在图像引导放射治疗的头颈部锥形束CT中通过缩小成像野宽度降低晶状体曝光剂量的建议:一项体模研究

Recommendation for reducing the crystalline lens exposure dose by reducing imaging field width in cone-beam computed tomography for image-guided radiation therapy: an anthropomorphic phantom study.
指南 / 共识 · Radiol Phys Technol · 2024-04-30

本研究在头颈部图像引导放射治疗(IGRT)的锥形束CT(CBCT)中,评估了将X射线孔径(叶片)一侧(X2)在X方向收窄对晶状体曝光剂量和位置匹配精度的影响。使用头体模和辐射光致发光剂量计测量,预设野为13.6 cm,叶片X2孔径从10.0 cm以0.5 cm递减至5.0 cm。结果:叶片X2孔径为5.0 cm时,右晶状体曝光剂量最大降低38.7%,左晶状体降低13.2%;叶片X2孔径13.6 cm与5.0 cm之间的位置校正量最大差异为1 mm,自动骨匹配精度相似。作者认为,在头颈部CBCT中,缩小叶片X2孔径是一种在保证位置匹配精度的同时降低晶状体曝光剂量的有用技术。

要点

  • 头颈部CBCT中缩小叶片X2孔径至5.0 cm可使右晶状体曝光剂量降低38.7%、左晶状体降低13.2%。
  • 叶片X2孔径13.6 cm与5.0 cm之间的位置校正量最大差异为1 mm,自动骨匹配精度相似。
  • 作者认为缩小叶片X2孔径可在保证位置匹配精度的同时降低晶状体曝光剂量。

核心结论

  • 在头颈部CBCT中,缩小叶片X2孔径是一种在保证位置匹配精度的同时降低晶状体曝光剂量的有用技术。

研究设计与样本

  • 体模研究:使用头体模,以辐射光致发光剂量计(RPLD,GD-352 M)测量眼表面剂量作为晶状体曝光剂量;预设野13.6 cm,叶片X2孔径从10.0 cm以0.5 cm递减至5.0 cm;自动骨匹配在叶片X2孔径13.6 cm和5.0 cm下各采集5次CBCT图像,头体模从-5.0 mm至+5.0 mm以1.0 mm递增移动。

诊断标准与关键数值

  • 叶片X2孔径为5.0 cm时,右晶状体曝光剂量最大降低38.7%,左晶状体最大降低13.2%。
  • 叶片X2孔径13.6 cm与5.0 cm之间的位置校正量最大差异为1 mm,自动骨匹配精度相似。

临床应用与阅片要点

  • 作者认为,在头颈部CBCT中缩小叶片X2孔径可在保证位置匹配精度的同时降低晶状体曝光剂量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

In cone-beam computed tomography (CBCT) for image-guided radiation therapy (IGRT) of the head, we evaluated the exposure dose reduction effect to the crystalline lens and position-matching accuracy by narrowing one side (X2) of the X-ray aperture (blade) in the X-direction. We defined the ocular surface dose of the head phantom as the crystalline lens exposure dose and measured using a radiophotoluminescence dosimeter (RPLD, GD-352 M) in the preset field (13.6 cm) and in each of the fields when blade X2 aperture was reduced in 0.5 cm increments from 10.0 to 5.0 cm. Auto-bone matching was performed on CBCT images acquired five times with blade X2 aperture set to 13.6 cm and 5.0 cm at each position when the head phantom was moved from - 5.0 to + 5.0 mm in 1.0 mm increment. The maximum reduction rate in the crystalline lens exposure dose was - 38.7% for the right lens and - 13.2% for the left lens when blade X2 aperture was 5.0 cm. The maximum difference in the amount of position correction between blade X2 aperture of 13.6 cm and 5.0 cm was 1 mm, and the accuracy of auto-bone matching was similar. In CBCT of the head, reduced blade X2 aperture is a useful technique for reducing the crystalline lens exposure dose while ensuring the accuracy of position matching.

来源:Radiol Phys Technol,查看原文。版权归原作者所有。

引用

  1. Yoshida T, Sasaki K, Hayakawa T, et al. Recommendation for reducing the crystalline lens exposure dose by reducing imaging field width in cone-beam computed tomography for image-guided radiation therapy: an anthropomorphic phantom study. Radiol Phys Technol. 2024. PMID: 38691308. doi:10.1007/s12194-024-00810-0