文献速递技术与安全腹部CT

光子计数探测器CT在不同扫描设置下脂肪定量的稳定性与准确性:体模研究

Stability and accuracy of fat quantification on photon-counting detector CT with various scan settings: A phantom study.
PubMed 文献 · Eur J Radiol · 2024-06-08

本研究用含不同脂肪分数的琼脂体模,在光子计数探测器CT上组合3种扫描模式、4种管电压和3种图像质量水平进行扫描,以常规螺旋120 kV IQ80图像建立的CT值-脂肪分数转换关系为参考。结果显示各扫描设置的组内相关系数均大于0.9、偏倚小于5%,均方根误差为1.0%–5.0%,提示70 keV重建的光子计数探测器CT脂肪定量准确、稳定,并具有超低剂量应用潜力。

要点

  • 光子计数探测器CT在70 keV重建下,各扫描设置的脂肪分数与参考标准一致性高,组内相关系数均大于0.9。
  • 各扫描设置偏倚均小于5%,均方根误差为1.0%–5.0%,高螺距120 kV IQ20扫描的均方根误差最低(1.0%)。
  • 螺旋120 kV IQ20与IQ80扫描的偏倚最低,平均值分别为1.19%和1.23%。

核心结论

  • 光子计数探测器CT在70 keV重建下于多种扫描设置中脂肪定量均准确、稳定,并具有超低辐射剂量下进行脂肪定量的较大潜力。

研究设计与样本

  • 体模研究,使用11个含不同脂肪分数(范围0%–100%)的琼脂体模,在光子计数探测器CT上组合3种扫描模式、4种管电压和3种图像质量水平,每种设置扫描两次,均以70 keV相同重建参数生成图像。

诊断标准与关键数值

  • 以常规螺旋120 kV IQ80扫描建立CT值至脂肪分数的转换关系作为参考标准;所有扫描设置的组内相关系数大于0.9(范围0.929–0.998,p<0.017),偏倚小于5%(范围-2.9%–5%),均方根误差为1.0%–5.0%。

临床应用与阅片要点

  • 常规螺旋120 kV IQ80扫描可用于建立CT值至脂肪分数的转换关系,作为其他扫描设置脂肪定量的参考。
  • 高螺距120 kV IQ20扫描的均方根误差最低(1.0%),螺旋120 kV IQ20与IQ80扫描的偏倚最低(平均值分别为1.19%和1.23%)。
  • 各扫描设置下组内相关系数均大于0.9且偏倚小于5%,提示70 keV重建的光子计数探测器CT脂肪定量在不同扫描设置间稳定。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Fat deposition is an important marker of many metabolic diseases. As a noninvasive and convenient examination method, CT has been widely used for fat quantification. With the clinical application of photon-counting detector (PCD)-CT, we aimed to investigate the accuracy, stability, and dose level of PCD-CT using various scan settings for fat quantification.

MATERIALS AND METHODS: Eleven agar-based lipid-containing phantoms (vials with different fat fractions [FFs]; range: 0 %-100 %) were scanned using PCD-CT. Three scanning types (sequence scan, regular spiral scan with a pitch of 0.8, and high-pitch spiral scan with a pitch of 3.2), four tube voltages (90, 120, 140, and 100 kV with a tin filter), and three image quality (IQ) levels (IQ levels of 20, 40, and 80) were alternated, and each scan setting was used twice. For each scan, a 70-keV image was generated using the same reconstruction parameters. A regular spiral scan at 120 kV with IQ80 was used to transfer the CT numbers of all scans to the FF. Intraclass correlation coefficient (ICC) and Bland-Altman analysis were implemented for accuracy and agreement evaluation, and group differences were compared using analysis of variance.

RESULTS: Excellent agreement and accuracy of FF derived by PCD-CT with all scan settings was demonstrated by high ICCs (>0.9; range: 0.929-0.998, p < 0.017) and low bias (<5% range: -2.9 %-5%). The root mean square error (RMSE) between the PCD-CT-acquired FF and the reference standard ranged from 1.0 % to 5.0 %, among which the high-pitch scan at 120 kV with IQ20 accounted for the lowest RMSE (1.0 %). The spiral scan at 120 kV with IQ20 and IQ80 yielded the lowest bias (mean value: 1.19 % and 1.23 %, respectively).

CONCLUSION: Fat quantification using PCD-CT reconstructed at 70 keV was accurate and stable under various scan settings. PCD-CT has great potential for fat quantification using ultralow radiation doses.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Zhang X, Bao S, Wang M, et al. Stability and accuracy of fat quantification on photon-counting detector CT with various scan settings: A phantom study. Eur J Radiol. 2024. PMID: 38878499. doi:10.1016/j.ejrad.2024.111545

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。