文献速递泌尿生殖腹部CT

全前列腺切除术后人工尿道括约肌植入并发症的影像学特征

Radiologic feature of complications after artificial urinary sphincter implantation following total prostatectomy.
PubMed 文献 · Abdom Radiol (NY) · 2024-06-08

本文阐述了全前列腺切除术后人工尿道括约肌(AUS)植入后并发症的影像学表现,重点描述电路泄漏导致的压力下降在CT上表现为管路周围积液、球囊塌陷、控制泵变形及管路内气体,并强调影像在评估尿路损伤、血肿/血清肿及感染中的作用。

要点

  • AUS功能障碍多因电路泄漏致内部压力下降,CT可见管路周围积液、球囊塌陷、控制泵变形及管路内气体。
  • 电路完整时功能障碍可源于泵无法按压(疼痛或阴囊外移位)或膀胱出血/血肿致尿路梗阻。
  • 影像可评估AUS植入或更换相关的尿路损伤、血肿/血清肿及感染。

核心结论

  • 理解AUS功能及其CT表现对准确评估AUS功能障碍和植入后并发症至关重要,可指导临床决策并改善患者预后。

研究设计与样本

  • 本文为阐述AUS功能及并发症影像学表现的综述性文章,未报告研究设计、样本量、中心数或参考标准。

关键影像表现/征象

  • 电路泄漏致内部压力下降时,影像可见管路周围积液、球囊塌陷、控制泵变形及管路内气体。
  • 电路完整时,功能障碍可表现为泵因疼痛或阴囊外移位而无法按压,或膀胱出血/血肿导致尿路梗阻。

临床应用与阅片要点

  • AUS系统由维持压力的球囊、作为操作开关的控制泵、压迫尿道的袖带及连接各组件的管路和接头组成,内部压力介质维持连续回路。
  • 评估AUS功能障碍时,应识别电路泄漏的影像征象(管路周围积液、球囊塌陷、控制泵变形、管路内气体)及电路完整时的其他病因(泵无法按压、膀胱出血/血肿致梗阻)。
  • 影像在评估AUS植入或更换相关的尿路损伤、血肿/血清肿及感染中起关键作用。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Incontinence following total prostatectomy for prostate cancer significantly impairs patient's quality of life. In severe cases, implantation of an artificial urinary sphincter (AUS) has shown favorable outcomes, enhancing continence by constricting the bulbous urethra. The AUS system consists of a pressure-maintaining balloon, control pump serving as the operational switch, cuff that constricts the urethra, and tubes and connectors that link these components, maintaining a continuous circuit through an internal pressure medium. Most instances of AUS dysfunction are attributed to circuit leaks leading to a reduction in internal pressure, which is identifiable on imaging by fluid accumulation around the circuit, balloon collapse, control pump deformation, and air within the circuit. When the AUS circuit is uncompromised, dysfunction may arise from issues such as the inability to compress the pump due to pain or displacement outside the scrotum or urinary tract obstruction caused by bladder hemorrhage/hematoma. Imaging plays a pivotal role in the evaluation of urinary tract injuries, hematomas/seromas, and infections associated with AUS placement or replacement. Understanding the function of AUS and its appearance on CT imaging is essential for accurately assessing AUS dysfunction and post-implantation complications, guiding clinical decision-making and improving patient care outcomes.

来源:Abdom Radiol (NY),查看原文。版权归原作者所有。

引用

  1. Oda S, Kuno H, Hiyama T, et al. Radiologic feature of complications after artificial urinary sphincter implantation following total prostatectomy. Abdom Radiol (NY). 2024. PMID: 38849538. doi:10.1007/s00261-024-04360-2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。