DANTE-CAIPI加速增强3D T1:基于深度学习的图像质量改善用于血管壁MRI
本研究纳入64例连续患者,对比传统增强3D T1-SPACE、DANTE-CAIPI-SPACE及DANTE-CAIPI-SPACE+DL三种序列的颅内血管壁MRI图像质量。结果显示DANTE-CAIPI-SPACE+DL较SPACE显著减少动静脉血流伪影,图像质量、SNR、伪影及病灶评估无显著差异,且定量SNR和CNR显著高于DANTE-CAIPI-SPACE。作者认为深度学习去噪的DANTE-CAIPI-SPACE可在维持或改善图像质量的同时缩短扫描时间。
要点
- DANTE-CAIPI-SPACE+DL较SPACE显著减少动脉和静脉血流伪影(均P < .001)。
- DANTE-CAIPI-SPACE+DL与SPACE在图像质量、SNR、伪影及病灶评估上无显著差异。
- DANTE-CAIPI-SPACE+DL的定量血管壁SNR和CNR显著高于DANTE-CAIPI-SPACE(均P < .001)。
核心结论
- 深度学习去噪的增强T1加权DANTE-CAIPI-SPACE加速血流抑制颅内血管壁MRI较传统增强颅内血管壁MRI血流抑制改善、扫描时间更短,定性及定量SNR指标相当,且SNR指标优于增强DANTE-CAIPI-SPACE颅内血管壁MRI;实施该技术有潜力在维持或改善图像质量的同时缩短扫描方案时间。
研究设计与样本
- 单中心连续纳入64例患者,均接受颅内血管壁MRI检查。
- 对比序列包括传统增强3D T1-SPACE、DANTE-CAIPI-SPACE及经展开式深度卷积网络去噪的DANTE-CAIPI-SPACE+DL。
- 采用4分或5分Likert量表评估整体图像质量、SNR、伪影严重程度、动静脉抑制及病灶评估,并行SNR和CNR定量评估。
关键影像表现/征象
- DANTE-CAIPI-SPACE+DL较SPACE动脉血流伪影评分1(1-1.75)vs 3(3-4),P < .001。
- DANTE-CAIPI-SPACE+DL较SPACE静脉血流伪影评分1(1-2)vs 3(3-4),P < .001。
- DANTE-CAIPI-SPACE+DL与SPACE在图像质量、SNR、伪影评分及病灶评估方面无显著差异。
- DANTE-CAIPI-SPACE+DL较DANTE-CAIPI-SPACE的SNR评分2(1-2)vs 3(2-3),P < .001。
- DANTE-CAIPI-SPACE与DANTE-CAIPI-SPACE+DL在图像质量、伪影、动脉血流伪影、静脉血流伪影及病灶评估方面无统计学显著差异。
诊断标准与关键数值
- DANTE-CAIPI-SPACE+DL定量血管壁SNR中位数9.71、CNR中位数4.24,均显著高于DANTE-CAIPI-SPACE的SNR 5.50、CNR 2.64(均P < .001)。
- SPACE定量血管壁SNR中位数10.82、CNR中位数5.21,与DANTE-CAIPI-SPACE+DL无显著差异。
临床应用与阅片要点
- DANTE-CAIPI-SPACE+DL可提供与SPACE相当的图像质量、SNR、伪影及病灶评估,同时血流抑制更优、扫描时间更短。
- DANTE-CAIPI-SPACE+DL较DANTE-CAIPI-SPACE改善SNR评分及定量SNR、CNR。
- 作者认为实施DL去噪的DANTE-CAIPI-SPACE颅内血管壁MRI有潜力在维持或改善图像质量的同时缩短扫描方案时间。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND AND PURPOSE: Accelerated and blood-suppressed postcontrast 3D intracranial vessel wall MRI (IVW) enables high-resolution rapid scanning but is associated with low SNR. We hypothesized that a deep-learning (DL) denoising algorithm applied to accelerated, blood-suppressed postcontrast IVW can yield high-quality images with reduced artifacts and higher SNR in shorter scan times.
MATERIALS AND METHODS: Sixty-four consecutive patients underwent IVW, including conventional postcontrast 3D T1-sampling perfection with application-optimized contrasts by using different flip angle evolution (SPACE) and delay alternating with nutation for tailored excitation (DANTE) blood-suppressed and CAIPIRINHIA-accelerated (CAIPI) 3D T1-weighted TSE postcontrast sequences (DANTE-CAIPI-SPACE). DANTE-CAIPI-SPACE acquisitions were then denoised by using an unrolled deep convolutional network (DANTE-CAIPI-SPACE+DL). SPACE, DANTE-CAIPI-SPACE, and DANTE-CAIPI-SPACE+DL images were compared for overall image quality, SNR, severity of artifacts, arterial and venous suppression, and lesion assessment by using 4-point or 5-point Likert scales. Quantitative evaluation of SNR and contrast-to-noise ratio (CNR) was performed.
RESULTS: DANTE-CAIPI-SPACE+DL showed significantly reduced arterial (1 [1-1.75] versus 3 [3-4], P < .001) and venous flow artifacts (1 [1-2] versus 3 [3-4], P < .001) compared with SPACE. There was no significant difference between DANTE-CAIPI-SPACE+DL and SPACE in terms of image quality, SNR, artifact ratings, and lesion assessment. For SNR ratings, DANTE-CAIPI-SPACE+DL was significantly better compared with DANTE-CAIPI-SPACE (2 [1-2], versus 3 [2-3], P < .001). No statistically significant differences were found between DANTE-CAIPI-SPACE and DANTE-CAIPI-SPACE+DL for image quality, artifact, arterial blood and venous blood flow artifacts, and lesion assessment. Quantitative vessel wall SNR and CNR median values were significantly higher for DANTE-CAIPI-SPACE+DL (SNR: 9.71, CNR: 4.24) compared with DANTE-CAIPI-SPACE (SNR: 5.50, CNR: 2.64) (P < .001 for each), but there was no significant difference between SPACE (SNR: 10.82, CNR: 5.21) and DANTE-CAIPI-SPACE+DL.
CONCLUSIONS: DL denoised postcontrast T1-weighted DANTE-CAIPI-SPACE accelerated and blood-suppressed IVW showed improved flow suppression with a shorter scan time and equivalent qualitative and quantitative SNR measures relative to conventional postcontrast IVW. It also improved SNR metrics relative to postcontrast DANTE-CAIPI-SPACE IVW. Implementing DL denoised DANTE-CAIPI-SPACE IVW has the potential to shorten protocol time while maintaining or improving the image quality of IVW.
引用
- Kharaji M, Canton G, Guo Y, et al. DANTE-CAIPI Accelerated Contrast-Enhanced 3D T1: Deep Learning-Based Image Quality Improvement for Vessel Wall MRI. AJNR Am J Neuroradiol. 2025. PMID: 39038956. doi:10.3174/ajnr.A8424
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。