利用对魔角不敏感的超短回波时间磁化传递定量评估肌腱变性:牛肌腱体模研究
本研究在3 T磁共振上对8条牛屈肌腱在胶原酶诱导降解前后及与B0场成0、27、55、90度时行三维超短回波时间(UTE)磁化传递(MT)和UTE T2*成像,UTE MTR变异系数低于UTE T2*,以组织病理为参考标准检测变性肌腱的AUC分别为0.918和0.865,UTE MTR对魔角效应不敏感。
要点
- UTE MTR的变异系数为9.64%-11.25%,低于UTE T2*的18.81%-24.06%(P < 0.001),提示UTE MTR更稳健。
- 以组织病理为参考标准,UTE MT检测变性肌腱的AUC为0.918,优于UTE T2*的0.865。
- 在55度采集角度下UTE T2*出现假性高值,而UTE MTR对魔角效应不敏感。
核心结论
- 超短回波时间磁化传递成像是对魔角效应稳健的肌腱变性定量方法,并可在临床合理时间内完成采集。
研究设计与样本
- 体外研究,8条牛屈肌腱在3 T磁共振上行三维UTE MT和UTE T2*研究应用序列(采集时间分别为4分04秒和6分38秒),于酶诱导降解前后及与B0场成0、27、55、90度时成像;肌腱分为两组,第1组(对照)用磷酸盐缓冲液处理,第2组用胶原酶I处理以诱导胶原变性,以组织病理为参考标准。
关键影像表现/征象
- UTE MTR的变异系数为9.64%-11.25%,UTE T2*的变异系数为18.81%-24.06%,UTE MTR显著低于UTE T2*(P < 0.001)。
- 在55度采集角度下UTE T2*出现假性高值,而UTE MTR值稳健,即对魔角效应不敏感。
诊断标准与关键数值
- 以组织病理为参考标准,UTE MT检测变性肌腱的AUC为0.918,UTE T2* mapping的AUC为0.865。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: The aim of this study was to qualitatively and quantitatively assess changes in bovine flexor tendons before and after collagen degradation and at different angles in relation to the static B 0 field using 3-dimensional ultra-short echo time (UTE) magnetization transfer (MT) imaging within a clinically feasible acquisition time.
MATERIALS AND METHODS: Eight bovine flexor tendons were examined at 3 T magnetic resonance imaging including 3-dimensional UTE MT and UTE T2* research application sequences (acquired within 4:04 and 6:38 minutes, respectively) before and after enzyme-induced degradation. The tendons were divided into 2 groups: group 1 (controls) treated with phosphate-buffered saline and group 2 treated with collagenase I to induce collagen degeneration. Magnetic resonance imaging was repeated at 0, 27, 55, and 90 degrees to the B 0 field. To calculate quantitative tissue properties, all tendons were semiautomatically segmented, and changes in quantitative UTE T2* and UTE MT ratios (MTRs) were compared at different angles and between groups. In addition to descriptive statistics, the coefficient of variation was calculated to compare UTE MT and UTE T2* imaging.
RESULTS: Ultra-short echo time MTR showed a significantly lower coefficient of variation compared with UTE T2* values, indicating a more robust imaging method (UTE MTR 9.64%-11.25%, UTE T2* 18.81%-24.06%, P < 0.001). Both methods showed good performance in detecting degenerated tendons using histopathology as reference standard, with UTE MT imaging having a better area under the curve than UTE T2* mapping (0.918 vs 0.865). Falsely high UTE T2* values were detected at the 55 degrees acquisition angle, whereas UTE MTR values were robust, that is, insensitive to the MAE.
CONCLUSIONS: Ultra-short echo time MT imaging is a reliable method for quantifying tendon degeneration that is robust to the MAE and can be acquired in a clinically reasonable time.