三维超声心动图在经导管二尖瓣置换术前预测左心室流出道面积的价值
本研究回顾性纳入2个三级中心2017年10月至2023年5月接受经导管二尖瓣置换术的105例连续患者,用专用软件在基线CT和三维经食管超声心动图(TEE)上投影虚拟瓣膜,测量二尖瓣环尺寸及新左心室流出道(neo-LVOT)尺寸并与基线CT对照。结果显示两种方法预测基线neo-LVOT面积无显著偏倚,但TEE显著低估二尖瓣环面积、周长及内外侧径,且两种方法均显著低估实际neo-LVOT面积,以专用二尖瓣生物瓣膜组为著;CT与TEE预测的neo-LVOT面积均与术后LVOT压差呈负相关。
要点
- TEE预测基线neo-LVOT面积与CT无显著偏倚,可用于术前评估。
- TEE显著低估二尖瓣环面积、周长及内外侧径,测瓣环尺寸需谨慎。
- TEE与CT均显著低估实际neo-LVOT面积,专用二尖瓣生物瓣膜组更明显。
核心结论
- TEE衍生的分析在预测neo-LVOT面积及TMVR术后峰值压差方面与CT衍生的指标结果相当。
研究设计与样本
- 回顾性研究,纳入2个三级中心2017年10月至2023年5月接受TMVR的105例连续患者,使用专用软件在基线CT和三维TEE上投影虚拟瓣膜,以基线CT为参考标准。
关键影像表现/征象
- 两种影像方法预测基线neo-LVOT无显著偏倚。
- TEE较CT显著低估二尖瓣环面积、周长及内外侧径。
- 两种方法均显著低估实际neo-LVOT面积,以专用二尖瓣生物瓣膜治疗组为著。
诊断标准与关键数值
- TEE预测与实际neo-LVOT面积的平均偏倚25.6 mm2,一致性界限-92.2 mm2至143.3 mm2,P < 0.001。
- CT预测与实际neo-LVOT面积的平均偏倚28.3 mm2,一致性界限-65.8 mm2至122.4 mm2,P = 0.046。
- TEE预测neo-LVOT面积与术后LVOT压差呈负相关,r2 = 0.481,P < 0.001。
- CT预测neo-LVOT面积与术后LVOT压差呈负相关,r2 = 0.401,P < 0.001。
临床应用与阅片要点
- 三维TEE可作为TMVR术前预测neo-LVOT面积的检查方法,结果与CT相当。
- TEE测量二尖瓣环尺寸时需注意其较CT显著低估面积、周长及内外侧径。
- TEE与CT预测的neo-LVOT面积均可能低估实际值,专用二尖瓣生物瓣膜组更明显。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: New postprocessing software facilitates 3-dimensional (3D) echocardiographic determination of mitral annular (MA) and neo-left ventricular outflow tract (neo-LVOT) dimensions in patients undergoing transcatheter mitral valve replacement (TMVR).
OBJECTIVES: This study aims to test the accuracy of 3D echocardiographic analysis as compared to baseline computed tomography (CT).
METHODS: A total of 105 consecutive patients who underwent TMVR at 2 tertiary care centers between October 2017 and May 2023 were retrospectively included. A virtual valve was projected in both baseline CT and 3D transesophageal echocardiography (TEE) using dedicated software. MA dimensions were measured in baseline images and neo-LVOT dimensions were measured in baseline and postprocedural images. All measurements were compared to baseline CT as a reference. The predicted neo-LVOT area was correlated with postprocedural peak LVOT gradients.
RESULTS: There was no significant bias in baseline neo-LVOT prediction between both imaging modalities. TEE significantly underestimated MA area, perimeter, and medial-lateral dimension compared to CT. Both modalities significantly underestimated the actual neo-LVOT area (mean bias pre/post TEE: 25.6 mm2, limit of agreement: -92.2 mm2 to 143.3 mm2; P < 0.001; mean bias pre/post CT: 28.3 mm2, limit of agreement: -65.8 mm2 to 122.4 mm2; P = 0.046), driven by neo-LVOT underestimation in the group treated with dedicated mitral valve bioprosthesis. Both CT- and TEE-predicted-neo-LVOT areas exhibited an inverse correlation with postprocedural LVOT gradients (r2 = 0.481; P < 0.001 for TEE and r2 = 0.401; P < 0.001 for CT).
CONCLUSIONS: TEE-derived analysis provides comparable results with CT-derived metrics in predicting the neo-LVOT area and peak gradient after TMVR.