光子计数探测器CT超高分辨率冠状动脉CT血管成像:量子迭代重建对图像质量及狭窄测量准确性影响的双中心研究

Ultra-high resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements.
PubMed 文献 · Eur J Radiol · 2024-05-18

研究在动态血管体模与102例患者中评估光子计数探测器CT超高分辨率冠状动脉CT血管成像不同量子迭代重建级别对图像质量及狭窄定量的影响,发现提高级别可降低噪声、提升信噪比与对比噪声比且不损害锐利度与狭窄测量准确性。

要点

  • 量子迭代重建级别升高可显著降低客观图像噪声并提高信噪比与对比噪声比(体外及体内均p<0.001)。
  • 锐利度与狭窄测量值在不同量子迭代重建级别间无显著差异(两两比较p>0.008)。
  • 体内主观噪声随级别升高显著降低、整体图像质量评分显著提高(两两比较p<0.001),主观锐利度无差异(p=0.15)。

核心结论

  • 量子迭代重建算法可提升冠状动脉CT血管成像数据集图像质量,且不损害图像锐利度或狭窄测量准确性,最高强度级别获益最明显。

研究设计与样本

  • 双中心研究:动态血管体模(含25%与50%狭窄两处钙化病变,心率60、80、100次/分)及102例患者体内冠状动脉CT血管成像,均以超高分辨率模式(层厚0.2 mm)采集并用锐利血管核(Bv64)重建四个量子迭代重建级别(1-4)。

关键影像表现/征象

  • 体模与体内图像客观噪声随量子迭代重建级别升高显著降低(均p<0.001)。
  • 信噪比与对比噪声比随级别升高显著升高(均p<0.001)。
  • 锐利度与百分直径狭窄值在不同级别间无显著差异(两两比较p>0.008)。
  • 体内主观噪声随级别升高显著降低,整体图像质量评分显著提高(两两比较p<0.001)。
  • 主观锐利度在不同级别间无差异(p=0.15)。

诊断标准与关键数值

  • 体模含25%与50%狭窄两处钙化病变,用于评估百分直径狭窄测量。
  • 图像质量指标包括噪声、信噪比、对比噪声比、锐利度及主观噪声、锐利度、整体图像质量。

临床应用与阅片要点

  • 量子迭代重建最高强度级别在提升图像质量方面获益最明显,且不影响狭窄测量准确性。
  • 不同量子迭代重建级别间锐利度及百分直径狭窄值无显著差异,提示狭窄定量结果对重建级别不敏感。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To assess the impact of different quantum iterative reconstruction (QIR) levels on objective and subjective image quality of ultra-high resolution (UHR) coronary CT angiography (CCTA) images and to determine the effect of strength levels on stenosis quantification using photon-counting detector (PCD)-CT.

METHOD: A dynamic vessel phantom containing two calcified lesions (25 % and 50 % stenosis) was scanned at heart rates of 60, 80 and 100 beats per minute with a PCD-CT system. In vivo CCTA examinations were performed in 102 patients. All scans were acquired in UHR mode (slice thickness0.2 mm) and reconstructed with four different QIR levels (1-4) using a sharp vascular kernel (Bv64). Image noise, signal-to-noise ratio (SNR), sharpness, and percent diameter stenosis (PDS) were quantified in the phantom, while noise, SNR, contrast-to-noise ratio (CNR), sharpness, and subjective quality metrics (noise, sharpness, overall image quality) were assessed in patient scans.

RESULTS: Increasing QIR levels resulted in significantly lower objective image noise (in vitro and in vivo: both p < 0.001), higher SNR (both p < 0.001) and CNR (both p < 0.001). Sharpness and PDS values did not differ significantly among QIRs (all pairwise p > 0.008). Subjective noise of in vivo images significantly decreased with increasing QIR levels, resulting in significantly higher image quality scores at increasing QIR levels (all pairwise p < 0.001). Qualitative sharpness, on the other hand, did not differ across different levels of QIR (p = 0.15).

CONCLUSIONS: The QIR algorithm may enhance the image quality of CCTA datasets without compromising image sharpness or accurate stenosis measurements, with the most prominent benefits at the highest strength level.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Vecsey-Nagy M, Varga-Szemes A, Schoepf UJ, et al. Ultra-high resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements. Eur J Radiol. 2024. PMID: 38805884. doi:10.1016/j.ejrad.2024.111517