MR弹性成像刚度测量的精密度与重测重复性:多中心体模研究

Precision and Test-Retest Repeatability of Stiffness Measurement with MR Elastography: A Multicenter Phantom Study.
PubMed 文献 · Radiology · 2024-01-01

该多中心体模研究评估了MR弹性成像(MRE)刚度测量的精密度与重测重复性。三种模拟不同肝硬度的体模在五个中心、10台1.5T和3T系统上使用2D GRE和/或2D SE EPI序列扫描。所有体模和两种序列合并的变异系数(CV)为5.8%,2D GRE的CV为4.5%,2D SE EPI的CV为7.8%。结论认为在相似成像协议、硬件和重建算法下,MRE在多中心环境中具有优异的体外精密度和同日重测重复性。

要点

  • 所有体模和两种序列合并的刚度测量CV为5.8%(95%CI:3.8,7.7)。
  • 2D GRE序列的CV为4.5%(95%CI:3.3,5.7),2D SE EPI序列的CV为7.8%(95%CI:3.1,12.6)。
  • 所有体模和两种序列合并的RC为5.8%(95%CI:3.7,7.8),2D GRE为4.9%(95%CI:2.7,7.0),2D SE EPI为7.0%(95%CI:2.9,11.2)。

核心结论

  • 在相似成像协议、硬件和重建算法下,MRE在多中心环境中具有优异的体外精密度和同日重测重复性。

研究设计与样本

  • 多中心体模研究,使用三种聚氯乙烯凝胶圆柱体模(体模1-3分别模拟软、中、硬硬度),于2021年1月至2022年1月在五个中心间循环,使用来自三家主要厂商的10台配备MRE的临床1.5-T和3-T系统,采用2D GRE和/或2D SE EPI序列扫描,各中心使用相似的MRE采集参数、硬件和重建算法。

诊断标准与关键数值

  • 精密度以变异系数(CV)评估,同日重测重复性以重复性系数(RC)评估;所有体模和两种序列合并的CV为5.8%(95%CI:3.8,7.7),2D GRE的CV为4.5%(95%CI:3.3,5.7),2D SE EPI的CV为7.8%(95%CI:3.1,12.6);所有体模和两种序列合并的RC为5.8%(95%CI:3.7,7.8),2D GRE的RC为4.9%(95%CI:2.7,7.0),2D SE EPI的RC为7.0%(95%CI:2.9,11.2)。

临床应用与阅片要点

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background MR elastography (MRE) has been shown to have excellent performance for noninvasive liver fibrosis staging. However, there is limited knowledge regarding the precision and test-retest repeatability of stiffness measurement with MRE in the multicenter setting. Purpose To determine the precision and test-retest repeatability of stiffness measurement with MRE across multiple centers using the same phantoms. Materials and Methods In this study, three cylindrical phantoms made of polyvinyl chloride gel mimicking different degrees of liver stiffness in humans (phantoms 1-3: soft, medium, and hard stiffness, respectively) were evaluated. Between January 2021 and January 2022, phantoms were circulated between five different centers and scanned with 10 MRE-equipped clinical 1.5-T and 3-T systems from three major vendors, using two-dimensional (2D) gradient-recalled echo (GRE) imaging and/or 2D spin-echo (SE) echo-planar imaging (EPI). Similar MRE acquisition parameters, hardware, and reconstruction algorithms were used at each center. Mean stiffness was measured by a single observer for each phantom and acquisition on a single section. Stiffness measurement precision and same-session test-retest repeatability were assessed using the coefficient of variation (CV) and the repeatability coefficient (RC), respectively. Results The mean precision represented by the CV was 5.8% (95% CI: 3.8, 7.7) for all phantoms and both sequences combined. For all phantoms, 2D GRE achieved a CV of 4.5% (95% CI: 3.3, 5.7) whereas 2D SE EPI achieved a CV of 7.8% (95% CI: 3.1, 12.6). The mean RC of stiffness measurement was 5.8% (95% CI: 3.7, 7.8) for all phantoms and both sequences combined, 4.9% (95% CI: 2.7, 7.0) for 2D GRE, and 7.0% (95% CI: 2.9, 11.2) for 2D SE EPI (all phantoms). Conclusion MRE had excellent in vitro precision and same-session test-retest repeatability in the multicenter setting when similar imaging protocols, hardware, and reconstruction algorithms were used. © RSNA, 2024 Supplemental material is available for this article. See also the editorial by Tang in this issue.