基于人工智能的头颈部CT血管造影自动与手动后处理的比较研究:多厂商多中心数据的独立外部验证
回顾性纳入三家三级医院共250例头颈部CTA,比较CerebralDoc自动后处理与手动后处理的图像质量、狭窄与动脉瘤诊断性能、后处理时间及辐射剂量;自动后处理多数评分不劣于或优于手动,诊断性能相当或更优,后处理时间与辐射剂量更优,但VR上V4段、M1段边缘及MIP上C4段图像质量需进一步优化。
要点
- 自动后处理在多数血管段图像质量不劣于或优于手动后处理。
- 自动后处理对狭窄和动脉瘤的诊断性能与手动后处理相当或更优。
- VR上V4段、M1段边缘及MIP上C4段自动后处理评分显著低于手动。
核心结论
- CerebralDoc自动后处理较手动后处理具有明显优势并具临床价值,但VR上V4段和M1段及MIP上C4段的图像质量需进一步优化。
研究设计与样本
- 回顾性研究,纳入三家三级医院250例患者,CT设备来自五家厂商,以原始CTA数据集及对应多平面重建为参考标准,比较CerebralDoc自动后处理与手动后处理。
关键影像表现/征象
- VR上V4段和M1段边缘评分在自动后处理显著低于手动后处理(各厂商均P<0.05)。
- MIP上C4段评分在自动后处理显著低于手动后处理(各厂商均P<0.05)。
- 自动后处理多数评分显示图像质量优于或相当于手动后处理。
诊断标准与关键数值
- 以原始CTA数据集及对应多平面重建为参考,55例存在显著狭窄(≥50%),33例有动脉瘤。
- 图像质量采用Likert量表评估。
- 自动后处理诊断性能对狭窄和动脉瘤不劣于或优于手动后处理。
- 自动后处理在后处理时间和辐射剂量上优于手动后处理(P<0.001)。
临床应用与阅片要点
- 自动后处理可用于头颈部CTA后处理,多数血管段图像质量与诊断性能不劣于手动。
- 阅片时需注意VR上V4段、M1段边缘及MIP上C4段自动后处理图像质量可能低于手动。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To externally validate the performance of automated postprocessing (AP) on head and neck CT Angiography (CTA) and compare it with manual postprocessing (MP).
METHODS: This retrospective study included head and neck CTA-exams of patients from three tertiary hospitals acquired on CT scanners from five manufacturers. AP was performed by CerebralDoc. The image quality was assessed using Likert scales, and the qualitative and quantitative diagnostic performance of arterial stenosis and aneurysm, postprocessing time, and scanning radiation dose were also evaluated.
RESULTS: A total of 250 patients were included. Among these, 55 patients exhibited significant stenosis (≥ 50%), and 33 patients had aneurysms, diagnosed using original CTA datasets and corresponding multiplanar reconstructions as the reference. While the scores of the V4 segment and the edge of the M1 segment on volume rendering (VR), as well as the C4 segment on maximum intensity projection (MIP), were significantly lower with AP compared to MP across vendors (all P < 0.05), most scores in AP demonstrated image quality that was either superior to or comparable with that of MP. Furthermore, the diagnostic performance of AP was either superior to or comparable with that of MP. Moreover, AP also exhibited advantages in terms of postprocessing time and radiation dose when compared to MP (P < 0.001).
CONCLUSION: The AP of CerebralDoc presents clear advantages over MP and holds significant clinical value. However, further optimization is required in the image quality of the V4 and M1 segments on VR as well as the C4 segment on MIP.
引用
- Li K, Yang Y, Niu S, et al. A Comparative Study of AI-Based Automated and Manual Postprocessing of Head and Neck CT Angiography: An Independent External Validation with Multi-Vendor and Multi-Center Data. Neuroradiology. 2024. PMID: 38753039. doi:10.1007/s00234-024-03379-y
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。