文献速递胸部CT

光子计数CT光谱重建在胸部CT中减少线束硬化伪影的潜力

Potential of photon counting computed tomography derived spectral reconstructions to reduce beam-hardening artifacts in chest CT.
PubMed 文献 · Eur J Radiol · 2024-03-29

该回顾性单中心研究对49例胸壁输液港系统肿瘤患者的胸部PCCT数据,测量19种光谱重建下输液港相关低密度和高密度伪影的CT值,并由两位放射科医师进行结构化伪影评估。结果显示keV与碘对比度及伪影强度显著相关,120 keV单能重建可完全减少严重和明显伪影,且不同单能重建间影像所见无显著差异。

要点

  • 120 keV单能重建可完全减少严重和明显伪影,较低keV重建则不能。
  • 锁骨下静脉碘浓度很高时,40 keV图像出现图像变形。
  • 不同单能重建之间影像所见无显著差异。

核心结论

  • PCCT光谱重建可用于胸部肿瘤影像,以减少输液港所致线束硬化伪影。
  • 评估分期图像数据集时,考虑120 keV虚拟单能图像尤其有帮助,因为该重建中伪影相对较低。

研究设计与样本

  • 回顾性单中心研究,纳入49例因肿瘤疾病分期行胸部PCCT的胸壁输液港系统患者。
  • 对19种光谱重建(多能成像、40至190 keV单能重建、碘图及虚拟平扫)测量8个ROI,参考标准为两位放射科医师的结构化伪影与影像所见评估。

关键影像表现/征象

  • 测量了输液港港体相关低密度和高密度伪影、双侧肌肉和血管的平均值及标准差HU值。
  • 120 keV单能重建与较低keV重建相比,可完全减少严重和明显伪影,差异有统计学意义(p < 0.001)。
  • 锁骨下静脉碘浓度很高的病例中,40 keV图像出现图像变形(p < 0.01)。

诊断标准与关键数值

  • keV与碘对比度及伪影强度均显著相关(所有p < 0.001)。
  • 不同单能重建之间影像所见无显著差异(所有p > 0.05)。

临床应用与阅片要点

  • 在肿瘤胸部PCCT分期中评估图像时,考虑使用120 keV虚拟单能图像可能尤其有帮助,因为该重建中伪影相对较低。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Aim of the recent study is to point out a method to optimize quality of CT scans in oncological patients with port systems. This study investigates the potential of photon counting computed tomography (PCCT) for reduction of beam hardening artifacts caused by port-implants in chest imaging by means of spectral reconstructions.

METHOD: In this retrospective single-center study, 8 ROIs for 19 spectral reconstructions (polyenergetic imaging, monoenergetic reconstructions from 40 to 190 keV as well as iodine maps and virtual non contrast (VNC)) of 49 patients with pectoral port systems undergoing PCCT of the chest for staging of oncologic disease were measured. Mean values and standard deviation (SD) Hounsfield unit measurements of port-chamber associated hypo- and hyperdense artifacts, bilateral muscles and vessels has been carried out. Also, a structured assessment of artifacts and imaging findings was performed by two radiologists.

RESULTS: A significant association of keV with iodine contrast as well as artifact intensity was noted (all p < 0.001). In qualitative assessment, utilization of 120 keV monoenergetic reconstructions could reduce severe and pronounced artifacts completely, as compared to lower keV reconstructions (p < 0.001). Regarding imaging findings, no significant difference between monoenergetic reconstructions was noted (all p > 0.05). In cases with very high iodine concentrations in the subclavian vein, image distortions were noted at 40 keV images (p < 0.01).

CONCLUSIONS: The present study demonstrates that PCCT derived spectral reconstructions can be used in oncological imaging of the thorax to reduce port-derived beam-hardening artefacts. When evaluating image data sets within a staging, it can be particularly helpful to consider the 120 keV VMIs, in which the artefacts are comparatively low.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Haag F, Hokamp NG, Overhoff D, et al. Potential of photon counting computed tomography derived spectral reconstructions to reduce beam-hardening artifacts in chest CT. Eur J Radiol. 2024. PMID: 38574510. doi:10.1016/j.ejrad.2024.111448

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。