文献速递外周血管技术与安全US

超声相干平面波复合用于高局部脉搏波速度估计的最佳参数:体模验证与10名受试者可行性

Optimum Parameters in Ultrasound Coherent Plane Wave Compounding for High LPWV Estimation: Validation on Phantom and Feasibility in 10 Subjects.
PubMed 文献 · J Ultrasound Med · 2024-05-03

本研究旨在确定基于相干平面波复合(CPWC)的高局部脉搏波速度(LPWV)估计中倾斜平面波数量与发射角度的最佳精细值。在LPWV为13.42±0.90 m/s的颈动脉体模上,采用3、5、7、9个倾斜平面波序列(角度间隔粗调步长1°,细调步长0.25°)采集射频数据,并招募两个年龄组的10名健康受试者评估体内可行性。结果显示,3个平面波(PRF 12 kHz,帧率4000 Hz)且间隔0.75°时,体模LPWV为13.52±0.08 m/s(最低CV 1.84%),10名受试者LPWV为5.49±1.46 m/s(最低CV 12.35%)。

要点

  • 3个倾斜平面波、角度间隔0.75°时LPWV测量重复性最佳。
  • 体模LPWV 13.52±0.08 m/s,变异系数1.84%。
  • 10名健康受试者LPWV 5.49±1.46 m/s,变异系数12.35%。

核心结论

  • 本研究所确定的最佳参数在体模和10名健康受试者中LPWV测量重复性最佳,支持在更大数据集上进一步研究其潜在应用。

研究设计与样本

  • 体模验证与在体可行性研究;使用Verasonics系统、L14-5/38线阵探头(128阵元)和CompuFlow1000搏动泵,在LPWV为13.42±0.90 m/s的颈动脉体模上采集射频数据,并招募两个年龄组的10名健康受试者。

诊断标准与关键数值

  • 体模参考LPWV为13.42±0.90 m/s。
  • 3个平面波(PRF 12 kHz,帧率4000 Hz)且间隔0.75°时,体模LPWV为13.52±0.08 m/s,CV最低为1.84%。
  • 相同参数下10名受试者LPWV为5.49±1.46 m/s,CV最低为12.35%。

临床应用与阅片要点

  • 作者认为所确定的最佳参数在体模和10名健康受试者中重复性最佳,支持在更大数据集上进一步研究其潜在应用。

局限性

  • 原文指出需在更大数据集上进一步研究以支持潜在应用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study is to determine the optimum and fine values of the number and transmission angles of tilted plane waves for coherent plane-wave compounding (CPWC)-based high local pulse wave velocity (LPWV) estimation.

METHODS: A Verasonics system incorporating a linear array probe L14-5/38 with 128 elements and a pulsatile pump, CompuFlow1000, were used to acquire radio frequency data of 3, 5, 7, and 9 tilted plane wave sequences with angle intervals from 0° to 12° with a coarse interval increment step of 1°, and the angle intervals from 0° to 2° with a fine interval increment step of 0.25° from a carotid vessel phantom with the LPWV of 13.42 ± 0.90 m/s. The mean value, standard deviation, and coefficients of variation (CV) of the estimated LPWVs were calculated to quantitatively assess the performance of different configurations for CPWC-based LPWV estimation. Ten healthy human subjects of two age groups were recruited to assess the in vivo feasibility of the optimum parameter values.

RESULTS: The CPWC technique with three plane waves (PRF of 12 kHz corresponding to a frame rate of 4000 Hz) with an interval of 0.75° had LPWVs of 13.52 ± 0.08 m/s with the lowest CV of 1.84% on the phantom, and 5.49 ± 1.46 m/s with the lowest CV of 12.35% on 10 subjects.

CONCLUSIONS: The optimum parameters determined in this study show the best repeatability of the LPWV measurements with a vessel phantom and 10 healthy subjects, which support further studies on larger datasets for potential applications.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Xiong L, Zhang Y, He B, et al. Optimum Parameters in Ultrasound Coherent Plane Wave Compounding for High LPWV Estimation: Validation on Phantom and Feasibility in 10 Subjects. J Ultrasound Med. 2024. PMID: 38700113. doi:10.1002/jum.16469

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。