超短回波时间MR成像显示Woven EndoBridge装置治疗颅内动脉瘤的可视化研究
前瞻性纳入12例患者(9例女性)共15个经Woven EndoBridge(WEB)装置治疗的颅内动脉瘤,行18次MRI检查,比较对比增强3D各向同性超短回波时间(UTE)MRI与标准TOF-MRA(增强及平扫)及DSA对装置形态、载瘤血管伪影性虚拟狭窄及WEB闭塞分级的显示。UTE在17/18次MRI中显示装置位置与形态优于TOF-MRA,载瘤血管伪影性虚拟狭窄显著低于TOF及增强TOF,再灌注分级正确17例(TOF与增强TOF各16例)。
要点
- UTE序列在17/18次MRI中显示WEB装置位置与形态优于TOF-MRA。
- UTE-MRI的载瘤血管伪影性虚拟狭窄显著低于TOF及增强TOF。
- 再灌注分级正确例数:UTE 17例,TOF 16例,增强TOF 16例。
核心结论
- 对比增强UTE是新型MRI序列,在WEB装置治疗颅内动脉瘤的无创、无辐射随访成像中较标准序列显示出优势。
研究设计与样本
- 前瞻性研究,纳入12例患者(9例女性)共15个治疗动脉瘤,行18次MRI检查。
- 随访UTE-MRI在同一台3T扫描仪上进行,并与TOF-MRA(增强及平扫)及DSA比较。
- 由两名介入神经放射科医师分别独立及共识评分。
关键影像表现/征象
- UTE序列在17/18次MRI中显示WEB装置位置与形态评为优于或明显优于TOF-MRA。
- UTE-MRI的载瘤血管伪影性虚拟狭窄显著低于TOF及增强TOF。
- DSA在8/18次随访中可见再灌注。
诊断标准与关键数值
- 再灌注分级正确例数:TOF 16例,增强TOF 16例,UTE 17例。
- 以DSA为参考,DSA在8/18次随访中显示再灌注。
临床应用与阅片要点
- 对比增强3D UTE序列可用于WEB装置治疗颅内动脉瘤随访中对装置形态与再灌注的无创、无辐射评估。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Assessing the treatment success of intracranial aneurysms treated with Woven EndoBridge (WEB) devices using MRI is important in follow-up imaging. Depicting both the device configuration as well as reperfusion is challenging due to susceptibility artifacts. We evaluated the usefulness of the contrast-enhanced 3D ultrashort TE (UTE) sequence in this setting.
MATERIALS AND METHODS: In this prospective study, 12 patients (9 women) with 15 treated aneurysms were included. These 12 patients underwent 18 MRI examinations. Follow-up UTE-MRI controls were performed on the same 3T scanner. We compared the visualization of device configuration, artifact-related virtual stenosis of the parent vessel, and the WEB occlusion scale in 3D isotropic UTE-MRI postcontrast with standard TOF-MRA with contrast-enhancement (CE) and without IV contrast as well as DSA. Two interventional neuroradiologists rated the images separately and in consensus.
RESULTS: Visualization of the WEB device position and configuration was rated superior or highly superior using the UTE sequence in 17/18 MRIs compared with TOF-MRA. Artifact-related virtual stenosis of the parent vessel was significantly lower in UTE-MRI compared with TOF and CE-TOF. Reperfusion was visible in 8/18 controls on DSA. TOF was able to grade reperfusion correctly in 16 cases; CE-TOF, in 16 cases; and UTE, in 17 cases.
CONCLUSIONS: Contrast-enhanced UTE is a novel MRI sequence that shows benefit compared with the standard sequences in noninvasive and radiation-free follow-up imaging of intracranial aneurysms treated using the WEB device.