直接腹部静脉血栓成像(DATI):一种无对比剂黑血MR技术用于腹部静脉血栓诊断
前瞻性纳入10名健康志愿者和28例疑似腹部静脉血栓(AVT)患者,在3.0 T MRI上采用基于呼吸导航SPACE序列联合DANTE黑血准备的DATI扫描,疑似患者另行对比增强CT静脉成像(CTV)作为参考。DATI对AVT诊断准确率96.6%、敏感度91.5%、特异度98.0%、阳性预测值92.3%、阴性预测值97.8%,图像质量和诊断信心均为良好至优秀,无需外源性对比剂。
要点
- DATI无需对比剂即可提供足够的血栓信号强度及血栓与暗静脉腔之间的对比。
- 以CTV为参考,DATI诊断AVT的准确率96.6%、敏感度91.5%、特异度98.0%。
- DATI图像质量评分读者1为3.50±0.64、读者2为3.42±0.63,诊断信心评分读者1为3.71±0.53、读者2为3.78±0.42。
核心结论
- DATI无需外源性对比剂即可提供良好至优秀的图像质量、有效的静脉血信号抑制和明确的血栓检出,用于AVT诊断。
研究设计与样本
- 前瞻性研究,纳入10名健康受试者和28例疑似AVT患者,在3.0 T MRI上采用基于呼吸导航SPACE序列联合DANTE黑血准备的DATI扫描,疑似患者另行CTV作为参考标准,由两名盲法放射科医师独立评估随机化图像。
关键影像表现/征象
- DATI可提供足够的血栓信号强度以及血栓与暗静脉腔之间的对比,用于AVT诊断。
- DATI图像质量评分读者1为3.50±0.64、读者2为3.42±0.63,κ=0.872。
- DATI诊断信心评分读者1为3.71±0.53、读者2为3.78±0.42,κ=0.804。
诊断标准与关键数值
- 以CTV为参考标准,DATI诊断AVT的准确率96.6%、敏感度91.5%、特异度98.0%、阳性预测值92.3%、阴性预测值97.8%。
- 图像质量和诊断信心采用1分(差)至4分(优秀)评分。
临床应用与阅片要点
- DATI无需外源性对比剂即可用于疑似AVT的诊断,图像质量和诊断信心良好至优秀。
- 两名读者间图像质量(κ=0.872)和诊断信心(κ=0.804)一致性良好。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To develop and evaluate a direct abdominal vein thrombus imaging (DATI) technique, based on a respiratory navigating SPACE sequence with DANTE black-blood preparation, for diagnosing abdominal vein thrombosis (AVT) without the use of exogenous contrast agents.
METHODS: We prospectively enrolled 10 healthy subjects and 28 suspected AVT patients who underwent DATI scans on 3.0 T MRI. Contrast-enhanced CT venography (CTV) was also conducted on the suspected AVT patients for comparison. All images were analyzed by two blinded radiologists who independently evaluated randomized images and gave image quality and diagnostic confidence scores (1-poor, 4-excellent) for DATI and CTV. The accuracy (ACC), sensitivity (SE), specificity (SP), positive predictive value (PPV), and negative predictive value (NPV) of CTV were calculated using CTV as a standard reference. The diagnostic agreement between DATI and CTV as well as the interobserver agreement were conducted using Cohen κ test.
RESULTS: The patient study demonstrated that DATI can provide adequate thrombus signal intensity and the contrast between the thrombus to dark venous lumen for the diagnosis of AVT. It offers good to excellent image quality (reader1/reader2: 3.50 ± 0.64/3.42 ± 0.63, κ = 0.872) and diagnostic confidence (reader1/reader2: 3.71 ± 0.53/3.78 ± 0.42, κ = 0.804) for the diagnosis of AVT. Taking CTV as a reference, DATI has high accuracy (96.6%), SE (91.5%), SP (98.0%), PPV (92.3%), and NPV (97.8%).
DATA CONCLUSION: DATI can provide good to excellent image quality, effective venous blood signal suppression, and definitive thrombus detection for the diagnosis of AVT without the use of exogenous contrast agents.
引用
- Liao L, Chen J, Liu Z, et al. Direct abdominal vein thrombus imaging (DATI): a contrast-free black-blood MR technique for the diagnosis of abdominal vein thrombosis. Abdom Radiol (NY). 2024. PMID: 38780634. doi:10.1007/s00261-024-04335-3
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。