术前定量磁敏感图检测颈内动脉狭窄患者颈动脉内膜切除术前组织学斑块内出血的诊断准确性

Diagnostic Accuracy of Preoperative Quantitative Susceptibility Mapping for Detecting Histologic Intraplaque Hemorrhage in Cervical ICA Stenosis in Patients Undergoing Carotid Endarterectomy.
PubMed 文献 · AJNR Am J Neuroradiol · 2024-10-03

本研究以组织学为参考标准,比较术前定量磁敏感图(QSM)与常规T1加权3D-FSE序列检测16例颈内动脉狭窄患者斑块内出血(IPH)的诊断准确性。QSM磁敏感度与组织学IPH相对面积的相关性(ρ=0.691)显著高于T1WI相对信号强度(ρ=0.413;P=0.0259),检测以IPH为主的组织学切片(相对面积>40.7%)的AUC在QSM为0.964、T1WI FSE为0.811。结论认为术前QSM检测组织学IPH较T1WI 3D-FSE更准确。

要点

  • QSM磁敏感度与组织学IPH相对面积的相关性(ρ=0.691)显著高于T1WI相对信号强度(ρ=0.413;P=0.0259)。
  • 检测以IPH为主的组织学切片(相对面积>40.7%)的AUC:QSM 0.964,T1WI FSE 0.811。
  • QSM与以IPH为主的组织学切片间存在边际同质性(P=0.0412),T1WI相对信号强度则无(P=0.1824)。

核心结论

  • 术前QSM检测颈内动脉狭窄的组织学IPH较术前T1加权3D-FSE成像更准确。

研究设计与样本

  • 以组织学为参考标准,对16例颈内动脉狭窄患者在颈动脉内膜切除术前采集颈动脉T1加权3D-FSE和QSM图像,在包括最严重狭窄处的3个轴位层面上分别测量颈内动脉相对信号强度和磁敏感度。

关键影像表现/征象

  • 在T1加权3D-FSE和QSM图像上分别测量颈内动脉相对信号强度和磁敏感度,测量层面包括最严重狭窄处。
  • 对颈动脉内膜切除术切除的3个与MR图像对应的颈动脉斑块横断面行H&E、抗血型糖蛋白A抗体和普鲁士蓝染色,计算组织学IPH相对面积。

诊断标准与关键数值

  • 以组织学IPH相对面积>40.7%定义以IPH为主的组织学切片。
  • 检测以IPH为主的组织学切片(相对面积>40.7%)的AUC:QSM 0.964,T1WI FSE相对信号强度0.811。
  • 磁敏感度与组织学IPH相对面积的相关系数ρ=0.691,相对信号强度与组织学IPH相对面积的相关系数ρ=0.413(P=0.0259)。
  • QSM与以IPH为主的组织学切片间边际同质性P=0.0412;T1WI FSE相对信号强度与以IPH为主的组织学切片间边际同质性P=0.1824。

临床应用与阅片要点

  • 作者认为术前QSM检测颈内动脉狭窄的组织学IPH较术前T1加权3D-FSE成像更准确。
  • 原文未据此给出具体阅片或报告推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Quantitative susceptibility mapping has been proposed to assess intraplaque hemorrhage (IPH) in the carotid artery. The purpose of this study was to compare the diagnostic accuracy of preoperative quantitative susceptibility mapping with that of the conventional T1-weighed 3D-FSE sequence for detecting IPH in cervical ICA stenosis in patients undergoing carotid endarterectomy by using histology as the reference standard.

MATERIALS AND METHODS: Carotid T1-weighted 3D-FSE and QSM images were obtained from 16 patients with cervical ICA stenosis before carotid endarterectomy. Relative signal intensity and susceptibility of the ICA were measured on 3 axial images, including the location of most severe stenosis on T1-weighted 3D-FSE and quantitative susceptibility mapping images, respectively. Three transverse sections of carotid plaques excised by carotid endarterectomy, which corresponded with images on MR imaging, were stained with H&E, antibody against glycophorin A, and Prussian blue, and the relative area of histologic IPH was calculated.

RESULTS: The correlation coefficient was significantly greater between susceptibility and relative area-histologic IPH (ρ = 0.691) than between relative signal intensity and relative area-histologic IPH (ρ = 0.413; P = .0259). The areas under the receiver operating characteristic curves for detecting histologic sections consisting primarily of IPH (relative area-histologic IPH > 40.7%) tended to be greater for susceptibility (0.964) than for T1WI FSE-relative signal intensity (0.811). Marginal homogeneity was observed between susceptibility and histologic sections consisting primarily of IPH (P = .0412), but not between T1-weighted FSE-relative signal intensity and histologic sections consisting primarily of IPH (P = .1824).

CONCLUSIONS: Pre-carotid endarterectomy quantitative susceptibility mapping detects histologic IPH in cervical ICA stenosis more accurately than preoperative T1-weighted 3D-FSE imaging.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Oomori D, Akamatsu Y, Uwano I, et al. Diagnostic Accuracy of Preoperative Quantitative Susceptibility Mapping for Detecting Histologic Intraplaque Hemorrhage in Cervical ICA Stenosis in Patients Undergoing Carotid Endarterectomy. AJNR Am J Neuroradiol. 2024. PMID: 38789122. doi:10.3174/ajnr.A8356