心血管植入式电子装置导线断裂或废弃患者的MRI检查
这项回顾性研究纳入80例携带废弃导线或导线碎片的CIED患者,共行107次1.5-T MRI检查,30天内未报告死亡、临床显著心律失常或不良临床事件;3例废弃导线患者出现捕获阈值、感知或导线阻抗复合终点的显著变化,多变量分析显示导线方向、长度、MRI类型及检查时长与该复合终点变化无关。
要点
- 携带废弃或断裂CIED导线患者行1.5-T MRI后30天内未报告死亡、临床显著心律失常或不良临床事件。
- 80例中3例废弃导线患者出现捕获阈值、感知或导线阻抗复合终点的显著变化。
- 多变量分析显示导线方向、长度、MRI类型及检查时长与复合终点显著变化无关。
核心结论
- 在导线长度和方向各异的废弃CIED导线或导线碎片患者中,使用1.5-T MRI与不良临床事件无关。
研究设计与样本
- 符合《健康保险可携性与责任法案》的回顾性研究,纳入2014年3月至2020年7月接受1.5-T MRI的CIED废弃导线或导线碎片患者80例、废弃导线或碎片126根、MRI检查107次;比较MRI前后CIED参数,并评估MRI当时及之后30天内的不良临床事件。
诊断标准与关键数值
- 临床显著变异定义为MRI前后询问时捕获阈值变化至少50%、感知变化至少40%或导线阻抗变化至少30%的复合终点。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Purpose To examine the clinical effect of lead length and lead orientation in patients with cardiac implantable electronic devices (CIEDs) and lead fragments or abandoned leads undergoing 1.5-T MRI. Materials and Methods This Health Insurance Portability and Accountability Act-compliant retrospective study included patients with CIEDs and abandoned leads or lead fragments undergoing 1.5-T MRI from March 2014 through July 2020. CIED settings before and after MRI were reviewed, with clinically significant variations defined as a composite of the change in capture threshold of at least 50%, in sensing of at least 40%, or in lead impedance of at least 30% between before MRI and after MRI interrogation. Adverse clinical events were assessed at MRI and up to 30 days after. Univariable and multivariable analysis was performed. Results Eighty patients with 126 abandoned CIED leads or lead fragments underwent 107 1.5-T MRI examinations. Sixty-seven patients (median age, 74 years; IQR, 66-78 years; 44 male patients, 23 female patients) had abandoned leads, and 13 (median age, 66 years; IQR, 52-74 years; nine male patients, four female patients) had lead fragments. There were no reported deaths, clinically significant arrhythmias, or adverse clinical events within 30 days of MRI. Three patients with abandoned leads had a significant change in the composite of capture threshold, sensing, or lead impedance. In a multivariable generalized estimating equation analysis, lead orientation, lead length, MRI type, and MRI duration were not associated with a significant change in the composite outcome. Conclusion Use of 1.5-T MRI in patients with abandoned CIED leads or lead fragments of varying length and orientation was not associated with adverse clinical events. Keywords: Cardiac Assist Devices, MRI, Cardiac Implantable Electronic Device Supplemental material is available for this article. © RSNA, 2024.
引用
- Greenhill MJ, Rangan P, Su W, et al. MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads. Radiol Cardiothorac Imaging. 2024. PMID: 38869431. doi:10.1148/ryct.230303
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。