文献速递妇产影像技术与安全US

自动多体积胎盘重建:三维能量多普勒超声与红外相机追踪

Automated multivolume placental reconstruction using three-dimensional power Doppler ultrasound and infrared camera tracking.
PubMed 文献 · Ultrasound Obstet Gynecol · 2025-04-03

本研究在仿体中验证了利用红外相机追踪换能器空间坐标、将多个三维能量多普勒超声体积自动拼接重建整个胎盘模型的准确性,并在17例中晚孕期单胎妊娠中评估临床可行性:胎盘体积重建成功率92%,每例至少获得一个完整体积(100%),重建胎盘血管树定性评价完整连续,全胎盘分割体积测量重复性高(ICC 0.96,95%CI 0.89-0.99)。

要点

  • 仿体研究中拼接后三维超声测量值与真实值平均相差1.51 mm(范围-4.78至4.23 mm)。
  • 临床可行性研究中胎盘体积重建成功率92%,每例至少获得一个完整体积(100%)。
  • 全胎盘分割体积测量重复性高,组内相关系数0.96(95%CI 0.89-0.99)。

核心结论

  • 作者提出一种利用多个三维能量多普勒超声体积自动重建中晚孕期胎盘整体结构及血管树的方法,仿体验证准确、临床可行,为开发胎盘功能不全筛查工具奠定基础,并可能扩展至成人器官灌注评估。

研究设计与样本

  • 研究分为仿体准确性验证和单中心横断面可行性研究两部分;仿体为模拟孕晚期胎盘尺寸的丙烯酸模型,内含12个不同高度长方体柱并浸于组织模拟液中;临床部分纳入17例无并发症中晚孕期单胎妊娠孕妇,采用标准化超声设置采集多个三维能量多普勒及灰阶胎盘体积,并用红外相机追踪坐标进行自动拼接。

关键影像表现/征象

  • 仿体在三种不同声束入射角度下各重建6个完整数据集,共18个扩展数据集,每个完整仿体数据集重建所需三维超声体积中位数为9个。
  • 648个评估数据点中67.1%完全对齐,局部距离误差均值2.92 mm(范围0-25.51 mm)。
  • 每个拼接三维超声体积测量120个柱间距离(共2160个距离),与仿体模型真实测量值平均相差1.51 mm(范围-4.78至4.23 mm)。
  • 临床可行性研究中17例受试者共采集49个三维超声体积数据集,胎盘体积集重建成功率92%,每例至少获得一个完整体积(100%)。
  • 重建胎盘血管树定性评价为全程存在、连续且细节丰富;体积采集与重建时间中位数10分钟。

诊断标准与关键数值

  • 重建准确性以局部距离误差(重叠三维超声体积中柱体对齐评估拼接准确性)和全局距离误差(拼接三维超声体积测量值与仿体模型真实测量值之差)评价。
  • 全胎盘分割体积测量的重复性以组内相关系数评价,为0.96(95%CI 0.89-0.99)。

临床应用与阅片要点

  • 本方法采用红外相机追踪换能器精确空间位置获得全局坐标,将多个重叠三维能量多普勒超声体积自动拼接为整个胎盘模型。
  • 临床可行性研究使用标准化超声设置采集多个三维能量多普勒及灰阶胎盘体积,拼接后对胎盘血管树进行定性评估及全胎盘分割体积测量。

局限性

  • 原文结论指出本研究为开发检测胎盘功能不全的实用筛查工具奠定基础,并可能扩展至成人器官灌注评估,未报告其他局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Placental insufficiency contributes to many obstetric pathologies, but there is no bedside clinical tool to evaluate placental perfusion. We have developed a method to acquire multiple three-dimensional (3D) power Doppler (PD) ultrasound (US) volumes of placental vasculature, with infrared camera tracking of the precise spatial location of the transducer providing global coordinates. These volumes are reconstructed automatically ('stitched') into a model of the entire placenta. The purpose of this study was to evaluate the accuracy of automated reconstruction in an US phantom and to assess the feasibility of this technique in second-to-third-trimester human placentae.

METHODS: A custom-designed acrylic phantom was constructed with dimensions mimicking a third-trimester placenta, containing 12 rectangular cuboid towers of various heights submersed in tissue-mimicking solution. Multiple overlapping 3D-US volumes of this phantom were acquired using three different insonation angles and infrared camera tracking. Data were transformed into a 3D cartesian volume and stitched automatically into six 3D-US volumes, each covering the entire phantom, for each of the three different insonation angles. Reconstruction accuracy was evaluated by calculating local distance error (assessment of towers in overlapping US volumes to determine accuracy of stitching) and global distance error (subtraction of true measurements in phantom model from corresponding measurements in stitched 3D-US volumes). A single-center, cross-sectional feasibility study was then conducted in women with an uncomplicated second-to-third-trimester singleton pregnancy, with data obtained using standardized ultrasound settings. Multiple 3D PD-US and grayscale volumes of the placentae were acquired with infrared camera-tracked coordinates. Volumes were stitched to create a model of placental vasculature, and these were assessed for quality and repeatability of volume measurement.

RESULTS: Six entire phantom datasets were reconstructed at each of three insonation angles, giving a total of 18 extended phanom datasets. A median of nine 3D-US volumes required to reconstruct the entire phantom. Twelve towers per volume were assessed on three separate occasions, generating 648 datapoints. Of these datapoints, 67.1% were perfectly aligned. The mean local distance error was 2.92 (range, 0-25.51) mm. Measurements between towers of 120 distances in each stitched 3D-US volume (2160 distances in total) differed by an average of 1.51 (range, -4.78 to 4.23) mm from the true measurements in the phantom model. In the feasibility study, 17 participants were scanned, and 49 3D-US volume datasets acquired, with 92% reconstruction success per placental volume set and at least one complete volume being obtained per participant (100% participant achievability). The median volume acquisition and reconstruction time was 10 min. Reconstructed placental vasculature was assessed qualitatively to be present, continuous and detailed throughout. Volume measurement of entire segmented placentae was highly repeatable (intraclass correlation coefficient, 0.96 (95% CI, 0.89-0.99)).

CONCLUSION: We present an automated method to model the entire structure and vasculature of second-to-third-trimester placentae using multiple 3D PD-US volumes, with verified accuracy and clinical feasibility. This study builds a foundation on which to develop a practical screening tool for detecting placental insufficiency and has the potential for expansion to evaluation of adult organ perfusion. © 2024 The Author(s). Ultrasound in Obstetrics & Gynecology published by John Wiley & Sons Ltd on behalf of International Society of Ultrasound in Obstetrics and Gynecology.

来源:Ultrasound Obstet Gynecol,查看原文。版权归原作者所有。

引用

  1. Xue A, Hanly R, Luichareonkit D, et al. Automated multivolume placental reconstruction using three-dimensional power Doppler ultrasound and infrared camera tracking. Ultrasound Obstet Gynecol. 2025. PMID: 38788047. doi:10.1002/uog.27708

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。