文献速递儿科影像胸部技术与安全CT

儿童晶状体散射剂量在轴扫40毫米与螺旋扫描160毫米探测器宽度CT模式间的比较

Comparison of pediatric lens scattered dose measurements between axial 40-mm and helical 160-mm detector width computed tomography scan modes.
PubMed 文献 · Pediatr Radiol · 2024-05-21

该研究使用两种CT设备(Revolution CT和Aquilion ONE)及三种儿童体模,比较螺旋扫描(40毫米探测器宽度)与轴扫(160毫米探测器宽度)在儿童胸部CT检查中晶状体散射剂量的差异。结果显示,使用Aquilion ONE时三种体模螺旋扫描的晶状体散射剂量均显著高于轴扫;使用Revolution CT时仅5岁体模螺旋扫描显著高于轴扫,新生儿和1岁体模无显著差异。作者认为散射剂量可能随CT设备和体型变化,但轴扫通常较低。

要点

  • 使用Aquilion ONE时,螺旋扫描晶状体散射剂量在新生儿、1岁和5岁体模中分别较轴扫高约70%、40%和30%(P<0.01)。
  • 使用Revolution CT时,5岁体模螺旋扫描晶状体散射剂量较轴扫高约20-35%(P<0.01)。
  • 使用Revolution CT时,新生儿和1岁体模螺旋与轴扫的晶状体散射剂量无显著差异。

核心结论

  • 使用Aquilion ONE时,螺旋扫描的晶状体散射剂量在所有三种体模中均显著高于轴扫;使用Revolution CT时,仅5岁体模螺旋扫描的晶状体散射剂量显著高于轴扫;这些结果提示散射剂量可能随CT设备和体型变化,但轴扫通常较低。

研究设计与样本

  • 研究使用两种CT设备:Revolution CT(GE Healthcare,256排,0.625毫米探测器)和Aquilion ONE GENESIS Edition(Canon Medical Systems,320排,0.5毫米探测器);采用三种儿童人体模,在左右晶状体放置光激发光剂量计(OSLD);比较儿童胸部CT检查中螺旋扫描(40毫米探测器宽度)与轴扫(160毫米探测器宽度)的散射剂量。

临床应用与阅片要点

  • 作者认为,尽管散射剂量可能随CT设备和体型变化,但轴扫通常较低。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND: Reports comparing field lens doses between helical scans with a 40-mm detector width and axial scans with a 160-mm detector width using different computed tomography (CT) scanners are currently scarce.

OBJECTIVE: To compare scatter doses for lenses between a helical scan with a 40-mm detector width and an axial scan with a 160-mm detector width when using different CT scanners in the context of pediatric chest examinations.

MATERIALS AND METHODS: Two different CT machines were used: Revolution CT (GE Healthcare, Waukesha, WI) with a 256-row, 0.625-mm multidetector; and Aquilion ONE GENESIS Edition (Canon Medical Systems, Otawara, Japan) with a 320-row, 0.5-mm multidetector. Three pediatric anthropomorphic phantoms were used, with optically stimulated luminescence dosimeters (OSLDs) placed on the left and right lenses. The scatter dose values measured by the OSLDs were compared between a helical scan with a 40-mm detector width and an axial scan with a 160-mm detector width during pediatric chest CT examinations.

RESULTS: Median equivalent doses for the helical and axial scans were 0.12 and 0.12 mSv/mGy for the newborn, 0.17 and 0.16 mSv/mGy for the 1-year-old, and 0.18 and 0.15 mSv/mGy for the 5-year-old, respectively, when using the Revolution CT. With the Revolution CT, no significant differences were observed in the scatter doses between helical and axial scans in the newborn and 1-year-old phantoms. However, the lens scatter dose for the helical scan was approximately 20-35% higher than that for the axial scan in the 5-year-old phantom (P<0.01). The median equivalent doses of eye lenses for the helical and axial scans were 0.12 and 0.07 mSv/mGy for the newborn, 0.07 and 0.05 mSv/mGy for the 1-year-old, and 0.14 and 0.12 mSv/mGy for the 5-year-old, respectively, when using the Aquilion ONE. With the Aquilion ONE, lens scatter doses for the helical scan were approximately 70%, 40%, and 30% higher in the newborn, 1-year-old, and 5-year-old phantoms, respectively, than those for the axial scan (P<0.01).

CONCLUSIONS: When using the Aquilion ONE, lens scatter doses for the helical scan were significantly higher in all three phantoms than those for the axial scan. In contrast, when using the Revolution CT, the lens scatter dose for the helical scan was significantly higher in the 5-year-old phantom than that for the axial scan. These results suggest that although scattered doses may vary with respect to the CT scanner and body size, they are generally lower in the case of axial scans.

来源:Pediatr Radiol,查看原文。版权归原作者所有。

引用

  1. Masuda T, Kiguchi M, Fujioka C, et al. Comparison of pediatric lens scattered dose measurements between axial 40-mm and helical 160-mm detector width computed tomography scan modes. Pediatr Radiol. 2024. PMID: 38769141. doi:10.1007/s00247-024-05947-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。