聚甲基丙烯酸甲酯软组织标记物与传统乳腺活检标记物长期超声闪烁可检测性及安全性的比较——猪模型临床前研究
本研究在3头猪体内植入聚甲基丙烯酸甲酯(PMMA)、3D打印及Tumark Q标记物,评估局部扰动和不同留置时间后6个月内超声多普勒闪烁的长期安全性与可检测性。结果显示无不良事件,所有标记物闪烁评分随时间逐渐下降,Q标记物平均闪烁评分最高,PMMA标记物在6个月和5个月时仍保持临床可操作的闪烁评分。
要点
- PMMA标记物在猪模型中安全性可接受,无不良事件发生。
- 所有标记物闪烁评分随时间逐渐下降,总体闪烁显著降低(P < 0.001)。
- PMMA标记物在留置6个月和5个月时平均闪烁评分仍具临床可操作性。
核心结论
- 在猪模型中,PMMA标记物安全性可接受,留置6个月后临床可操作的闪烁仍有助于检测猪淋巴结中的PMMA标记物;Q标记物随时间保持的闪烁最强。
研究设计与样本
- 经机构动物护理与使用委员会批准的6个月动物实验,涉及3头猪,植入PMMA(Zimmer Biomet)、3D打印和Tumark Q标记物,每头猪在不同时间植入,部分标记物附近用补片材料或乙醇诱导局部炎症反应,每月行超声半定量闪烁评分,终点行超声引导下可检测标记物淋巴结定位并手术切除后行组织形态计量学分析。
关键影像表现/征象
- 所有三头猪所有标记物的闪烁评分随时间逐渐下降。
- Q标记物平均闪烁评分最高,其次为PMMA标记物、PMMA加补片、Q加乙醇。
- 3D打印标记物加补片和PMMA加乙醇评分最低。
- 所有导丝定位的淋巴结均成功切除。
- 尽管标记物周围组织鞘百分比不同且总体闪烁随时间显著降低(P < 0.001),使用通用电气C1-6探头和9L探头时PMMA平均闪烁评分在6个月和5个月时仍具临床可操作性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: We have studied the use of polymethyl methacrylate (PMMA) as an alternative biopsy marker that is readily detectable with ultrasound Doppler twinkling in cases of in vitro, ex vivo, or limited duration in vivo settings. This study investigates the long-term safety and ultrasound Doppler twinkling detectability of a PMMA breast biopsy marker following local perturbations and different dwell times in a 6-mo animal experiment.
METHODS: This study, which was approved by our Institutional Animal Care and Use Committee, involved three pigs and utilized various markers, including PMMA (Zimmer Biomet), 3D-printed, and Tumark Q markers. Markers were implanted at different times for each pig. Mesh material or ethanol was used to induce a local inflammatory reaction near certain markers. A semiquantitative twinkling score assessed twinkling for actionable localization during monthly ultrasounds. At the primary endpoint, ultrasound-guided localization of lymph nodes with detectable markers was performed. Following surgical resection of the localized nodes, histomorphometric analysis was conducted to evaluate for tissue ingrowth and the formation of a tissue rind around the markers.
RESULTS: No adverse events occurred. Twinkling scores of all markers for all three pigs decreased gradually over time. The Q marker exhibited the highest mean twinkling score followed by the PMMA marker, PMMA with mesh, and Q with ethanol. The 3D-printed marker with mesh and PMMA with ethanol had the lowest scores. All wire-localized lymph nodes were successfully resected. Despite varying percentages of tissue rind around the markers and a significant reduction in overall twinkling (p < 0.001) over time, mean PMMA twinkling scores remained clinically actionable at 6 and 5 mo using a General Electric C1-6 probe and 9L-probe, respectively.
CONCLUSIONS: In this porcine model, the PMMA marker demonstrates an acceptable safety profile. Clinically actionable twinkling aids PMMA marker detection even after 6 mo of dwell time in porcine lymph nodes. The Q marker maintained the greatest twinkling over time compared to all the other markers studied.
引用
- Lee CU, Urban MW, Hesley GK, et al. Long-Term Ultrasound Twinkling Detectability and Safety of a Polymethyl Methacrylate Soft Tissue Marker Compared to Conventional Breast Biopsy Markers-A Preclinical Study in a Porcine Model. Ultrasound Med Biol. 2024. PMID: 38575416. doi:10.1016/j.ultrasmedbio.2024.03.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。