文献速递腹部技术与安全CT

低管电压联合深度学习重建在薄层腹部CT中降低辐射与对比剂剂量的前瞻性临床试验

Low tube voltage and deep-learning reconstruction for reducing radiation and contrast medium doses in thin-slice abdominal CT: a prospective clinical trial.
PubMed 文献 · Eur Radiol · 2024-05-16

前瞻性纳入148例患者,比较120-kVp联合混合迭代重建(HIR)与80-kVp联合HIR或深度学习重建(DLR)的薄层腹部CT。80-kVp方案辐射剂量和碘剂量分别较120-kVp降低40.4%和36.3%,80-kVp-DLR获得最高CNR和主观评分,噪声纹理与边缘锐利度与120-kVp-HIR无显著差异。

要点

  • 80-kVp联合DLR在薄层腹部CT中提供优于120-kVp-HIR的客观与主观图像质量。
  • 80-kVp方案较120-kVp方案降低辐射剂量40.4%、碘剂量36.3%。
  • 80-kVp-DLR与120-kVp-HIR的噪声纹理和边缘锐利度无显著差异。

核心结论

  • 与120-kVp-HIR方案相比,80-kVp联合DLR在薄层腹部CT中获得更优的主观和客观图像质量,同时降低辐射剂量和碘对比剂剂量。

研究设计与样本

  • 前瞻性临床试验纳入148例接受增强腹部CT的患者,120-kVp方案(600 mgL/kg,n=74)与80-kVp方案(360 mgL/kg,n=74)各74例,层厚0.5 mm,120-kVp图像用HIR重建,80-kVp图像用HIR和DLR重建。

关键影像表现/征象

  • 80-kVp图像血管和实质器官CT值高于120-kVp图像(均p < 0.001)。
  • 80-kVp-HIR图像噪声高于120-kVp-HIR,80-kVp-DLR图像噪声低于120-kVp-HIR(均p < 0.001)。
  • 80-kVp-DLR获得最高CNR和主观评分(均p < 0.001)。
  • 120-kVp-HIR与80-kVp-DLR的平均NPS频率和ERS无显著差异(p ≥ 0.38)。

诊断标准与关键数值

  • 主观图像质量采用4分制评分,80-kVp-DLR评分最高(p < 0.001)。
  • 80-kVp方案SSDE为8.1 ± 0.9 mGy,较120-kVp的13.6 ± 2.7 mGy降低40.4%(p < 0.001)。
  • 80-kVp方案碘剂量为21.2 ± 3.9 gL,较120-kVp的33.3 ± 4.3 gL降低36.3%(p < 0.001)。

临床应用与阅片要点

  • 低管电压80-kVp联合DLR可用于薄层腹部CT,在降低辐射和碘对比剂剂量的同时获得更优图像质量。
  • 80-kVp联合DLR可能提高薄层腹部CT的诊断信心和患者安全。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To investigate the feasibility of low-radiation dose and low iodinated contrast medium (ICM) dose protocol combining low-tube voltage and deep-learning reconstruction (DLR) algorithm in thin-slice abdominal CT.

METHODS: This prospective study included 148 patients who underwent contrast-enhanced abdominal CT with either 120-kVp (600 mgL/kg, n = 74) or 80-kVp protocol (360 mgL/kg, n = 74). The 120-kVp images were reconstructed using hybrid iterative reconstruction (HIR) (120-kVp-HIR), while 80-kVp images were reconstructed using HIR (80-kVp-HIR) and DLR (80-kVp-DLR) with 0.5 mm thickness. Size-specific dose estimate (SSDE) and iodine dose were compared between protocols. Image noise, CT attenuation, and contrast-to-noise ratio (CNR) were quantified. Noise power spectrum (NPS) and edge rise slope (ERS) were used to evaluate noise texture and edge sharpness, respectively. The subjective image quality was rated on a 4-point scale.

RESULTS: SSDE and iodine doses of 80-kVp were 40.4% (8.1 ± 0.9 vs. 13.6 ± 2.7 mGy) and 36.3% (21.2 ± 3.9 vs. 33.3 ± 4.3 gL) lower, respectively, than those of 120-kVp (both, p < 0.001). CT attenuation of vessels and solid organs was higher in 80-kVp than in 120-kVp images (all, p < 0.001). Image noise of 80-kVp-HIR and 80-kVp-DLR was higher and lower, respectively than that of 120-kVp-HIR (both p < 0.001). The highest CNR and subjective scores were attained in 80-kVp-DLR (all, p < 0.001). There were no significant differences in average NPS frequency and ERS between 120-kVp-HIR and 80-kVp-DLR (p ≥ 0.38).

CONCLUSION: Compared with the 120-kVp-HIR protocol, the combined use of 80-kVp and DLR techniques yielded superior subjective and objective image quality with reduced radiation and ICM doses at thin-section abdominal CT.

CLINICAL RELEVANCE STATEMENT: Scanning at low-tube voltage (80-kVp) combined with the deep-learning reconstruction algorithm may enhance diagnostic efficiency and patient safety by improving image quality and reducing radiation and contrast doses of thin-slice abdominal CT.

KEY POINTS: Reducing radiation and iodine doses is desirable; however, contrast and noise degradation can be detrimental. The 80-kVp scan with the deep-learning reconstruction technique provided better images with lower radiation and contrast doses. This technique may be efficient for improving diagnostic confidence and patient safety in thin-slice abdominal CT.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Yoshida K, Nagayama Y, Funama Y, et al. Low tube voltage and deep-learning reconstruction for reducing radiation and contrast medium doses in thin-slice abdominal CT: a prospective clinical trial. Eur Radiol. 2024. PMID: 38753193. doi:10.1007/s00330-024-10793-6

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。