超快三维经皮超分辨超声:基于行列阵列特定相干波束形成与滚动声学子孔径处理的体外、兔及人体研究

Ultrafast 3-D Transcutaneous Super Resolution Ultrasound Using Row-Column Array Specific Coherence-Based Beamforming and Rolling Acoustic Sub-aperture Processing: In Vitro, in Rabbit and in Human Study.
PubMed 文献 · Ultrasound Med Biol · 2024-05-03

本研究开发了一套结合行列阵列特定相干波束形成与滚动声学子孔径处理的图像重建流程,用于经皮三维超分辨超声成像;体外实验显示该流程将虚假定位点比例从约26%降至约15%,噪声降低约7 dB,有效帧率提升至超过4000 fps;在兔肾和人体甲状腺中非侵入性地获得了大视野超分辨微血管图像。

要点

  • 滚动窗口重建结合相干波束形成与声学子孔径处理,可降低行列阵列旁瓣伪影与噪声。
  • 体外交叉管实验中虚假定位点比例从约26%降至约15%,噪声降低约7 dB。
  • 有效帧率提升至超过4000 fps,支持在兔肾和人体甲状腺经皮获取超分辨微血管图像。

核心结论

  • 所提出的方法使用行列阵列可在体内及人体中非侵入性地产生大视野超分辨微血管图像。

研究设计与样本

  • 研究开发了包含滚动窗口图像重建的处理流程,整合行列阵列特定相干波束形成与声学子孔径处理,并在体外交叉管、兔模型及人体中进行验证。

关键影像表现/征象

  • 体外交叉管实验中,与正交平面波复合相比,该流程将虚假定位点比例从约26%降至约15%。
  • 噪声可降低约7 dB,有效帧率增加至超过4000 fps。
  • 在体内,超声定位显微术非侵入性地生成了兔肾和人体甲状腺的图像。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: This study aimed to realise 3-D super-resolution ultrasound imaging transcutaneously with a row-column array which has far fewer independent electronic channels and a wider field of view than typical fully addressed 2-D matrix arrays. The in vivo image quality of the row-column array is generally poor, particularly when imaging non-invasively. This study aimed to develop a suite of image formation and post-processing methods to improve image quality and demonstrate the feasibility of ultrasound localisation microscopy using a row-column array, transcutaneously on a rabbit model and in a human.

METHODS: To achieve this, a processing pipeline was developed which included a new type of rolling window image reconstruction, which integrated a row-column array specific coherence-based beamforming technique with acoustic sub-aperture processing. This and other processing steps reduced the 'secondary' lobe artefacts, and noise and increased the effective frame rate, thereby enabling ultrasound localisation images to be produced.

RESULTS: Using an in vitro cross tube, it was found that the procedure reduced the percentage of 'false' locations from ∼26% to ∼15% compared to orthogonal plane wave compounding. Additionally, it was found that the noise could be reduced by ∼7 dB and the effective frame rate was increased to over 4000 fps. In vivo, ultrasound localisation microscopy was used to produce images non-invasively of a rabbit kidney and a human thyroid.

CONCLUSION: It has been demonstrated that the proposed methods using a row-column array can produce large field of view super-resolution microvascular images in vivo and in a human non-invasively.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Hansen-Shearer J, Yan J, Lerendegui M, et al. Ultrafast 3-D Transcutaneous Super Resolution Ultrasound Using Row-Column Array Specific Coherence-Based Beamforming and Rolling Acoustic Sub-aperture Processing: In Vitro, in Rabbit and in Human Study. Ultrasound Med Biol. 2024. PMID: 38702285. doi:10.1016/j.ultrasmedbio.2024.03.020