四维心腔内超声在左心耳手术中的逐步推荐操作流程
本文基于初始临床经验,提出四维心腔内超声(4D ICE)配合NUVISION™ 4D ICE导管与GE E95及S70N超声系统用于左心耳封堵术的优化逐步操作流程,并在1例左心耳封堵术中与经食管超声心动图(TEE)比较:ICE左心耳测量值与TEE相似,二维及多切面模式图像分辨率最佳,FlexiSlice多平面重建图像质量略低,获得可比图像且ICE导管重新定位需求更少。
要点
- 4D ICE优化流程可用于左心耳封堵术,左心耳测量值与TEE相似。
- 二维及多切面(三平面、双平面)模式图像分辨率最佳。
- FlexiSlice多平面重建可提供有价值信息但图像质量略低。
核心结论
- 采用优化的4D ICE导管操作流程可实现高效左心耳封堵术,图像分辨率更高且与TEE相当。
研究设计与样本
- 单例左心耳封堵术病例比较4D ICE与TEE;基于初始临床经验制定NUVISION™ 4D ICE导管配合GE E95和S70N超声系统的优化工作流程。
关键影像表现/征象
- ICE左心耳测量值与TEE相似。
- 二维及多切面模式(三平面和双平面)下ICE图像分辨率最佳。
- FlexiSlice多平面重建工具从4D容积数据生成正面图像,提供有价值信息但图像质量略低。
- 该病例中TEE与ICE获得可比图像,但ICE导管重新定位需求更少。
临床应用与阅片要点
- 采用优化的4D ICE导管工作流程可高效进行左心耳封堵术。
- 4D ICE可提供与TEE相当的图像,且可能无需全身麻醉。
- 二维及多切面模式图像分辨率最佳,FlexiSlice重建图像质量略低但仍有价值。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
INTRODUCTION: Four-dimensional (4D) intracardiac echocardiography (ICE) is a novel cardiac imaging modality that has been applied to various workflows, including catheter ablation, tricuspid valve repair, and left atrial appendage occlusion (LAAO). The use of this type of advanced ICE imaging may ultimately allow for the replacement of transesophageal echocardiography (TEE) for LAAO, providing comparable imaging quality while eliminating the need for general anesthesia.
METHODS: Based on our initial clinical experience with 4D ICE in LAAO, we have developed an optimized workflow for the use of the NUVISION™ 4D ICE Catheter in conjunction with the GE E95 and S70N Ultrasound Systems in LAAO. In this manuscript, we provide a step-by-step guide to using 4D ICE in conjunction with compatible imaging consoles. We have also evaluated the performance of 4D ICE with the NUVISION Ultrasound Catheter versus TEE in one LAAO case and present those results here.
RESULTS: In our comparison of 4D ICE using our optimized workflow with TEE in an LAAO case, ICE LAA measurements were similar to those from TEE. The best image resolution was seen via ICE in 2-dimensional and multislice modes (triplane and biplane). The FlexiSlice multiplanar reconstruction tool, which creates an en-face image derived from a 4D volume set, also provided valuable information but yielded slightly lower image quality, as expected for these volume-derived images. For this case, comparable images were obtained with TEE and ICE but with less need to reposition the ICE catheter.
CONCLUSION: The use of optimized 4D ICE catheter workflow recommendations allows for efficient LAAO procedures, with higher resolution imaging, comparable to TEE.
引用
- Gidney B, Della Rocca DG, Horton R, et al. Step-by-step recommendations utilizing four-dimensional intracardiac echocardiography in left atrial appendage procedures. J Cardiovasc Electrophysiol. 2024. PMID: 38887842. doi:10.1111/jce.16309
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。