商用超声造影剂在膀胱体模中用于次谐波辅助压力估测(SHAPE)的比较研究

A Comparative Study of Commercially Available Ultrasound Contrast Agents for Sub-harmonic-Aided Pressure Estimation (SHAPE) in a Bladder Phantom.
PubMed 文献 · Ultrasound Med Biol · 2024-07-25

本研究在膀胱体模中比较 Definity、Lumason、Sonazoid 与 Optison 四种商用微泡的 SHAPE 测压性能。所有造影剂均显示压力变化与 SHAPE 信号呈预测的负线性关系,但稳定性、线性相关、压力敏感度与误差各异;总体 Definity 与 Lumason 表现最高。结果表明 SHAPE 信号随膀胱体模压力升高而降低,SHAPE 测压效果随微泡配方而异,支持后续在人体可行性研究中使用 Lumason 与 Definity。

要点

  • 四种商用微泡的 SHAPE 信号均随膀胱体模压力升高而降低,与造影剂或 CMG 阶段无关。
  • Definity 与 Lumason 在基于 SHAPE 的膀胱体模压力评估中总体表现最高。
  • SHAPE 测压效果依赖微泡配方,支持选用 Lumason 与 Definity 推进人体可行性研究。

核心结论

  • SHAPE 信号随膀胱体模压力升高而降低,与所用造影剂或 CMG 阶段无关,提示 SHAPE 可能用于无导尿管测量膀胱压力。
  • SHAPE 测压效果随微泡配方不同而变化,Definity 与 Lumason 总体表现最高。
  • 上述观察支持在人体受试者可行性研究中使用 Lumason 与 Definity,以推进经 SHAPE 无导尿管测量排尿期膀胱压力的方案。

研究设计与样本

  • 研究为在既往设计并建造的膀胱体模中比较四种商用超声造影微泡的体外实验。
  • 测试造影剂为 Definity、Lumason、Sonazoid 与 Optison。
  • 采用标准临床膀胱测压(CMG)系统将微泡-生理盐水混合液注入膀胱体模以测量压力。
  • 超声成像使用 GE Healthcare LOGIQ E10 扫描仪。

关键影像表现/征象

  • 所有造影剂均显示压力变化与 SHAPE 信号之间呈预测的负线性关系。
  • 各造影剂在稳定性、线性相关性、对压力的敏感度及误差方面彼此不同。
  • 总体而言,Definity 与 Lumason 在基于 SHAPE 的膀胱体模压力评估中表现最高。

临床应用与阅片要点

  • 作者认为结果支持在人体受试者可行性研究中使用 Lumason 与 Definity,以推进经 SHAPE 无导尿管测量排尿期膀胱压力的方案。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: The goal of this study was to evaluate the performance of different commercial ultrasound contrast microbubbles (MBs) when measuring bladder phantom pressure with sub-harmonic-aided pressure estimation (SHAPE) methodology. We hypothesized that SHAPE performance is dependent on MB formulation. This study aimed to advance the SHAPE application for bladder pressure measurements in humans.

METHODS: Using a previously designed and built bladder phantom, we tested four different commercial agents: Definity, Lumason, Sonazoid and Optison. A standard clinical cystometrogram (CMG) system was used to infuse a MB-saline mixture into the bladder phantom to measure pressure. Ultrasound imaging was performed using the GE Healthcare LOGIQ E10 scanner.

RESULTS: All agents showed a predicted inverse linear relationship between change in pressure and SHAPE signal. However, they differ from each other in terms of stability, linear correlation, sensitivity to pressure and error. Generally, Definity and Lumason showed the highest performance during the SHAPE-based bladder phantom pressure assessments.

CONCLUSION: Our results show that the SHAPE signal decreases as bladder phantom pressures increases, regardless of the agent or CMG phase, suggesting the possibility of using SHAPE for measuring bladder pressure without a catheter. However, the efficacy of SHAPE in measuring pressure varies by MB formulation. These observations support using Lumason and Definity in a human subject feasibility study as we advance toward a catheter-free solution for measuring voiding bladder pressure via SHAPE.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。