膝关节软骨厚度的变异:对五家厂商MRI进行全自动三维分析
研究纳入10名22-60岁健康志愿者,使用五家厂商的3特斯拉MRI全自动定量七个区域软骨厚度,Kruskal-Wallis检验显示七个区域P值均大于0.05,各厂商设备间无显著差异,350次测量中测量间误差≤0.20 mm者占95%,最大误差源于滑液被误提取为软骨。
要点
- 五家厂商3T MRI全自动测量膝关节软骨厚度,七个区域组间比较P值均大于0.05。
- 350次测量中测量间误差≤0.05 mm占53%、≤0.10 mm占75%、≤0.20 mm占95%。
- 内侧胫骨软骨测量间误差最大,原因为滑液被误提取为软骨。
核心结论
- MRI机型的选择不影响软骨厚度测量,研究假设被拒绝。
- 总体上95%的测量间误差在0.20 mm以内,最大误差由滑液被误提取为软骨所致。
研究设计与样本
- 研究对象为10名22-60岁健康志愿者。
- 使用五家不同公司的3特斯拉MRI设备。
- 对七个区域进行全自动软骨厚度定量,以五台MRI机型所测厚度的中位数与目标厚度之差的绝对值定义测量间误差并用直方图分析。
关键影像表现/征象
- 未观察到可归因于特定机型仪器的异常趋势。
- 内侧胫骨软骨的测量间误差最大,分析显示原因为滑液被误提取为软骨。
诊断标准与关键数值
- 七个区域Kruskal-Wallis检验的P值均大于0.05。
- 350次测量中测量间误差≤0.05 mm占53%、≤0.10 mm占75%、≤0.20 mm占95%。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Measurements of knee cartilage thickness derived from MR images are attractive biomarkers for osteoarthritis research. Although some cross-sectional multivendor studies exist, none have employed fully automatic three-dimensional MRI analysis. Our objective was to evaluate the variations in knee cartilage thickness measurements obtained using automated methods and MRI instruments from five different vendors.
METHODS: The subjects were 10 healthy volunteers aged 22-60 years. MRI models with 3 Tesla strength from five different companies were used. Cartilage thickness was quantified fully automatically for seven regions. We hypothesized that "the MRI model influences cartilage thickness measurements." Inter-measurement error, defined as the absolute difference between the targeted and median thicknesses determined by the five MRI models, was analyzed using histograms. The factors generating the largest inter-measurement error were also examined.
RESULTS: No exceptional trends attributable to a specific instrument model were observed, and the p-value from the Kruskal-Wallis test exceeded 0.05 in all seven regions. Therefore, the study hypothesis was rejected. Of the 350 measurements, the inter-measurement error was ≤0.05 mm in 53 %, ≤0.10 mm in 75 %, and ≤0.20 mm in 95 %. Analysis of the medial tibial cartilage, which had the largest inter-measurement error, revealed mis-extraction of synovial fluid as cartilage.
CONCLUSIONS: The choice of MRI model did not influence cartilage thickness measurements. Overall, 95 % of the inter-measurement errors were within 0.20 mm. The greatest error resulted from mis-extracting synovial fluid as cartilage.
引用
- Katano H, Nagai K, Kaneko H, et al. Variations in knee cartilage thickness: Fully automatic three-dimensional analysis of MRIs from five manufacturers. Eur J Radiol. 2024. PMID: 38815306. doi:10.1016/j.ejrad.2024.111528
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。