文献速递腹部技术与安全CT

双源拆分CT单次扫描模拟多档辐射剂量与剂量匹配单能CT的肝脏病灶检出比较

Dual-Split CT to Simulate Multiple Radiation Doses From a Single Scan-Liver Lesion Detection Compared With Dose-Matched Single-Energy CT.
PubMed 文献 · Invest Radiol · 2024-07-30

研究用含5-10 mm肝脏病灶的腹部体模,在第三代双源CT单能双源模式下以100 kVp和5、2.5、1.25 mGy三档剂量扫描,经线性图像融合生成5档模拟剂量,并以通道化Hotelling模型观察者测病灶检出AUC,与剂量匹配单能CT比较;结论为双源拆分CT模拟多档剂量与剂量匹配单能CT的病灶检出能力相当,噪声纹理略更斑驳,模拟剂量可用于评估降剂量对病灶检出的影响而无需重复扫描患者。

要点

  • 双源拆分CT单次扫描可生成多档模拟辐射剂量,用于评估降剂量对肝脏病灶检出的影响,无需重复扫描患者。
  • 模拟图像噪声纹理较单能CT略更斑驳:1.25 mGy时平均空间频率0.16 mm⁻¹ vs 0.14 mm⁻¹。
  • 5 mGy时两种模式AUC相近;2.5与1.25 mGy时双源拆分模式AUC略低但无显著差异。

核心结论

  • 双源拆分CT单次扫描生成的多个模拟辐射剂量与剂量匹配单能CT的病灶检出能力相当。
  • 模拟图像噪声纹理略更斑驳。
  • 双源拆分CT模拟剂量可用于评估辐射剂量降低对病灶检出能力的影响,而无需对患者重复扫描。

研究设计与样本

  • 研究类型为体模研究,使用含5-10 mm高衰减与低衰减肝脏病灶的腹部体模。
  • 采用第三代双源CT单能双源模式,100 kVp,3档辐射剂量(5、2.5、1.25 mGy),A管管电流为67%、B管为33%。
  • 每档剂量经线性图像融合生成5档模拟辐射剂量(100%、67%、55%、45%、39%、33%),每档设置重复10次。
  • 体模另以传统单源单能模式在等效剂量下成像作为对照。
  • 病灶检出以通道化Hotelling模型观察者(10个密集高斯通道)的AUC评估,噪声纹理以噪声功率谱平均空间频率(f av)评估。

关键影像表现/征象

  • f av随辐射剂量降低而下降,且模拟图像与单能图像不同(如1.25 mGy时模拟图像0.16 mm⁻¹ vs 单能图像0.14 mm⁻¹)。
  • 双源拆分CT的噪声纹理较单能CT略更斑驳。

诊断标准与关键数值

  • 高衰减病灶:单能CT在5、2.5、1.25 mGy的平均AUC分别为0.92-0.99、0.81-0.96、0.68-0.89;双源拆分CT分别为0.91-0.99、0.78-0.91、0.70-0.85。
  • 低衰减病灶:单能CT在5、2.5、1.25 mGy的AUC分别为0.90-0.98、0.75-0.93、0.69-0.86;双源拆分CT分别为0.92-0.99、0.76-0.87、0.67-0.81。
  • 5 mGy时两种模式AUC相近;2.5与1.25 mGy时双源拆分模式AUC略低但无显著差异。

临床应用与阅片要点

  • 双源拆分CT单次扫描生成的模拟剂量可用于评估降剂量对肝脏病灶检出能力的影响,无需重复扫描患者。
  • 解读模拟图像时需注意其噪声纹理较单能CT略更斑驳,f av在1.25 mGy时为0.16 mm⁻¹ vs 0.14 mm⁻¹。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to evaluate the potential use of simulated radiation doses from a dual-split CT scan for dose optimization by comparing their lesion detectability to dose-matched single-energy CT acquisitions at different radiation dose levels using a mathematical model observer.

MATERIALS AND METHODS: An anthropomorphic abdominal phantom with liver lesions (5-10 mm, both hyperattenuating and hypoattenuating) was imaged using a third-generation dual-source CT in single-energy dual-source mode at 100 kVp and 3 radiation doses (5, 2.5, 1.25 mGy). The tube current was 67% for tube A and 33% for tube B. For each dose, 5 simulated radiation doses (100%, 67%, 55%, 45%, 39%, and 33%) were generated through linear image blending. The phantom was also imaged using traditional single-source single-energy mode at equivalent doses. Each setup was repeated 10 times. Image noise texture was evaluated by the average spatial frequency (f av ) of the noise power spectrum. Liver lesion detection was measured by the area under the receiver operating curve (AUC), using a channelized Hotelling model observer with 10 dense Gaussian channels.

RESULTS: F av decreased at lower radiation doses and differed between simulated and single-energy images (eg, 0.16 mm -1 vs 0.14 mm -1 for simulated and single-energy images at 1.25 mGy), indicating slightly blotchier noise texture for dual-split CT. For hyperattenuating lesions, the mean AUC ranged between 0.92-0.99, 0.81-0.96, and 0.68-0.89 for single-energy, and between 0.91-0.99, 0.78-0.91, and 0.70-0.85 for dual-split at 5 mGy, 2.5 mGy, and 1.25 mGy, respectively. For hypoattenuating lesions, the AUC ranged between 0.90-0.98, 0.75-0.93, and 0.69-0.86 for the single-energy, and between 0.92-0.99, 0.76-0.87, and 0.67-0.81 for dual-split at 5 mGy, 2.5 mGy, and 1.25 mGy, respectively. AUC values were similar between both modes at 5 mGy, and slightly lower, albeit not significantly, for the dual-split mode at 2.5 and 1.25 mGy.

CONCLUSIONS: Lesion detectability was comparable between multiple simulated radiation doses from a dual-split CT scan and dose-matched single-energy CT. Noise texture was slightly blotchier in the simulated images. Simulated doses using dual-split CT can be used to assess the impact of radiation dose reduction on lesion detectability without the need for repeated patient scans.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Racine D, Niemann T, Nemeth B, et al. Dual-Split CT to Simulate Multiple Radiation Doses From a Single Scan-Liver Lesion Detection Compared With Dose-Matched Single-Energy CT. Invest Radiol. 2024. PMID: 39074298. doi:10.1097/RLI.0000000000001111

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。