深度学习去噪使颈部CT辐射暴露降至常规重建极限以下
回顾性纳入临床怀疑颈部肿瘤患者,比较100%、模拟50%和25%辐射剂量下原始迭代重建与AI去噪后处理的图像质量。去噪在各级剂量下均显著降低噪声、提高CNR,主观评分在100%和50%剂量下显著更优,且原始100%与去噪25%主观评分无显著差异。
要点
- 去噪后处理在100%、50%、25%剂量下均显著降低图像噪声并提高CNR(p < 0.001)。
- 主观图像质量、诊断信心、锐利度和对比度在100%和50%剂量下去噪均显著优于原始重建(p < 0.001)。
- 原始100%剂量与去噪25%剂量的主观评分无显著差异(p = 0.906)。
核心结论
- 所研究的去噪算法可使颈部CT在辐射剂量降至常规25%时获得诊断质量图像,显著减少患者暴露。
研究设计与样本
- 回顾性研究,纳入2023年9月1日至2023年12月1日临床怀疑颈部肿瘤患者,同一单源扫描仪扫描,重建为原始迭代重建100%、模拟50%和25%剂量,每种数据经去噪软件后处理,三名不同经验放射科医生在随机盲法强迫选择设置下主观评分。
诊断标准与关键数值
- 在100%、50%、25%各辐射剂量水平,去噪较原始重建图像噪声显著更低、CNR显著更高(p < 0.001);主观图像质量、诊断信心、锐利度和对比度在100%和50%剂量下去噪显著高于原始(p < 0.001);原始100%与去噪25%主观评分无显著差异(p = 0.906)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Neck computed tomography (NCT) is essential for diagnosing suspected neck tumors and abscesses, but radiation exposure can be an issue. In conventional reconstruction techniques, limiting radiation dose comes at the cost of diminished diagnostic accuracy. Therefore, this study aimed to evaluate the effects of an AI-based denoising post-processing software solution in low-dose neck computer tomography.
MATERIALS AND METHODS: From 01 September 2023 to 01 December 2023, we retrospectively included patients with clinically suspected neck tumors from the same single-source scanner. The scans were reconstructed using Advanced Modeled Iterative Reconstruction (Original) at 100% and simulated 50% and 25% radiation doses. Each dataset was post-processed using a novel denoising software solution (Denoising). Three radiologists with varying experience levels subjectively rated image quality, diagnostic confidence, sharpness, and contrast for all pairwise combinations of radiation dose and reconstruction mode in a randomized, blinded forced-choice setup. Objective image quality was assessed using ROI measurements of mean CT numbers, noise, and a contrast-to-noise ratio (CNR). An adequately corrected mixed-effects analysis was used to compare objective and subjective image quality.
RESULTS: At each radiation dose level, pairwise comparisons showed significantly lower image noise and higher CNR for Denoising than for Original (p < 0.001). In subjective analysis, image quality, diagnostic confidence, sharpness, and contrast were significantly higher for Denoising than for Original at 100 and 50 % (p < 0.001). However, there were no significant differences in the subjective ratings between Original 100 % and Denoising 25 % (p = 0.906).
CONCLUSIONS: The investigated denoising algorithm enables diagnostic-quality neck CT images with radiation doses reduced to 25% of conventional levels, significantly minimizing patient exposure.