文献速递腹部儿科影像CT

光子计数CT系统上通用120 kV双源超高螺距方案用于所有体型儿童腹部:与能量积分CT比较的体模研究

Universal 120-kV Dual-Source Ultra-High Pitch Protocol on the Photon-Counting CT System for Pediatric Abdomen of All Sizes: A Phantom Investigation Comparing With Energy-Integrating CT.
PubMed 文献 · Invest Radiol · 2024-04-10

该体模研究比较了光子计数CT(PCCT)采用通用120 kV超高螺距及虚拟单能图像(VMI)方案与能量积分CT(EICT)采用体型依赖kV双源Flash方案在模拟新生儿至成人儿童腹部中的图像质量。结果显示PCCT VMI在70 keV噪声与EICT相当或更低,50 keV碘对比噪声比(CNR)与EICT相当或更高,噪声功率谱曲线形状相似。作者结论为PCCT通用方案在噪声和碘CNR方面表现等效或更优。

要点

  • PCCT通用120 kV超高螺距方案在70 keV VMI噪声与EICT相当或更低。
  • PCCT 50 keV VMI碘CNR与EICT相当或更高。
  • PCCT VMI噪声功率谱曲线形状与EICT相似,低keV噪声更高。

核心结论

  • 通用PCCT 120 kV超高螺距联合后处理VMI在儿童腹部CT噪声(70 keV)和碘CNR(50 keV)方面较EICT体型特异性kV图像表现等效或更优。

研究设计与样本

  • 体模研究,使用1个固体水插入物和3个碘(2、5、10 mg I/mL)插入物置于模拟新生儿、5岁、10岁及成人尺寸儿童腹部的可变尺寸体模中;EICT采用临床体型特异性kV双源方案(螺距3.0),PCCT(Siemens Alpha)采用通用120 kV方案,全剂量CTDI vol分别为新生儿0.30 mGy、5岁0.71 mGy、10岁1.05 mGy、成人7.40 mGy,并采集半剂量扫描;PCCT VMI在40–80 keV以5 keV间隔生成。

诊断标准与关键数值

  • PCCT 70 keV VMI噪声:新生儿7.0 ± 0.6 HU、成人14.7 ± 1.6 HU,与EICT(新生儿7.8 ± 0.8 HU、成人15.3 ± 1.5 HU)相比P > 0.05(t检验)或P < 0.05(t检验);PCCT 50 keV VMI碘CNR:新生儿50.8 ± 8.4、成人27.3 ± 2.8,与EICT(新生儿57.5 ± 6.7、成人13.8 ± 1.7)相比P > 0.05(t检验)或P < 0.05(t检验)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The purpose of this study is to determine if a universal 120-kV ultra-high pitch and virtual monoenergetic images (VMIs) protocol on the photon-counting computed tomography (PCCT) system can provide sufficient image quality for pediatric abdominal imaging, regardless of size, compared with protocols using a size-dependent kV and dual-source flash mode on the energy-integrating CT (EICT) system.

MATERIALS AND METHODS: One solid water insert and 3 iodine (2, 5, 10 mg I/mL) inserts were attached or inserted into phantoms of variable sizes, simulating the abdomens of a newborn, 5-year-old, 10-year-old, and adult-sized pediatric patients. Each phantom setting was scanned on an EICT using clinical size-specific kV dual-source protocols with a pitch of 3.0. The scans were performed with fixed scanning parameters, and the CTDI vol values of full dose were 0.30, 0.71, 1.05, and 7.40 mGy for newborn to adult size, respectively. In addition, half dose scans were acquired on EICT. Each phantom was then scanned on a PCCT (Siemens Alpha) using a universal 120-kV protocol with the same full dose and half dose as determined above on the EICT scanner. All other parameters matched to EICT settings. Virtual monoenergetic images were generated from PCCT scans between 40 and 80 keV with a 5-keV interval. Image quality metrics were compared between PCCT VMIs and EICT, including image noise (measured as standard deviation of solid water), contrast-to-noise ratio (CNR) (measured at iodine inserts with solid water as background), and noise power spectrum (measured in uniform phantom regions).

RESULTS: Noise at a PCCT VMI of 70 keV (7.0 ± 0.6 HU for newborn, 14.7 ± 1.6 HU for adult) is comparable ( P > 0.05, t test) or significantly lower ( P < 0.05, t test) compared with EICT (7.8 ± 0.8 HU for newborn, 15.3 ± 1.5 HU for adult). Iodine CNR from PCCT VMI at 50 keV (50.8 ± 8.4 for newborn, 27.3 ± 2.8 for adult) is comparable ( P > 0.05, t test) or significantly higher ( P < 0.05, t test) to the corresponding EICT measurements (57.5 ± 6.7 for newborn, 13.8 ± 1.7 for adult). The noise power spectrum curve shape of PCCT VMI is similar to EICT, despite PCCT VMI exhibiting higher noise at low keV levels.

CONCLUSIONS: The universal PCCT 120 kV with ultra-high pitch and postprocessed VMIs demonstrated equivalent or improved performance in noise (70 keV) and iodine CNR (50 keV) for pediatric abdominal CT, compared with size-specific kV images on the EICT.

来源:Invest Radiol,查看原文。版权归原作者所有。