文献速递乳腺CT乳腺X线

专用乳腺CT与乳腺X线摄影对不同乳腺密度乳腺癌检出能力的比较

Detectability of Breast Cancer in Dedicated Breast CT Compared With Mammography Dependent on Breast Density.
PubMed 文献 · Invest Radiol · 2024-07-05

该回顾性研究纳入90例女性105个活检证实恶性乳腺病灶,由2名独立阅片者评估平扫与增强螺旋乳腺CT及乳腺X线摄影的检出率、BI-RADS描述及诊断信心。增强螺旋乳腺CT检出率99.1%且不受乳腺密度影响,平扫螺旋乳腺CT检出率低于乳腺X线摄影并随密度升高而下降。

要点

  • 增强螺旋乳腺CT检出率99.1%,各乳腺密度下均高(A/B/C/D为100%/100%/100%/87.5%)。
  • 平扫螺旋乳腺CT检出率66.7%和61.9%,低于乳腺X线摄影的83.8%和80.0%。
  • 乳腺X线摄影检出率随密度升高而下降(A=100%,D=50.0%和71.4%)。

核心结论

  • 增强螺旋乳腺CT检出率和诊断信心很高且不受乳腺密度影响,平扫螺旋乳腺CT检出率低于乳腺X线摄影并随密度升高而下降。

研究设计与样本

  • 回顾性机构审查委员会批准的研究,纳入90例女性105个活检证实恶性乳腺病灶,2名独立阅片者评估,参考标准为活检病理。

关键影像表现/征象

  • 肿块性病灶在CT与乳腺X线摄影检出率相当,结构扭曲和微钙化在螺旋乳腺CT检出较少。
  • 诊断信心为非常高或高的比例:增强螺旋乳腺CT 97.2%,平扫螺旋乳腺CT 74.1%,乳腺X线摄影81.4%。

诊断标准与关键数值

  • 阅片者1和2检出率:乳腺X线摄影83.8%和80.0%,增强螺旋乳腺CT 99.1%和99.1%(P<0.05),平扫螺旋乳腺CT 66.7%和61.9%(P<0.05)。
  • 增强螺旋乳腺CT各密度检出率(两阅片者):A/B/C/D均为100%/100%/100%/87.5%。
  • 乳腺X线摄影各密度检出率:A=100%,B=89.1%和95.1%,C=73.1%,D=50.0%和71.4%(P<0.05)。
  • 平扫螺旋乳腺CT各密度检出率:A=87.5%和90.7%,B=65.5%和69.1%,C=54.8%和60.0%,D=37.5%(P<0.05)。

临床应用与阅片要点

  • 增强螺旋乳腺CT在不同乳腺密度下均保持高检出率,可用于评估致密乳腺中的恶性病灶。
  • 平扫螺旋乳腺CT对结构扭曲和微钙化的检出较少,阅片时需结合乳腺X线摄影。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To evaluate the detectability of non-contrast-enhanced and contrast-enhanced spiral breast computed tomography ([non]-CE-SBCT) compared with mammography. Secondary objectives are to determine detectability depending on breast density and to evaluate appearance of breast malignancies according to BI-RADS descriptors.

METHODS: This retrospective institutional review board-approved study included 90 women with 105 biopsy-proven malignant breast lesions. Breast density, BI-RADS descriptors, and detectability were evaluated by 2 independent readers. Diagnostic confidence was rated on a 4-point Likert scale.

RESULTS: For readers 1 and 2, detectability was 83.8% and 80.0% for mammography, 99.1% and 99.1% for CE-SBCT ( P < 0.05), and 66.7% and 61.9% for non-CE-SBCT ( P < 0.05). With both readers, detectability in CE-SBCT was high for density A/B/C/D (both 100%/100%/100%/87.5%). Detectability of readers declined with increasing density for mammography (density A = 100%, B = 89.1% and 95.1%, C = 73.1%, D = 50.0% and 71.4%; P < 0.05) and for non-CE-SBCT (density A = 87.5% and 90.7%, B = 65.5% and 69.1%, C = 54.8% and 60.0%, D = 37.5%; P < 0.05). Mass lesions were detected with CT as often as with mammography, whereas architectural distortions and microcalcifications were detected less often with SBCT. Diagnostic confidence was very high or high in 97.2% for CE-SBCT, in 74.1% for non-CE-SBCT, and in 81.4% for mammography.

CONCLUSIONS: Detectability and diagnostic confidence were very high in CE-SBCT, regardless of breast density. The detectability of non-CE-SBCT was lower than that of mammography and declined with increasing breast density.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Wetzl M, Heilingbrunner T, Heindl F, et al. Detectability of Breast Cancer in Dedicated Breast CT Compared With Mammography Dependent on Breast Density. Invest Radiol. 2024. PMID: 38949016. doi:10.1097/RLI.0000000000001105

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。