光子计数CT与双源CT在经导管主动脉瓣植入术规划中的比较

Photon-Counting Versus Dual-Source CT for Transcatheter Aortic Valve Implantation Planning.
PubMed 文献 · Acad Radiol · 2024-06-21

回顾性研究纳入300例主动脉瓣狭窄的TAVI候选者,比较光子计数CT与双源CT的图像质量、TAVI规划适用性及辐射剂量。光子计数CT的SNR和CNR较低,但视觉图像质量和规划适用性更高,有效辐射剂量更低(8.8 mSv vs 15.3 mSv)。

要点

  • 光子计数CT视觉图像质量和TAVI规划适用性评分均高于双源CT(4.8 vs 3.3,P < .001),79.2%的光子计数CT图像评为「优秀」,双源CT仅5%。
  • 光子计数CT在99%的病例中可显示主动脉瓣铰链点并描绘股动脉入路,双源CT为85%(P < 0.01)。
  • 光子计数CT有效辐射剂量低于双源CT(8.8 ± 4.5 mSv vs 15.3 ± 5.8 mSv,P < .001)。

核心结论

  • 与当前最先进的双源CT相比,光子计数CT可改善图像质量、有效避免用于TAVI规划的非诊断性CT检查,并与更低的辐射剂量相关;放射科和心脏科医生认为光子计数CT图像更适合TAVI规划。

研究设计与样本

  • 回顾性研究纳入2022年1月至2023年7月间连续接受对比增强主动脉-髂-股动脉光子计数CT和/或双源CT的主动脉瓣狭窄TAVI候选者,共300例患者(136例女性,中位年龄81岁,IQR 76-84)行302次CT检查,其中光子计数CT 202次、双源CT 100次,2例患者两种检查均接受。

关键影像表现/征象

  • 光子计数CT的SNR和CNR显著低于双源CT(33.0 ± 10.5 vs 47.3 ± 16.4 和 47.3 ± 14.8 vs 59.3 ± 21.9,P < .001)。
  • 光子计数CT视觉图像质量评分高于双源CT(4.8 vs 3.3,P < .001),图像质量评为「优秀」的比例为光子计数CT 160/202(79.2%)vs 双源CT 5/100(5%)。
  • 读者认为图像适合显示主动脉瓣铰链点和描绘股动脉入路在光子计数CT为99%、双源CT为85%(P < 0.01),适用性评分光子计数CT显著高于双源CT(4.8 vs 3.3,P < .001)。

诊断标准与关键数值

  • 图像质量和TAVI规划适用性由四名独立专家读者(两名心脏放射科医生、两名心脏科医生)采用5分制评估。
  • 放射科医生之间和心脏科医生之间的总体读者间一致性均为中等(κ = 0.60)。
  • 有效辐射剂量由CTDI和DLP计算:光子计数CT平均CTDI 22.4、DLP 519.4、eRD 8.8 ± 4.5 mSv;双源CT平均CTDI 62.9、DLP 895.5、eRD 15.3 ± 5.8 mSv(均P < .001)。

临床应用与阅片要点

  • 光子计数CT在TAVI规划中显示主动脉瓣铰链点和股动脉入路方面适用性更高、辐射剂量更低,放射科和心脏科医生认为其图像更适合TAVI规划。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Cardiovascular CT is required for planning transcatheter aortic valve implantation (TAVI).

PURPOSE: To compare image quality, suitability for TAVI planning, and radiation dose of photon-counting CT (PCCT) with that of dual-source CT (DSCT).

MATERIAL AND METHODS: Retrospective study on consecutive TAVI candidates with aortic valve stenosis who underwent contrast-enhanced aorto-ilio-femoral PCCT and/or DSCT between 01/2022 and 07/2023. Signal-to-noise (SNR) and contrast-to-noise ratio (CNR) were calculated by standardized ROI analysis. Image quality and suitability for TAVI planning were assessed by four independent expert readers (two cardiac radiologists, two cardiologists) on a 5-point-scale. CT dose index (CTDI) and dose-length-product (DLP) were used to calculate effective radiation dose (eRD).

RESULTS: 300 patients (136 female, median age: 81 years, IQR: 76-84) underwent 302 CT examinations, with PCCT in 202, DSCT in 100; two patients underwent both. Although SNR and CNR were significantly lower in PCCT vs. DSCT images (33.0 ± 10.5 vs. 47.3 ± 16.4 and 47.3 ± 14.8 vs. 59.3 ± 21.9, P < .001, respectively), visual image quality was higher in PCCT vs. DSCT (4.8 vs. 3.3, P < .001), with moderate overall interreader agreement among radiologists and among cardiologists (κ = 0.60, respectively). Image quality was rated as "excellent" in 160/202 (79.2%) of PCCT vs. 5/100 (5%) of DSCT cases. Readers found images suitable to depict the aortic valve hinge points and to map the femoral access path in 99% of PCCT vs. 85% of DSCT (P < 0.01), with suitability ranked significantly higher in PCCT vs. DSCT (4.8 vs. 3.3, P < .001). Mean CTDI and DLP, and thus eRD, were significantly lower for PCCT vs. DSCT (22.4 vs. 62.9; 519.4 vs. 895.5, and 8.8 ± 4.5 mSv vs. 15.3 ± 5.8 mSv; all P < .001).

CONCLUSION: PCCT improves image quality, effectively avoids non-diagnostic CT imaging for TAVI planning, and is associated with a lower radiation dose compared to state-of-the-art DSCT. Radiologists and cardiologists found PCCT images more suitable for TAVI planning.