超声背向散射系数用于脂肪定量受测量深度影响

Ultrasound backscatter coefficient for fat quantification is affected by the measurement depth.
PubMed 文献 · Abdom Radiol (NY) · 2024-06-04

本研究回顾性分析104例受试者,使用配备超声衍生脂肪分数(UDFF)算法的Sequoia超声系统,将感兴趣区上缘置于肝包膜下1.5至5 cm处,提取背向散射系数(BSC)数据。BSC随深度对数增加每1 cm下降-13.98 dB/cm-steradian(95%CI -21.016至-5.379,P=0.001),且皮肤至肝包膜距离和硬度也是BSC的独立预测因子。结论认为BSC估计存在深度依赖性,需要标准化采集方案。

要点

  • BSC随测量深度增加而下降,深度是对BSC的独立预测因子。
  • 皮肤至肝包膜距离和肝硬度也是BSC的独立预测因子。
  • 比较BSC结果或评估随时间变化时需标准化采集深度。

核心结论

  • 肝脏BSC的估计表现出深度依赖性,显著影响结果,需要标准化采集方案以比较结果和可靠评估随时间的变化。

研究设计与样本

  • 回顾性研究,共104例个体(56例女性;年龄57.9±13.0岁;BMI 29.0±6.5 kg/m2;皮肤至肝包膜距离2.3±0.7 cm;肝硬度7.5±5.5 kPa),使用配备UDFF算法的Sequoia超声系统,将感兴趣区上缘置于肝包膜下1.5、2、3、4、5 cm处,离线提取BSC数据,采用分数多项式回归分析。

诊断标准与关键数值

  • BSC随深度对数增加每1 cm下降-13.98 dB/cm-steradian(系数-13.98;95%CI -21.016至-5.379;P=0.001)。
  • 最佳拟合模型包含深度作为线性因子和平方根的组合。
  • 皮肤至肝包膜距离和硬度也是BSC的独立预测因子。

临床应用与阅片要点

  • 日常阅片和报告中需注意BSC测量深度会影响结果,比较不同深度或不同时间点的BSC值时应谨慎。
  • 作者结论认为需要标准化采集协议以比较结果和可靠评估变化。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: It has been reported that the estimate of ultrasound attenuation coefficient (AC) is affected by depth of measurement, with linear decrease of values with depth. It is unknown whether backscatter coefficient (BSC) has similar behavior.

METHODS: This retrospective study was performed with Sequoia US system equipped with ultrasound derived fat fraction (UDFF) algorithm (Siemens Healthineers, Issaquah, WA, USA) that combines BSC with AC. UDFF was obtained positioning upper edge of the region of interest at 1.5,2,3,4,5 cm below liver capsule. BSC data were extracted from UDFF offline. A fractional polynomial regression, which selects the best model considering the polynomial development of the variables of interest, was used. Covariates included were age, sex, skin-to-liver-capsule distance, stiffness. Distance was included as linear factor or with a power ranging from - 2 to 3, and the best fitting model was chosen according to partial F test. Body mass index (BMI) was not included because of collinearity with skin-to-liver capsule distance.

RESULTS: 104 individuals (56 females; age: 57.9 ± 13.0 years; BMI: 29.0 ± 6.5 kg/m2; skin-to-liver-capsule distance: 2.3 ± 0.7 cm; liver stiffness: 7.5 ± 5.5 kiloPascal) were studied. Best fitting model for BSC included a combination of depth as linear factor and square root. BSC showed a decrease of - 13.98 dB/cm-steradian for each logarithmic increase of 1 cm depth (coefficient: - 13.98; 95% CI: - 21.016; - 5.379; p = .001). Skin-to-liver-capsule distance and stiffness also were independent predictors of BSC.

CONCLUSIONS: The estimation of the BSC in the liver exhibits a depth dependence that significantly affects results. A standardized acquisition protocol is needed to compare results and reliably assess changes over time.