专用光子计数CT用于乳腺微钙化的检测与分类:与数字乳腺断层合成的个体内比较

Dedicated Photon-Counting CT for Detection and Classification of Microcalcifications: An Intraindividual Comparison With Digital Breast Tomosynthesis.
PubMed 文献 · Invest Radiol · 2024-06-27

前瞻性个体内研究纳入22例DBT筛查发现的BI-RADS 4/5类微钙化女性,活检前行专用光子计数乳腺CT(PC-BCT)并与DBT比较。PC-BCT在非后部区域可可靠检出微钙化,分布评估与DBT高度一致,但形态学评估一致性差,且不能评估单个钙化形态。

要点

  • PC-BCT对非后部乳腺微钙化的检出与DBT MIP重建(RM-1)清晰度相似(4.83 vs 4.86,P=0.66)。
  • PC-BCT与DBT的分布描述符一致性好(κ=0.91),但形态学描述符一致性差(κ=0.44)。
  • PC-BCT最小可辨钙化颗粒中位0.2 mm,优于RM-1的0.6 mm和RM-2的1.1 mm(P<0.001)。

核心结论

  • PC-BCT可可靠检出非后部乳腺微钙化并评估其分布,但不能评估单个微钙化的形态。

研究设计与样本

  • 前瞻性个体内研究,纳入22例DBT筛查发现BI-RADS 4/5类微钙化的女性,活检前行PC-BCT,重建为0.15 mm各向同性(RM-1)和0.3 mm(RM-2)及薄层MIP,两名放射科医师独立评分并与DBT直接比较。

关键影像表现/征象

  • 2/22例后部区域微钙化在PC-BCT上未检出,可能因未包含在PC-BCT扫描容积内。
  • 其余20例中微钙化可检出,读者间一致性高(κ>0.8)。
  • 微钙化清晰度:DBT与RM-1的MIP重建相似(4.83±0.38 vs 4.86±0.35,P=0.66),其余PC-BCT重建显著更低(P<0.05):RM-2为2.11±0.92,RM-2的MIP为2.64±0.80,RM-1为3.50±1.23。
  • 18/20例PC-BCT与DBT分布描述符相同(κ=0.91),仅5/20例形态学描述符相同(κ=0.44)。
  • 最小可辨微钙化颗粒中位尺寸:DBT 0.2 mm vs RM-1 0.6 mm vs RM-2 1.1 mm(P<0.001),可能因 blooming 效应。

诊断标准与关键数值

  • 微钙化检出采用5分制评分,读者间一致性κ>0.8。
  • 分布和形态按BI-RADS描述符评估,分布一致性κ=0.91,形态一致性κ=0.44。
  • 平均腺体剂量:PC-BCT 7.04 mGy vs DBT 6.88 mGy(P=0.67)。

临床应用与阅片要点

  • PC-BCT可用于非后部乳腺微钙化的检出和分布评估,但不宜用于评估单个微钙化形态。
  • 后部区域微钙化可能因未包含在扫描容积内而漏检,阅片时需注意。
  • RM-1的MIP重建清晰度与DBT相似,RM-1和RM-2常规重建清晰度显著低于DBT。

局限性

  • 2/22例后部区域微钙化在PC-BCT上未检出,可能因未包含在PC-BCT扫描容积内。
  • 形态学描述符一致性差(κ=0.44),PC-BCT不能评估单个微钙化形态。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Clinical experience regarding the use of dedicated photon-counting breast CT (PC-BCT) for diagnosis of breast microcalcifications is scarce. This study systematically compares the detection and classification of breast microcalcifications using a dedicated breast photon-counting CT, especially designed for examining the breast, in comparison with digital breast tomosynthesis (DBT).

MATERIALS AND METHODS: This is a prospective intraindividual study on women with DBT screening-detected BI-RADS-4/-5 microcalcifications who underwent PC-BCT before biopsy. PC-BCT images were reconstructed with a noninterpolated spatial resolution of 0.15 × 0.15 × 0.15 mm (reconstruction mode 1 [RM-1]) and with 0.3 × 0.3 × 0.3 mm (reconstruction mode 2 [RM-2]), plus thin-slab maximum intensity projection (MIP) reconstructions. Two radiologists independently rated the detection of microcalcifications in direct comparison with DBT on a 5-point scale. The distribution and morphology of microcalcifications were then rated according to BI-RADS. The size of the smallest discernible microcalcification particle was measured. For PC-BCT, the average glandular dose was determined by Monte Carlo simulations; for DBT, the information provided by the DBT system was used.

RESULTS: Between September 2022 and July 2023, 22 participants (mean age, 61; range, 42-85 years) with microcalcifications (16 malignant; 6 benign) were included. In 2/22 with microcalcifications in the posterior region, microcalcifications were not detectable on PC-BCT, likely because they were not included in the PC-BCT volume. In the remaining 20 participants, microcalcifications were detectable. With high between-reader agreement (κ > 0.8), conspicuity of microcalcifications was rated similar for DBT and MIPs of RM-1 (mean, 4.83 ± 0.38 vs 4.86 ± 0.35) ( P = 0.66), but was significantly lower ( P < 0.05) for the remaining PC-BCT reconstructions: 2.11 ± 0.92 (RM-2), 2.64 ± 0.80 (MIPs of RM-2), and 3.50 ± 1.23 (RM-1). Identical distribution qualifiers were assigned for PC-BCT and DBT in 18/20 participants, with excellent agreement (κ = 0.91), whereas identical morphologic qualifiers were assigned in only 5/20, with poor agreement (κ = 0.44). The median size of smallest discernible microcalcification particle was 0.2 versus 0.6 versus 1.1 mm in DBT versus RM-1 versus RM-2 ( P < 0.001), likely due to blooming effects. Average glandular dose was 7.04 mGy (PC-BCT) versus 6.88 mGy (DBT) ( P = 0.67).

CONCLUSIONS: PC-BCT allows reliable detection of in-breast microcalcifications as long as they are not located in the posterior part of the breast and allows assessment of their distribution, but not of their individual morphology.

来源:Invest Radiol,查看原文。版权归原作者所有。