深度学习图像重建在超低剂量胸部CT诊断肺结节中的临床价值

Clinical value of deep learning image reconstruction on the diagnosis of pulmonary nodule for ultra-low-dose chest CT imaging.
PubMed 文献 · Clin Radiol · 2024-04-20

前瞻性研究纳入142例需肺部检查者,比较超低剂量CT(UL-A 0.57 mSv、UL-B 0.33 mSv)与标准CT(4.32 mSv)及不同重建算法。标准CT检出710个结节(UL-A 358个、UL-B 352个)。超低剂量CT结节检出敏感度高,DLIR较ASIR-V显著降低噪声、提高图像质量及结节测量准确度。

要点

  • 超低剂量CT对肺结节检出敏感度高,50%ASIR-V、DLIR-M与DLIR-H敏感度均为96.90%。
  • DLIR-H在超低剂量CT中噪声、SNR、CNR表现优于50%ASIR-V,主观评分相当或更高。
  • DLIR-H结节体积绝对百分比误差(3.10%)低于50%ASIR-V(8.29%)(P<.001)。

核心结论

  • 超低剂量CT扫描对肺结节检出具有高敏感度;与ASIR-V相比,DLIR可显著降低图像噪声、提高图像质量及结节测量准确度。

研究设计与样本

  • 前瞻性研究纳入142例需肺部检查的参与者,均行超低剂量CT(UL-A 0.57±0.04 mSv或UL-B 0.33±0.03 mSv)与标准CT(SDCT 4.32±0.33 mSv)平扫;以SDCT采用50%强度ASIR-V(50%ASIR-V)为参考标准,超低剂量CT分别用50%ASIR-V、DLIR中等强度(DLIR-M)和高强度(DLIR-H)重建。

诊断标准与关键数值

  • 标准CT共检出710个结节,其中UL-A组358个、UL-B组352个;超低剂量CT中50%ASIR-V、DLIR-M和DLIR-H的结节检出敏感度均为96.90%;多因素分析显示BMI、结节直径和类型是结节检出敏感度的独立预测因素(p<.001);DLIR-H体积绝对百分比误差为3.10%±95.11%,50%ASIR-V为8.29%±99.14%(P<.001)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To compare the image quality and pulmonary nodule detectability between deep learning image reconstruction (DLIR) and adaptive statistical iterative reconstruction-Veo (ASIR-V) in ultra-low-dose CT (ULD-CT).

METHODS: 142 participants required lung examination who underwent simultaneously ULD-CT (UL-A, 0.57 ± 0.04 mSv or UL-B, 0.33 ± 0.03 mSv), and standard CT (SDCT, 4.32 ± 0.33 mSv) plain scans were included in this prospective study. SDCT was the reference standard using ASIR-V at 50% strength (50%ASIR-V). ULD-CT was reconstructed with 50%ASIR-V, DLIR at medium and high strength (DLIR-M, DLIR-H). The noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), and subjective scores were measured. The presence and accuracy of nodules were analyzed using a combination of a deep learning-based nodule evaluation system and a radiologist.

RESULTS: A total of 710 nodules were detected by SDCT, including 358 nodules in UL-A and 352 nodules in UL-B. DLIR-H exhibited superior noise, SNR, and CNR performance, and achieved comparable or even higher subjective scores compared to 50%ASIR-V in ULD-CT. Nodules sensitivity detection of 50%ASIR-V, DLIR-M, and DLIR-H in ULD-CT were identical (96.90%). In multivariate analysis, body mass index (BMI), nodule diameter, and type were independent predictors for the sensitivity of nodule detection (p<.001). DLIR-H provided a lower absolute percent error (APE) in volume (3.10% ± 95.11% vs 8.29% ± 99.14%) compared to 50%ASIR-V of ULD-CT (P<.001).

CONCLUSIONS: ULD-CT scanning has a high sensitivity for detecting pulmonary nodules. Compared with ASIR-V, DLIR can significantly reduce image noise, and improve image quality, and accuracy of the nodule measurement in ULD-CT.