文献速递泌尿生殖技术与安全MR

提升实践中的前列腺MR图像质量——ACR前列腺MR图像质量改进协作组初步结果

Improving Prostate MR Image Quality in Practice-Initial Results From the ACR Prostate MR Image Quality Improvement Collaborative.
PubMed 文献 · J Am Coll Radiol · 2024-05-09

美国5家机构参与ACR学习网络前列腺MR图像质量改进协作,采用结构化改进方法并以PI-QUAL评分≥4为质量达标标准。审计2380例检查,周平均达标率从基线67%升至项目完成时87%,目标为至少85%。常用干预包括MR协议调整、患者准备说明、人员培训和审计流程建设。

要点

  • 以PI-QUAL评分≥4为质量达标标准,协作组目标为达标检查比例至少85%。
  • 2380例审计检查的周平均达标率从基线67%(范围60%-74%)升至项目完成时87%(范围80%-97%)。
  • 常用干预为MR协议调整、患者准备说明制定与实施、人员培训及审计流程机制建设。

核心结论

  • 学习网络模式可使多家机构同时提升前列腺MR图像质量,首批队列的关键经验可为更广泛推广提供路线图。

研究设计与样本

  • 质量改进协作项目,美国5家机构团队接受结构化改进方法培训,以PI-QUAL系统审计图像,采用计划-执行-研究-行动循环设计并测试干预,共审计2380例检查。

诊断标准与关键数值

  • 以PI-QUAL评分≥4定义为满足图像质量标准;协作组目标为将PI-QUAL≥4的检查比例提高至至少85%。

临床应用与阅片要点

  • 各机构通过MR协议调整、患者准备说明、人员培训和审计流程机制等干预提升前列腺MR图像质量,并以运行图监测达标率、每周会议回顾进展。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Variability in prostate MRI quality is an increasingly recognized problem that negatively affects patient care. This report aims to describe the results and key learnings of the first cohort of the ACR Learning Network Prostate MR Image Quality Improvement Collaborative.

METHODS: Teams from five organizations in the United States were trained on a structured improvement method. After reaching a consensus on image quality and auditing their images using the Prostate Imaging Quality (PI-QUAL) system, teams conducted a current state analysis to identify barriers to obtaining high-quality images. Through plan-do-study-act cycles involving frontline staff, each site designed and tested interventions targeting image quality key drivers. The percentage of examinations meeting quality criteria (ie, PI-QUAL score ≥4) was plotted on a run chart, and project progress was reviewed in weekly meetings. At the collaborative level, the goal was to increase the percentage of examinations with PI-QUAL ≥4 to at least 85%.

RESULTS: Across 2,380 examinations audited, the mean weekly rates of prostate MR examinations meeting image quality criteria increased from 67% (range: 60%-74%) at baseline to 87% (range: 80%-97%) upon program completion. The most commonly employed interventions were MR protocol adjustments, development and implementation of patient preparation instructions, personnel training, and development of an auditing process mechanism.

CONCLUSION: A learning network model, in which organizations share knowledge and work together toward a common goal, can improve prostate MR image quality at multiple sites simultaneously. The inaugural cohort's key learnings provide a road map for improvement on a broader scale.

来源:J Am Coll Radiol,查看原文。版权归原作者所有。

引用

  1. Purysko AS, Zacharias-Andrews K, Tomkins KG, et al. Improving Prostate MR Image Quality in Practice-Initial Results From the ACR Prostate MR Image Quality Improvement Collaborative. J Am Coll Radiol. 2024. PMID: 38729590. doi:10.1016/j.jacr.2024.04.008

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。