文献速递儿科影像头颈CT

光子计数探测器CT与能量积分探测器CT在儿童颞骨CT中的图像质量与辐射剂量比较

Comparison of Image Quality and Radiation Dose in Pediatric Temporal Bone CT Using Photon-Counting Detector CT and Energy-Integrating Detector CT.
PubMed 文献 · AJNR Am J Neuroradiol · 2024-09-09

回顾性纳入110例儿童颞骨CT(光子计数探测器CT 52例、能量积分探测器CT 58例),两名独立阅片者用4分法评估4个解剖结构的分辨率及整体图像质量。光子计数探测器CT所有项目评分均显著高于能量积分探测器CT(P<.05),剂量长度乘积降低43%,6岁以下亚组降低超过70%。

要点

  • 光子计数探测器CT在鼓膜、砧镫关节、镫骨脚、蜗轴及整体图像质量评分均显著高于能量积分探测器CT(P<.05)。
  • 光子计数探测器CT较能量积分探测器CT剂量长度乘积降低43%,6岁以下儿童亚组降低超过70%。
  • 阅片者间一致性良好至优秀(组内相关系数0.6-0.81)。

核心结论

  • 光子计数探测器颞骨CT较能量积分探测器CT以更低辐射剂量获得显著更优的图像质量。

研究设计与样本

  • 回顾性研究,共纳入110例儿童颞骨CT扫描,其中光子计数探测器CT 52例、能量积分探测器CT 58例。
  • 两名独立阅片者采用4分法评估鼓膜、砧镫关节、镫骨脚、蜗轴4个解剖结构的空间分辨率及整体图像质量,并评估阅片者间一致性。
  • 比较每例CT的剂量长度乘积,并按年龄(小于3岁、3-5岁、6-11岁、12岁及以上)进行亚组分析。

关键影像表现/征象

  • 光子计数探测器CT鼓膜评分2.9 vs 能量积分探测器CT 2.4。
  • 光子计数探测器CT砧镫关节评分3.6 vs 能量积分探测器CT 2.6。
  • 光子计数探测器CT镫骨脚评分3.2 vs 能量积分探测器CT 2.4。
  • 光子计数探测器CT蜗轴评分3.4 vs 能量积分探测器CT 2.8。
  • 光子计数探测器CT整体图像质量评分3.6 vs 能量积分探测器CT 2.8,所有项目差异均P<.05。

诊断标准与关键数值

  • 采用4分法评估鼓膜、砧镫关节、镫骨脚、蜗轴的空间分辨率及整体图像质量,光子计数探测器CT各项目评分均显著高于能量积分探测器CT(P<.05)。
  • 阅片者间一致性组内相关系数0.6-0.81,为良好至优秀。
  • 光子计数探测器CT较能量积分探测器CT剂量长度乘积降低43%,6岁以下儿童亚组降低超过70%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Currently, there is a lack of research directly comparing photon-counting detector CT (PCD-CT) and energy-integrating detector CT (EID-CT) in pediatric temporal bone CT imaging. The purpose of this study was to compare the image quality and radiation dose of temporal bone CT scans in pediatric patients acquired with PCD-CT and EID-CT.

MATERIALS AND METHODS: The retrospective study included a total of 110 pediatric temporal bone CT scans (PCD-CT, n = 52; EID-CT, n = 58). Two independent readers evaluated the spatial resolution of 4 anatomic structures (tympanic membrane, incudostapedial joint, stapedial crura, and cochlear modiolus) and overall image quality by using a 4-point scale. Interreader agreement was assessed. Dose-length product for each CT was compared, and subgroup analyses were performed based on age (younger than 3 years, 3-5 years, 6-11 years, and 12 years and above).

RESULTS: PCD-CT demonstrated statistically significantly higher scores than EID-CT for all items (tympanic membrane, 2.9 versus 2.4; incudostapedial joint, 3.6 versus 2.6; stapedial crura, 3.2 versus 2.4; cochlear modiolus, 3.4 versus 2.8; overall image quality, 3.6 versus 2.8; P < .05). Interreader agreement ranged from good to excellent (interclass correlation coefficients, 0.6-0.81). PCD-CT exhibited a 43% dose reduction compared with EID-CT, with a particularly substantial reduction of over 70% in the subgroups of children younger than 6 years.

CONCLUSIONS: PCD temporal bone CT achieves significantly superior imaging quality at a lower radiation dose compared with EID-CT.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。

引用

  1. Lee JS, Kim J, Bapuraj JR, et al. Comparison of Image Quality and Radiation Dose in Pediatric Temporal Bone CT Using Photon-Counting Detector CT and Energy-Integrating Detector CT. AJNR Am J Neuroradiol. 2024. PMID: 38589057. doi:10.3174/ajnr.A8276

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。