腹部CT的图像质量与辐射剂量:一项多中心研究

Image quality and radiation doses in abdominal CT: A multicenter study.
PubMed 文献 · Eur J Radiol · 2024-07-24

该多中心研究用体模在33个中心40台CT扫描仪上按本地腹部CT方案扫描一次,评估肝脏和肾脏实质的CNR和SNR并由三名放射科医生按五点Likert量表对13项图像质量标准评分,结果显示CTDIvol范围为3.5至12 mGy,CTDIvol与CNR仅呈弱相关(r=0.270,p=0.092),不同扫描仪间剂量与图像质量差异较大。

要点

  • 不同CT扫描仪间辐射剂量和图像质量差异较大,CTDIvol范围为3.5至12 mGy(中位数5.3 mGy)。
  • CTDIvol与CNR仅呈弱相关(r=0.270,p=0.092),较高剂量并不持续改善CNR。
  • 同一剂量水平CTDIvol下不同扫描仪间CNR存在差异,最旧扫描仪中位CTDIvol高15%、中位CNR低12%。

核心结论

  • 不同CT扫描仪间辐射剂量和图像质量差异较大,CTDIvol与CNR之间的弱相关提示较高剂量并不持续改善CNR,表明需要对腹部CT检查的图像质量和辐射剂量进行系统评估和优化。

研究设计与样本

  • 多中心研究,在33个中心40台CT扫描仪(来自四家厂商)上按本地腹部CT方案对体模扫描一次,评估肝脏和肾脏实质的CNR和SNR,并由三名经验丰富的放射科医生按五点Likert量表对13项图像质量标准进行视觉分级分析,记录每次扫描的CTDIvol。

关键影像表现/征象

  • CTDIvol范围为3.5至12 mGy(中位数5.3 mGy,第三四分位数6.7 mGy)。
  • 肝脏实质SNR范围为4.4至14.4(中位数8.5),CNR范围为2.7至11.2(中位数6.1)。
  • 同一剂量水平CTDIvol下不同扫描仪间CNR存在差异。
  • 最旧扫描仪中位CTDIvol高15%、中位CNR低12%。
  • CTDIvol与CNR呈弱相关(r=0.270,p=0.092)。

诊断标准与关键数值

  • 定量图像质量以肝脏和肾脏实质的CNR和SNR评估。
  • 定性图像质量由三名经验丰富的放射科医生按五点Likert量表对13项图像质量标准进行视觉分级分析。
  • 读者间信度ICC:低剂量组ICC=0.889,中剂量组ICC=0.767,高剂量组ICC=0.847;低CNR组ICC=0.803,中CNR组ICC=0.811。

临床应用与阅片要点

  • 不同扫描仪间剂量与图像质量差异较大,CTDIvol与CNR仅弱相关,提示较高剂量并不持续改善CNR。
  • 同一剂量水平下不同扫描仪间CNR存在差异,最旧扫描仪中位CTDIvol高15%且中位CNR低12%。
  • 作者认为需要对腹部CT检查的图像质量和辐射剂量进行系统评估和优化。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To benchmark image quality and corresponding radiation doses for acute abdominal CT examination across different laboratories and CT manufacturers.

METHOD: An anthropomorphic phantom was scanned once with local abdominal CT protocols at 40 CT scanners, from four vendors, in thirty-three sites. Quantitative image quality was evaluated by CNR and SNR in the liver and kidney parenchyma. Qualitative image quality was assessed by visual grading analysis performed by three experienced radiologists using a five-point Likert scale to score thirteen image quality criteria. The CTDIvol was recorded for each scan. Pearson's correlation coefficient was calculated for the continuous variables, and the intraclass correlation coefficient was used to investigate interrater reliability between the radiologists.

RESULTS: CTDIvol ranged from 3.5 to 12 mGy (median 5.3 mGy, third quartile 6.7 mGy). SNR in liver parenchyma ranged from 4.4 to 14.4 (median 8.5), and CNR ranged from 2.7 to 11.2 (median 6.1). A weak correlation was found between CTDIvol and CNR (r = 0.270, p = 0.092). Variations in CNR across scanners at the same dose level CTDIvol were observed. No significant difference in CTDIvol or CNR was found based on scanner installation year. The oldest scanners had a 15 % higher median CTDIvol and a 12 % lower median CNR. The ICC showed acceptable agreement for all dose groups: low (ICC=0.889), medium (ICC=0.767), high (ICC=0.847), and in low (ICC=0.803) and medium (ICC=0.811) CNR groups.

CONCLUSION: There was large variation in radiation dose and image quality across the different CT scanners. Interestingly, the weak correlation between CTDIvol and CNR indicates that higher doses do not consistently improve CNR, indicating a need for systematic assessment and optimization of image quality and radiation doses for the abdominal CT examination.