亚毫西弗腹部光子计数CT与能量积分探测器CT对尿路结石检出:对诊断信心的影响
本回顾性单中心研究纳入507例疑似尿路结石患者,比较腹部平扫光子计数探测器(PCD)CT与能量积分探测器(EID)CT的辐射剂量和图像质量。PCD CT有效剂量0.79 mSv低于EID CT的1.39 mSv(降低44%),且肾脏、腰大肌、闭孔肌水平信噪比分别高30%、23%、17%(P < .001),三位阅片者诊断信心无差异(中位5分,P ≥ .57)。
要点
- PCD CT有效剂量0.79 mSv,较EID CT的1.39 mSv降低44%。
- PCD CT肾脏、腰大肌、闭孔肌信噪比分别较EID CT高30%、23%、17%(P < .001)。
- 两位CT方案诊断信心均为中位5分(IQR 5-5),差异无统计学意义(P ≥ .57)。
核心结论
- 亚毫西弗腹部PCD CT可提供高质量图像用于尿路结石诊断,辐射暴露较EID CT降低44%且信噪比更高,阅片者诊断信心无差异证据。
研究设计与样本
- 回顾性单中心研究,纳入2022年2月至2023年3月疑似尿路结石行腹部平扫PCD CT或EID CT的患者507例(PCD CT组229例,EID CT组278例;平均年龄51.7岁±17.4,男性317例)。
关键影像表现/征象
- 在肾脏、腰大肌和闭孔肌三个解剖层面进行信号与噪声测量。
- 无论采用PCD CT或EID CT方案、无论结石位置,阅片者间一致性均接近完美(Krippendorff α ≥ 0.99;95% CI:0.99,1)。
诊断标准与关键数值
- 诊断信心采用五分制记录:1分为几乎没有信心,5分为完全有信心。
- PCD CT组三位阅片者分别检出肾结石129、127、129例,EID CT组分别检出94、94、94例。
- PCD CT组三位阅片者分别检出尿路结石113、114、114例,EID CT组分别检出152、153、152例。
- PCD CT组有效剂量0.79 mSv(IQR 0.63-0.99 mSv),EID CT组1.39 mSv(IQR 1.01-1.87 mSv)。
- PCD CT组肾脏、腰大肌、闭孔肌水平信噪比分别较EID CT组高30%、23%、17%(P < .001)。
临床应用与阅片要点
- 亚毫西弗腹部PCD CT协议含锡预滤过,可用于疑似尿路结石的平扫检查。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Contrast-unenhanced abdominal CT is the imaging standard for urinary calculi detection; however, studies comparing photon-counting detector (PCD) CT and energy-integrating detector (EID) CT dose-reduction potentials are lacking. Purpose To compare the radiation dose and image quality of optimized EID CT with those of an experimental PCD CT scan protocol including tin prefiltration in patients suspected of having urinary calculi. Materials and Methods This retrospective single-center study included patients who underwent unenhanced abdominal PCD CT or EID CT for suspected urinary caliculi between February 2022 and March 2023. Signal and noise measurements were performed at three anatomic levels (kidney, psoas, and obturator muscle). Nephrolithiasis and/or urolithiasis presence was independently assessed by three radiologists, and diagnostic confidence was recorded on a five-point scale (1, little to no confidence; 5, complete confidence). Reader agreement was determined by calculating Krippendorff α. Results A total of 507 patients (mean age, 51.7 years ± 17.4 [SD]; 317 male patients) were included (PCD CT group, 229 patients; EID CT group, 278 patients). Readers 1, 2, and 3 detected nephrolithiasis in 129, 127, and 129 patients and 94, 94, and 94 patients, whereas the readers detected urolithiasis in 113, 114, and 114 patients and 152, 153, and 152 patients in the PCD CT and EID CT groups, respectively. Regardless of protocol (PCD CT or EID CT) or calculus localization, near perfect interreader agreement was found (α ≥ 0.99; 95% CI: 0.99, 1). There was no evidence of a difference in reader confidence between PCD CT and EID CT (median confidence, 5; IQR, 5-5; P ≥ .57). The effective doses were 0.79 mSv (IQR, 0.63-0.99 mSv) and 1.39 mSv (IQR, 1.01-1.87 mSv) for PCD CT and EID CT, respectively. Despite the lower radiation exposure, the signal-to-noise ratios at the kidney, psoas, and obturator levels were 30%, 23%, and 17% higher, respectively, in the PCD CT group (P < .001). Conclusion Submillisievert abdominal PCD CT provided high-quality images for the diagnosis of urinary calculi; radiation exposure was reduced by 44% with a higher signal-to-noise ratio than with EID CT and with no evidence of a difference in reader confidence. Published under a CC BY 4.0 license. Supplemental material is available for this article. See also the editorial by Nezami and Malayeri in this issue.