同一批髋关节置换术后患者1.5T与3T前列腺MRI的图像质量比较:包含PI-QUAL第2版的多阅片者评估

Image quality comparison of 1.5T and 3T prostate MRIs of the same post-hip arthroplasty patients: multi-rater assessments including PI-QUAL version 2.
PubMed 文献 · Abdom Radiol (NY) · 2024-07-09

该单中心回顾性研究纳入20例髋关节置换术后患者,配对比较1.5T与3T前列腺MRI图像质量。三位盲法腹部放射科医师评估显示,1.5T的DWI磁敏感伪影评分显著高于3T(2.80±0.41 vs 2.35±0.93,p=0.014),T2WI和DCE的磁敏感伪影评分及PI-QUAL评分均无显著差异。作者结论认为使用1.5T扫描仪可能更有利于减少髋关节假体在DWI上产生的磁敏感伪影。

要点

  • 髋关节置换术后患者行前列腺MRI时,DWI上1.5T的磁敏感伪影评分显著高于3T。
  • T2WI和DCE的磁敏感伪影评分及PI-QUAL评分在1.5T与3T间无显著差异。
  • DWI磁敏感伪影评分的阅片者间一致性为中等(多阅片者广义kappa 0.60)。

核心结论

  • 使用1.5T扫描仪可能更有利于减少髋关节假体在DWI上产生的磁敏感伪影。

研究设计与样本

  • 单中心回顾性研究,纳入20例独特患者的20对1.5T与3T前列腺MRI检查,由三位盲法腹部放射科医师评估DWI(50 s/mm²)、T2WI和DCE图像质量。

关键影像表现/征象

  • DWI(50 s/mm²)上1.5T的磁敏感伪影评分显著高于3T(均值±标准差,2.80±0.41 vs 2.35±0.93,p=0.014)。
  • T2WI和DCE的磁敏感伪影评分在1.5T与3T间无显著差异。
  • PI-QUAL评分在1.5T与3T间无显著差异。

诊断标准与关键数值

  • 磁敏感伪影程度采用三点量表分级,3分表示伪影最少,1分表示伪影最多。
  • 图像质量同时采用前列腺影像质量(PI-QUAL)第2版评估。
  • DWI磁敏感伪影评分的阅片者间一致性为中等(多阅片者广义kappa 0.60),T2WI为完美一致(不适用),DCE为高度一致(0.65)。
  • PI-QUAL评分的阅片者间一致性为一般(0.27)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To compare the image quality of 1.5T and 3T prostate MRIs of the same post-hip arthroplasty patients, with a specific focus on the degree of susceptibility artifacts.

METHODS: This single-center retrospective study included post-hip arthroplasty patients who underwent 1.5T prostate MRIs between 2021 and 2023, as well as comparative 3T prostate MRIs. Three blinded abdominal radiologists retrospectively reviewed their diffusion-weighted imaging (DWI, 50 s/mm2), T2-weighted imaging (T2WI), and dynamic contrast-enhanced imaging (DCE) to evaluate the image quality. The degree of susceptibility artifacts was categorized using a three-point scale, with 3 indicating the least artifact and 1 indicating the most. Image quality was also evaluated using Prostate Imaging Quality (PI-QUAL) version 2. The median of the three raters' scores was compared between 1.5T and 3T prostate MRIs using the Wilcoxon signed-rank test. The inter-rater agreement was evaluated using the multi-rater generalized kappa.

RESULTS: Twenty pairs of 1.5T and 3T prostate MRI examinations from 20 unique patients were included. The DWI susceptibility artifact score at 1.5T was significantly higher than at 3T (mean score ± standard deviation, 2.80 ± 0.41 vs. 2.35 ± 0.93, p = 0.014). In contrast, no significant differences were observed in the susceptibility artifact scores in T2WI and DCE, or in the PI-QUAL score. The inter-reader agreement in the susceptibility artifact score was moderate (multi-rater generalized kappa: 0.60) in DWI, perfect in T2WI (not applicable), and substantial (0.65) in DCE. The inter-reader agreement was fair (0.27) in the PI-QUAL score.

CONCLUSION: Using 1.5T scanners may be preferable to reduce susceptibility artifacts from hip prostheses in DWI.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。