27-0.43]; P = .03). Conclusion Whole-brain FCI can be used to identify subacute ischemic stroke by T1 relaxation mechanisms at field strengths as low as 0.2 mT. Research Registry no. 1813 Published under a CC BY 4.0 license. Supplemental material is available for this article.
引用
- Mallikourti V, Ross PJ, Maier O, et al. Field-Cycling MRI for Identifying Minor Ischemic Stroke Below 0.2 T. Radiology. 2024. PMID: 39189899. doi:10.1148/radiol.232972
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。