新型运动校正算法提升CT血流储备分数诊断性能
本研究回顾性纳入151例在3个月内接受冠状动脉CT血管成像(CCTA)和有创冠状动脉造影(ICA)/血流储备分数(FFR)检查的患者,比较标准图像(STD)与第一代(SSF1)和第二代(SSF2)运动校正算法处理后的CT-FFR诊断性能。SSF2组整体图像质量评分中位数3.67,高于STD组和SSF1组的3(P<0.001);SSF2组CT-FFR与FFR相关性最佳(r=0.652,P<0.001),一致性平均差0.04。病变水平SSF2组诊断缺血性病变的AUC为0.887,优于SSF1组的0.795(P<0.001)。
要点
- SSF2运动校正算法可显著提高CCTA图像质量,整体质量评分中位数3.67 vs STD和SSF1组3(P<0.001)。
- SSF2组CT-FFR与有创FFR相关性最佳(r=0.652,P<0.001),一致性平均差0.04。
- 病变水平SSF2组诊断缺血性病变AUC 0.887,优于SSF1组0.795(P<0.001)。
核心结论
- SSF2算法显著改善了CCTA图像质量,并提升了评估狭窄严重程度和计算CT-FFR的诊断性能。
研究设计与样本
- 回顾性纳入151例在3个月内接受CCTA和ICA/FFR检查的患者,CCTA图像经迭代重建后分别用SSF1和SSF2算法处理,分为STD、SSF1和SSF2三组。
诊断标准与关键数值
- 阻塞性狭窄定义为左主干直径狭窄≥50%或其他心外膜血管直径狭窄≥70%;缺血定义为FFR≤0.8或ICA显示直径狭窄≥90%。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: This study aimed to investigate the diagnostic performance of computed tomography (CT) fractional flow reserve (CT-FFR) derived from standard images (STD) and images processed via first-generation (SnapShot Freeze, SSF1) and second-generation (SnapShot Freeze 2, SSF2) motion correction algorithms.
METHODS: 151 patients who underwent coronary CT angiography (CCTA) and invasive coronary angiography (ICA)/FFR within 3 months were retrospectively included. CCTA images were reconstructed using an iterative reconstruction technique and then further processed through SSF1 and SSF2 algorithms. All images were divided into three groups: STD, SSF1, and SSF2. Obstructive stenosis was defined as a diameter stenosis of ≥ 50 % in the left main artery or ≥ 70 % in other epicardial vessels. Stenosis with an FFR of ≤ 0.8 or a diameter stenosis of ≥ 90 % (as revealed via ICA) was considered ischemic. In patients with multiple lesions, the lesion with lowest CT-FFR was used for patient-level analysis.
RESULTS: The overall quality score in SSF2 group (median = 3.67) was markedly higher than that in STD (median = 3) and SSF1 (median = 3) groups (P < 0.001). The best correlation (r = 0.652, P < 0.001) and consistency (mean difference = 0.04) between the CT-FFR and FFR values were observed in the SSF2 group. At the per-lesion level, CT-FFRSSF2 outperformed CT-FFRSSF1 in diagnosing ischemic lesions (area under the curve = 0.887 vs. 0.795, P < 0.001). At the per-patient level, the SSF2 group also demonstrated the highest diagnostic performance.
CONCLUSION: The SSF2 algorithm significantly improved CCTA image quality and enhanced its diagnostic performance for evaluating stenosis severity and CT-FFR calculations.
引用
- Yang W, Yu L, Yu Y, et al. Novel motion correction algorithm improves diagnostic performance of CT fractional flow reserve. Eur J Radiol. 2024. PMID: 38838412. doi:10.1016/j.ejrad.2024.111538
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。