两全其美:光子计数探测器CT自由呼吸超高清肺动脉造影诊断急性肺栓塞
本研究前瞻性比较光子计数探测器(PCD)CT与能量积分探测器(EID)CT在自由呼吸超高清肺动脉造影中的图像质量和辐射剂量。结果显示PCD组图像质量更优、无不可诊断扫描、辐射剂量更低(1.33±0.47 vs 1.80±0.82 mSv),且能重建碘图提供功能信息。
要点
- PCD-CT在叶级肺动脉的CNR和SNR更高,图像质量优良率达96.1%,无不可诊断扫描。
- PCD-CT辐射剂量更低,有效剂量1.33±0.47 mSv,EID-CT为1.80±0.82 mSv。
- 仅PCD-CT结合超高清可重建碘图,提供额外功能信息。
核心结论
- 自由呼吸超高清PCD-CT肺动脉造影可获得更优图像质量、显著降低辐射剂量并提供完整光谱信息。
研究设计与样本
- 前瞻性研究,51例PCD-CT肺动脉造影与51例第三代双源EID-CT肺动脉造影匹配比较;均采用超高清自由呼吸方案(40 mL对比剂,螺距3.2),PCD为140 kV,EID为70-100 kV。
关键影像表现/征象
- PCD组肺动脉CT衰减值更高(所有P<0.05)。
- PCD组叶级肺动脉CNR和SNR更高(P<0.05),肺动脉干内无差异(P>0.05)。
- PCD-CT图像质量优良率96.1%,EID-CT为50.9%;PCD-CT无不可诊断扫描,EID-CT有3例(5.9%)不可诊断。
诊断标准与关键数值
- PCD-CT有效剂量1.33±0.47 mSv,EID-CT为1.80±0.82 mSv(P<0.05)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
RATIONALE AND OBJECTIVES: To assess image quality and radiation dose of ultra-high-pitch CT pulmonary angiography (CTPA) with free-breathing technique for diagnosis of pulmonary embolism using a photon-counting detector (PCD) CT compared to matched energy-integrating detector (EID)-based single-energy CTPA.
MATERIALS AND METHODS: Fifty-one PCD-CTPAs were prospectively compared to 51 CTPAs on a third-generation dual-source EID-CT. CTPAs were acquired with an ultra-high-pitch protocol with free-breathing technique (40 mL contrast medium, pitch 3.2) at 140 kV (PCD) and 70-100 kV (EID). Iodine maps were reconstructed from spectral PCD-CTPAs. Image quality of CTPAs and iodine maps was assessed independently by three radiologists. Additionally, CT attenuation numbers within pulmonary arteries as well as signal-to-noise and contrast-to-noise ratios (SNR, CNR) were compared. Administered radiation dose was compared.
RESULTS: CT attenuation was higher in the PCD-group (all P < 0.05). CNR and SNR were higher in lobar pulmonary arteries in PCD-CTPAs (P < 0.05), whereas no difference was ascertained within the pulmonary trunk (P > 0.05). Image quality of PCD-CTPA was rated best by all readers (excellent/good image quality in 96.1% of PCD-CTPAs vs. 50.9% of EID-CTPAs). PCD-CT produced no non-diagnostic scans vs. three non-diagnostic (5.9%) EID-CTPAs. Radiation dose was lower with PCD-CT than with EID-CT (effective dose 1.33 ± 0.47 vs. 1.80 ± 0.82 mSv; all P < 0.05).
CONCLUSION: Ultra-high-pitch CTPA with free-breathing technique with PCD-CT allows for superior image quality with significantly reduced radiation dose and full spectral information. With the ultra-high pitch, only PCD-CTPA enables reconstruction of iodine maps containing additional functional information.
引用
- Pannenbecker P, Heidenreich JF, Huflage H, et al. The Best of Both Worlds: Ultra-high-pitch Pulmonary Angiography with Free-Breathing Technique by Means of Photon-Counting Detector CT for Diagnosis of Acute Pulmonary Embolism. Acad Radiol. 2024. PMID: 38969575. doi:10.1016/j.acra.2024.06.028
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。