双源光子计数探测器CT同时定量脂肪与铁含量:体模研究

Using Dual-source Photon-counting Detector CT to Simultaneously Quantify Fat and Iron Content: A Phantom Study.
PubMed 文献 · Acad Radiol · 2024-05-20

本研究用双源光子计数探测器CT扫描纯脂肪、纯铁及脂肪-铁沉积体模,采用铁特异性三物质分解算法生成虚拟非铁与虚拟铁含量图像,并在不同管电压、图像质量设置及量子迭代重建强度下评估定量准确性。结果显示已知脂肪含量与虚拟非铁值、已知铁含量与虚拟铁值均呈显著线性相关,且脂肪与铁的定量互不受共存成分干扰,不同扫描参数间无显著差异。作者结论为光子计数探测器CT可准确同时定量脂肪与铁,包括较低辐射剂量的扫描参数。

要点

  • 纯脂肪体模中已知脂肪含量与虚拟非铁值呈显著线性相关(r=0.981-0.999,p<0.001)。
  • 纯铁体模中已知铁含量与虚拟铁值呈显著线性相关(r=0.897-0.975,p<0.001)。
  • 脂肪-铁体模中脂肪定量不受75、150、225 µmol/g铁过载影响(p=0.174-0.519)。

核心结论

  • 光子计数探测器CT可准确同时定量脂肪与铁含量,包括采用较低辐射剂量的扫描参数。

研究设计与样本

  • 体模研究,采用纯脂肪、纯铁及脂肪-铁沉积体模,经双源光子计数探测器CT以120和140 kV两种管电压及80和145两种图像质量设置扫描。
  • 使用铁特异性三物质分解算法生成虚拟非铁与虚拟铁含量图像,并在量子迭代重建强度1-4级下评估。

关键影像表现/征象

  • 纯脂肪体模中已知脂肪含量与虚拟非铁值呈显著线性相关(r=0.981-0.999,p<0.001)。
  • 纯铁体模中已知铁含量与虚拟铁值呈显著线性相关(r=0.897-0.975,p<0.001)。
  • 脂肪-铁体模中脂肪含量5-30%时脂肪含量与虚拟非铁值呈良好线性(r=0.919-0.990,p<0.001),且虚拟非铁值不受75、150、225 µmol/g铁过载影响(p=0.174-0.519)。
  • 脂肪-铁体模中铁定量在75-225 µmol/g范围呈线性(r=0.961-0.994,p<0.001),虚拟铁值不受共存5%、20%、30%脂肪沉积干扰(p=0.943-0.999)。
  • 脂肪与铁测量的Bland-Altman分析在不同管电压和图像质量设置间无显著差异(均p>0.05),量子迭代重建1-4级间虚拟非铁值与虚拟铁值无显著差异。

诊断标准与关键数值

  • 纯脂肪体模脂肪含量与虚拟非铁值线性相关r=0.981-0.999(p<0.001)。
  • 纯铁体模铁含量与虚拟铁值线性相关r=0.897-0.975(p<0.001)。
  • 脂肪-铁体模脂肪含量5-30%时脂肪含量与虚拟非铁值线性相关r=0.919-0.990(p<0.001)。
  • 脂肪-铁体模铁定量线性范围75-225 µmol/g,铁含量与虚拟铁值相关r=0.961-0.994(p<0.001)。

临床应用与阅片要点

  • 虚拟非铁与虚拟铁值在不同管电压、图像质量设置及量子迭代重建强度1-4级间无显著差异,提示可在较低辐射剂量参数下同时定量脂肪与铁。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To evaluate the feasibility of using photon-counting detector computed tomography (PCD CT) to simultaneously quantify fat and iron content MATERIALS AND METHODS: Phantoms with pure fat, pure iron and fat-iron deposition were scanned by two tube voltages (120 and 140 kV) and two image quality (IQ) settings (80 and 145). Using an iron-specific three-material decomposition algorithm, virtual noniron (VNI) and virtual iron content (VIC) images were generated at quantum iterative reconstruction (QIR) strength levels 1-4.

RESULTS: Significant linear correlations were observed between known fat content (FC) and VNI for pure fat phantoms (r = 0.981-0.999, p < 0.001) and between known iron content (IC) and VIC for pure iron phantoms (r = 0.897-0.975, p < 0.001). In fat-iron phantoms, the measurement for fat content of 5-30% demonstrated good linearity between FC and VNI (r = 0.919-0.990, p < 0.001), and VNI were not affected by 75, 150, and 225 µmol/g iron overload (p = 0.174-0.519). The measurement for iron demonstrated a linear range of 75-225 µmol/g between IC and VIC (r = 0.961-0.994, p < 0.001) and VIC was not confounded by the coexisting 5%, 20%, and 30% fat deposition (p = 0.943-0.999). The Bland-Altman of fat and iron measurements were not significantly different at varying tube voltages and IQ settings (all p > 0.05). No significant difference in VNI and VIC at QIR 1-4.

CONCLUSION: PCD CT can accurately and simultaneously quantify fat and iron, including scan parameters with lower radiation dose.

来源:Acad Radiol,查看原文。版权归原作者所有。