光子计数探测器CT在排泄期检测肾结石:3D打印肾脏体模中虚拟平扫与虚拟非碘重建的对比

Photon-Counting Detector CT for Kidney Stone Detection in Excretory Phase CT-Comparison Between Virtual Non-contrast and Virtual Non-iodine Reconstructions in a 3D Printed Kidney Phantom.
PubMed 文献 · Acad Radiol · 2024-04-18

在3D打印肾脏体模中,于120 kV和140 kV、5/10/15 mGy条件下比较光子计数探测器CT排泄期的虚拟非碘(VNI)与虚拟平扫(VNC)算法。VNC在对比剂减除和结石可见性上准确性高于VNI;辐射剂量和管电压对结石减除无显著影响;两种算法仍常见对比剂减除不彻底和结石部分减除。

要点

  • VNC对对比剂减除和肾结石显示的准确性优于VNI。
  • 两种算法均常见对比剂减除不彻底和结石部分减除。
  • 尿酸结石和小于5 mm结石在两种算法中均更易被减除。

核心结论

  • VNC在对比剂减除和肾结石可见性方面准确性高于VNI。
  • 辐射剂量和管电压对结石减除无显著影响。
  • 两种算法仍频繁出现对比剂减除不彻底和肾结石部分减除。

研究设计与样本

  • 研究类型为体模研究,使用3D打印肾脏体模,内含草酸钙结石和尿酸结石,置于增强肾盏内。
  • 肾盏密度模拟200例排泄期患者的平均密度(110 kV下为916 HU)。
  • 在临床双源光子计数探测器CT上于120 kV和140 kV成像,辐射剂量设为5、10、15 mGy。
  • 应用VNI(PureCalcium)和VNC算法,由两名盲法阅片者使用视觉量表评价图像质量、对比剂减除程度和结石减除程度。

关键影像表现/征象

  • 尿酸结石和小于5 mm的结石在VNC和VNI中均比草酸钙结石和较大结石更易被减除。

诊断标准与关键数值

  • 读者2对VNI的整体图像质量评分高于VNC(4.90 vs 4.00,P < .05),读者1则无显著差异(4.96 vs 5.00,P > .05)。
  • 两名读者对VNC和VNI对比剂减除的一致性均为中等至高度一致(Krippendorff's alpha范围0.628–0.748)。
  • VNI对比剂减除不彻底的发生率:读者1为29% vs VNC 15%(P < .05),读者2为24% vs VNC 20%(P > .05)。
  • 结石减除率VNI高于VNC:读者1为22% vs 16%,读者2为25% vs 10%(P < .05)。

临床应用与阅片要点

  • 在排泄期光子计数探测器CT中,VNC较VNI能更准确地减除对比剂并显示肾结石。
  • 尿酸结石和小于5 mm结石在两种算法中均易被减除,阅片时需注意。

局限性

  • 两种算法仍频繁出现对比剂减除不彻底和肾结石部分减除。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To evaluate and compare the effectiveness of contrast media subtraction and kidney stone detection between a virtual non-iodine reconstruction algorithm (VNI; PureCalcium) and a virtual non-contrast (VNC) algorithm in excretory phase photon-counting detector computed tomography (PCD-CT), using a 3D printed kidney phantom under various tube voltages and radiation doses.

MATERIALS AND METHODS: A 3D-printed kidney phantom, holding Calcium Oxalate (CaOx) and uric acid stones within contrast-enhanced calyces, was created. The calyx density mirrored the average density observed in 200 excretory phase patients (916 HU at 110 kV). Imaging was conducted on a clinical dual-source PCD-CT at 120 kV and 140 kV, with radiation doses set at 5, 10, and 15 mGy. VNI and VNC algorithms were applied. Two blinded readers evaluated the image quality, along with the degree of contrast media and kidney stone subtraction, using visual scales. Krippendorff's alpha was calculated to determine inter-reader agreement, and the Chi-squared test was employed for comparing ordinal data.

RESULTS: Reader 2 rated overall image quality higher for VNI than VNC (4.90 vs. 4.00; P < .05), while Reader 1 found no significant difference (4.96 vs. 5.00; P > .05). Substantial agreement was observed between readers for contrast media subtraction in both VNC and VNI (Krippendorff's alpha range: 0.628-0.748). Incomplete contrast media subtraction occurred more frequently with VNI for both readers (Reader 1: 29% vs. 15%; P < .05; Reader 2: 24% vs. 20%; P > .05). Uric acid and smaller stones (<5 mm) were more likely to be subtracted than CaOx and larger stones in both VNC and VNI. Overall, a higher rate of stone subtraction was noted with VNI compared to VNC (Reader 1: 22% vs. 16%; Reader 2: 25% vs. 10%; P < .05). Neither radiation dose nor tube voltage significantly influenced stone subtraction (P > .05).

CONCLUSION: VNC demonstrated greater accuracy than VNI for contrast media subtraction and kidney stone visibility. Radiation dose and tube voltage had no significant impact. Nonetheless, both algorithms still exhibited frequent incomplete contrast media subtraction and partial kidney stone subtraction.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Breiding PS, Gomollon AMT, Martini K, et al. Photon-Counting Detector CT for Kidney Stone Detection in Excretory Phase CT-Comparison Between Virtual Non-contrast and Virtual Non-iodine Reconstructions in a 3D Printed Kidney Phantom. Acad Radiol. 2024. PMID: 38641450. doi:10.1016/j.acra.2024.04.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。